RayStation software version 4.0.3.4 RayStation is a software system designed for treatment planning and analysis of radiation therapy.
Enforcement
- Recall Number
- Z-1806-2014
- Event ID
- 68162
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- RAYSEARCH LABORATORIES AB
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- June 25, 2014
- Initiation Date
- April 25, 2014
- Classification Date
- June 16, 2014
- Termination Date
- June 26, 2017
- Address
- Sveavaegen 9, N/A, Stockholm, N/A, N/A, Sweden
Description
RayStation software version 4.0.3.4 RayStation is a software system designed for treatment planning and analysis of radiation therapy.
This notice concerns a problem with the computation of ROI voxel volumes for cases where the CT has variable slice spacing. The dose grid volumetric representation of an ROI can be wrong. This affects all dose-volume properties for such ROIs, including DVH, dose statistics, clinical goals and constraints or objective functions. Furthermore, if material override ROIs are used, or if the external is not represented by contours, the bug can trigger an error in the 3D dose distribution.
Software version 4.0.3.4
Worldwide Distribution-USA (nationwide) including the states of AZ, CA, FL, IL, MA, MD, MO, MS, NJ, NV, NY, OH, PA, TN, VA, and WA, and the countries of Australia, Belgium, Germany, France, Italy, New Zealand, Republic of Korea, Switzerland, Spain, and United Kingdom.
25 sites/245 units (US), 25 sites/124 units (foreign)