FDA Enforcement Class II Terminated

RayStation software version 4.0.3.4 RayStation is a software system designed for treatment planning and analysis of radiation therapy.

Recall: Z-1806-2014 · Reported June 25, 2014

Enforcement

Recall Number
Z-1806-2014
Event ID
68162
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
RAYSEARCH LABORATORIES AB
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
June 25, 2014
Initiation Date
April 25, 2014
Classification Date
June 16, 2014
Termination Date
June 26, 2017
Address
Sveavaegen 9, N/A, Stockholm, N/A, N/A, Sweden

Description

RayStation software version 4.0.3.4 RayStation is a software system designed for treatment planning and analysis of radiation therapy.

Reason

This notice concerns a problem with the computation of ROI voxel volumes for cases where the CT has variable slice spacing. The dose grid volumetric representation of an ROI can be wrong. This affects all dose-volume properties for such ROIs, including DVH, dose statistics, clinical goals and constraints or objective functions. Furthermore, if material override ROIs are used, or if the external is not represented by contours, the bug can trigger an error in the 3D dose distribution.

Code Info

Software version 4.0.3.4

Distribution

Worldwide Distribution-USA (nationwide) including the states of AZ, CA, FL, IL, MA, MD, MO, MS, NJ, NV, NY, OH, PA, TN, VA, and WA, and the countries of Australia, Belgium, Germany, France, Italy, New Zealand, Republic of Korea, Switzerland, Spain, and United Kingdom.

Quantity

25 sites/245 units (US), 25 sites/124 units (foreign)