FDA Enforcement Class II Terminated

VITROS (R) Chemistry Products FS Calibrator 1, Product Code/ Catalog Number 6801873 -- For in vitro diagnostic use only --- VITROS Chemistry Products FS Calibrator 1 is used in conjunction with VITROS Chemistry Products Calibrator Kits to calibrate the VITROS 5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Recall: Z-1805-2014 · Reported June 25, 2014

Enforcement

Recall Number
Z-1805-2014
Event ID
68286
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ortho-Clinical Diagnostics
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 25, 2014
Initiation Date
May 12, 2014
Classification Date
June 16, 2014
Termination Date
March 23, 2017
Address
1000 Lee Road, N/A, Rochester, NY, 14606, United States

Description

VITROS (R) Chemistry Products FS Calibrator 1, Product Code/ Catalog Number 6801873 -- For in vitro diagnostic use only --- VITROS Chemistry Products FS Calibrator 1 is used in conjunction with VITROS Chemistry Products Calibrator Kits to calibrate the VITROS 5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Reason

Calibrator values assigned for the calibrator kit lots were released using a reference method that was undergoing validation for a new lot of European Reference Material (ERM). The validation process was not complete prior to calibrator value assignment for VITROS (R) Chemistry Products hsCRP Reagent (Product Code 6801739), Generation (GEN) 28.

Code Info

Lots 91131 (exp. 07-Mar-2015) and 91241 (exp. 28-Feb-2016)

Distribution

Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.

Quantity

Domestic - 464 units; Foreign - 361 units