FDA Enforcement
Class II
Terminated
Q-Tel. The Q-Tel RMS system is a computer-based cardiac and pulmonary rehabilitation data acquisition and editing system.
Recall: Z-1818-2020
·
Reported May 6, 2020
Enforcement
- Recall Number
- Z-1818-2020
- Event ID
- 85297
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- WELCH ALLYN, INC/MORTARA
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 6, 2020
- Initiation Date
- March 18, 2020
- Classification Date
- April 29, 2020
- Termination Date
- February 9, 2026
- Address
- 7865 N 86th St, N/A, Milwaukee, WI, 53224-3431, United States
Description
Q-Tel. The Q-Tel RMS system is a computer-based cardiac and pulmonary rehabilitation data acquisition and editing system.
Reason
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
Code Info
Same as X12+ Model Number X12PLUS-XXX-XXXXX
Distribution
US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.
Quantity
78,521 units total