FDA Enforcement Class II Terminated

Navigation System II-Cart, 7700-100-000 with Articulated Arm Camera, 7700-103-001 and Articulated Arm Camera, 7700-103-001U Product Usage: The Navigation System II-Cart is intended to be used as a platform for computer assisted surgery as part of the Stryker Navigation System. The intended use of the complete Stryker Navigation System is defined by the Stryker application modules.

Recall: Z-1810-2015 · Reported June 24, 2015

Enforcement

Recall Number
Z-1810-2015
Event ID
71268
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Stryker Instruments Div. of Stryker Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 24, 2015
Initiation Date
May 4, 2015
Classification Date
June 16, 2015
Termination Date
May 16, 2016
Address
4100 E Milham Ave, Portage, MI, 49002-9704, United States

Description

Navigation System II-Cart, 7700-100-000 with Articulated Arm Camera, 7700-103-001 and Articulated Arm Camera, 7700-103-001U Product Usage: The Navigation System II-Cart is intended to be used as a platform for computer assisted surgery as part of the Stryker Navigation System. The intended use of the complete Stryker Navigation System is defined by the Stryker application modules.

Reason

Stryker initiated a device recall of the Navigation System II-Cart, Camera Articulated Arm because there is a potential that the weld seam between the main stud and the two flaps on the Articulated Arm was not welded in the correct location, which may result in a potential failure of the arm joint.

Code Info

Serial numbers 100019, 100984, 100965, 100295, 100979, 100773, 100946, 100797, 100947, 100956, 100966, 100649, 100967, 100944, 100941, 100945, 100970, 100950, 100976, 100951, 100728, 100952, 100248, 100953, 100641, 100954, 100463, 100955, 100981, 100957, 100978, 100958, 100106, 100959, 100939, 100960, 100973, 100961, 100236, 100962, 100237, 100963, 100417, 100964. 100940, 100968, 100975, 100969, 100977, 100972, 100105, 100974, 100214, 100980, 100492, 100982, 100493, 100983, 100948, 100986, 100704, 100987, 100949, 100988, 100227, 100989, 100971, 100133, 100134, 100942.

Distribution

Worldwide Distribution - US Nationwide in the states of AR, CO, FL, GA, IN, KS, KY, MA, MI, NC, NE, NJ, NY, OH, OK, PA, TX, WA, WY and the countries of Canada, Europe, and Asia.

Quantity

72