Custom Spine ISSYS LP Inverted Screw System. 8.5MM X 45 MM ISSYS LP Screw Rx Non-Sterile. Custom Spine, Inc. 9 Campus Dr. Parsippany, NJ 07054 USA. TEL: 973-808-0019 The ISSYS LP Spinal Fixation System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the lumbar and/or sacral spine for pedicular and non-pedicular fixation.
Enforcement
- Recall Number
- Z-1819-2013
- Event ID
- 65068
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Custom Spine, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- August 7, 2013
- Initiation Date
- March 28, 2013
- Classification Date
- July 29, 2013
- Termination Date
- October 24, 2013
- Address
- 1140 Parsippany Blvd Ste 201, N/A, Parsippany, NJ, 07054-1887, United States
Description
Custom Spine ISSYS LP Inverted Screw System. 8.5MM X 45 MM ISSYS LP Screw Rx Non-Sterile. Custom Spine, Inc. 9 Campus Dr. Parsippany, NJ 07054 USA. TEL: 973-808-0019 The ISSYS LP Spinal Fixation System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the lumbar and/or sacral spine for pedicular and non-pedicular fixation.
On March 28, 2013, Custom Spine, Inc. issued a minor removal of a lot of 18 screws that were found to be out of specification. The screw length was 5 mm shorter in length than the required specification for the screw size. The product is laser marked 8.5mm X 45mm, but its length measured 40mm.
Catalog Number 048545; Lot Number L363
Nationwide Distribution including Nebraska and Virginia.
18 units