FDA Enforcement Class II Terminated

Lead diagnostic electrocardiograph under the following device names: ELI 10, ELI 350, ELI 150c, ELI 250c, ELI 230, ELI 280, ELI 380, R-Scribe and Connex Cardio. Product Usage: A non-invasive prescription device indicated for use to acquire, analyze, display, transmit and print electrocardiograms.

Recall: Z-1810-2020 · Reported May 6, 2020

Enforcement

Recall Number
Z-1810-2020
Event ID
85297
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
WELCH ALLYN, INC/MORTARA
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 6, 2020
Initiation Date
March 18, 2020
Classification Date
April 29, 2020
Termination Date
February 9, 2026
Address
7865 N 86th St, N/A, Milwaukee, WI, 53224-3431, United States

Description

Lead diagnostic electrocardiograph under the following device names: ELI 10, ELI 350, ELI 150c, ELI 250c, ELI 230, ELI 280, ELI 380, R-Scribe and Connex Cardio. Product Usage: A non-invasive prescription device indicated for use to acquire, analyze, display, transmit and print electrocardiograms.

Reason

Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

Code Info

Model Numbers: ELI10-XXX-XXXXX, ELI350-XXX-XXXXX, ELI150C-XXX-XXXXX, ELI250C-XXX-XXXXX, ELI230-XXX-XXXXX, ELI280-XXX-XXXXX, ELI380-XYYZZ, RSCRIBE-xxx-xxxxx and CC-XXX-XXXX.

Distribution

US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.

Quantity

78,521 units total