FDA Enforcement Class II Ongoing

Synthes Large External Fixation, MR Conditional, Tube to Tube Clamp, Part Number 390.007, Lot Number 6800599 Synthes Large External Fixation is intended for use to provide treatment for long bone and pelvic fractures that require external fixation.

Recall: Z-1816-2013 · Reported August 7, 2013

Enforcement

Recall Number
Z-1816-2013
Event ID
65669
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Synthes USA HQ, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 7, 2013
Initiation Date
October 12, 2012
Classification Date
July 29, 2013
Address
1302 Wrights Ln E, N/A, West Chester, PA, 19380-3417, United States

Description

Synthes Large External Fixation, MR Conditional, Tube to Tube Clamp, Part Number 390.007, Lot Number 6800599 Synthes Large External Fixation is intended for use to provide treatment for long bone and pelvic fractures that require external fixation.

Reason

It was discovered internally that there was a discrepancy between the insert provided with the Tube to Tube Clamp and what was called for in the Bill of Materials. Following a detailed investigation, it was determined that the incorrect insert was used which indicates that the device is MR Safe, when it is actually MR Conditional.

Code Info

Part Number 390.007, Lot Number 6800599

Distribution

US Distribution including the states of CA, MN, NC and TN.

Quantity

8