FDA Enforcement Class II Terminated

H12+ and patient cable for the H12+ Holter Recorders with the following device name: PAT CBL 10WIRE LEADFORM AHA SNAP, PAT CBL 10WIRE LEADFORM IEC SNAP, PAT CBL 10WIRE LEADFORM XL AHA SNAP, PAT CBL 10WIRE LEADFORM XL IEC SNAP. The 24-hour H12+ recorder uses a single AA alkaline battery to provide continuous 12-lead data recorded over a 24-hour period and a removable 24-hour compact flash (CF) card for data storage.

Recall: Z-1812-2020 · Reported May 6, 2020

Enforcement

Recall Number
Z-1812-2020
Event ID
85297
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
WELCH ALLYN, INC/MORTARA
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 6, 2020
Initiation Date
March 18, 2020
Classification Date
April 29, 2020
Termination Date
February 9, 2026
Address
7865 N 86th St, N/A, Milwaukee, WI, 53224-3431, United States

Description

H12+ and patient cable for the H12+ Holter Recorders with the following device name: PAT CBL 10WIRE LEADFORM AHA SNAP, PAT CBL 10WIRE LEADFORM IEC SNAP, PAT CBL 10WIRE LEADFORM XL AHA SNAP, PAT CBL 10WIRE LEADFORM XL IEC SNAP. The 24-hour H12+ recorder uses a single AA alkaline battery to provide continuous 12-lead data recorded over a 24-hour period and a removable 24-hour compact flash (CF) card for data storage.

Reason

Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

Code Info

Model Numbers: H12PLUS-XXX-XXXXX, 9293-017-50, 9293-017-51, 9293-026-50, 9293-026-51.

Distribution

US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.

Quantity

78,521 units total