FDA Enforcement Class III Ongoing

Starling Monitor, Product code CMMST5 and Starling Monitor SV, Product code CMM-ST5 (same products); and accessories/spare parts Starling Battery Pack, Product Code CMABATT; and Starling Patient Cable, Product code CMASC10.

Recall: Z-1817-2022 · Reported October 5, 2022

Enforcement

Recall Number
Z-1817-2022
Event ID
90809
Classification
Class III
Status
Ongoing
Product Type
Devices
Firm
Baxter Healthcare Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 5, 2022
Initiation Date
September 2, 2022
Classification Date
September 26, 2022
Address
1 Baxter Pkwy, N/A, Deerfield, IL, 60015-4625, United States

Description

Starling Monitor, Product code CMMST5 and Starling Monitor SV, Product code CMM-ST5 (same products); and accessories/spare parts Starling Battery Pack, Product Code CMABATT; and Starling Patient Cable, Product code CMASC10.

Reason

Labeling has incorrect or incomplete Unique Device Identification (UDI) information; and the barcodes are not compliant with GS-1 standards.

Code Info

All serial numbers after 6/2016. Product code CMMST5 - Incorrect UDI-DI: 7290013030269; correct UDI-DI: 07290013030269. Product code CMM-ST5 - Incorrect UDI-DI: 7290013030030; correct UDI-DI: 07290013030030. Product code CMABATT - Incorrect UDI-DI: 7290013030160; correct UDI-DI: 07290013030160. Product code CMASC10 - Incorrect UDI-DI: 7290013030054; correct UDI-DI: 07290013030054.

Distribution

US Nationwide Distribution and government/military distribution.

Quantity

9,671 units