ORTHOFIX Centurion POCT System Drill Guide PN 69-1010 and PN 69-1012. Drill Guides are used to interface with various drill tips during the drilling procedure of bone preparation.
Recall: Z-1819-2015
·
Reported July 1, 2015
Enforcement
- Recall Number
- Z-1819-2015
- Event ID
- 71262
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Orthofix, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 1, 2015
- Initiation Date
- May 11, 2015
- Classification Date
- June 22, 2015
- Termination Date
- February 17, 2016
- Address
- 3451 Plano Pkwy, Lewisville, TX, 75056-9453, United States
Description
ORTHOFIX Centurion POCT System Drill Guide PN 69-1010 and PN 69-1012. Drill Guides are used to interface with various drill tips during the drilling procedure of bone preparation.
Reason
There is a possibility that the thumb button remains in the depressed (unlocked) position allowing the threaded sleeve (drill stop) to unintentionally rotate and change the depth setting.
Code Info
All lot codes
Distribution
Nationwide Distribution including CA, VA, FL, ID, CO, OR, PA, OK, SC, and TX.
Quantity
36 units