FDA Enforcement Class II Ongoing

VITROS Immunodiagnostic Products Free T3 Calibrators-IVD are for use with Free T3 Reagent Pack for the quantitative measurement of free triiodothyronine (FT3) in human serum and plasma (EDTA or heparin). Product Code: 111 2820

Recall: Z-1819-2024 · Reported May 22, 2024

Enforcement

Recall Number
Z-1819-2024
Event ID
94436
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
QUIDELORTHO
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 22, 2024
Initiation Date
April 9, 2024
Classification Date
May 14, 2024
Address
100 Indigo Creek Dr, N/A, Rochester, NY, 14626-5101, United States

Description

VITROS Immunodiagnostic Products Free T3 Calibrators-IVD are for use with Free T3 Reagent Pack for the quantitative measurement of free triiodothyronine (FT3) in human serum and plasma (EDTA or heparin). Product Code: 111 2820

Reason

Free T3 Calibrators used with T3 Reagent Packs may produce positively biased results for patient and quality control samples. The positive bias may cause erroneous patient results or a delay in testing that may impact patient management, clinical assessment, and the results of other thyroid function tests.

Code Info

UDI-DI: 10758750008438 Lot Numbers: 3042 Expiry: 09-Apr-2024; 3052 Expiry: 02-May-2024; 3061 Expiry: 06-Jun-2024; 3080 Expiry: 02-May-2024; 3090 Expiry: 14-Jun-2024; 3100 Expiry: 29-Jul-2024; 3110 Expiry: 13-Aug-2024; 3126 Expiry: 04-Sep-2024

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore , Spain, Sweden, The Netherlands, United Kingdom.

Quantity

N/A