FDA Enforcement
Class II
Ongoing
META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690
Recall: Z-1806-2025
·
Reported May 28, 2025
Enforcement
- Recall Number
- Z-1806-2025
- Event ID
- 96586
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Smith & Nephew, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 28, 2025
- Initiation Date
- March 18, 2025
- Classification Date
- May 19, 2025
- Address
- 1450 E Brooks Rd, N/A, Memphis, TN, 38116-1804, United States
Description
META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690
Reason
Integrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, which causes the screw to be non-functional, which would require replacement with another adequate size screw, and exchange may exceed 30 minutes.
Code Info
UDI-DI: 00885556132975. Lot: 24CT81426. Expiration: 25-Mar-2034
Distribution
Worldwide - US Nationwide distribution in the states of FL, NH, TX, CA, OH, NY, AL, IA, MA, NV, MO, WA, LA, TN, NC, KY, NM, NJ, AZ and the country of ZA.
Quantity
75