Starling Monitor, Product code CMMST5 and Starling Monitor SV, Product code CMM-ST5 (same products); and accessories/spare parts Starling Battery Pack, Product Code CMABATT; and Starling Patient Cable, Product code CMASC10.
Enforcement
- Recall Number
- Z-1817-2022
- Event ID
- 90809
- Classification
- Class III
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Baxter Healthcare Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 5, 2022
- Initiation Date
- September 2, 2022
- Classification Date
- September 26, 2022
- Address
- 1 Baxter Pkwy, N/A, Deerfield, IL, 60015-4625, United States
Description
Starling Monitor, Product code CMMST5 and Starling Monitor SV, Product code CMM-ST5 (same products); and accessories/spare parts Starling Battery Pack, Product Code CMABATT; and Starling Patient Cable, Product code CMASC10.
Labeling has incorrect or incomplete Unique Device Identification (UDI) information; and the barcodes are not compliant with GS-1 standards.
All serial numbers after 6/2016. Product code CMMST5 - Incorrect UDI-DI: 7290013030269; correct UDI-DI: 07290013030269. Product code CMM-ST5 - Incorrect UDI-DI: 7290013030030; correct UDI-DI: 07290013030030. Product code CMABATT - Incorrect UDI-DI: 7290013030160; correct UDI-DI: 07290013030160. Product code CMASC10 - Incorrect UDI-DI: 7290013030054; correct UDI-DI: 07290013030054.
US Nationwide Distribution and government/military distribution.
9,671 units