iLab Polaris Multi-Modality Guidance System; iLab Ultrasound Imaging System
Enforcement
- Recall Number
- Z-1804-2020
- Event ID
- 85271
- Classification
- Class II
- Status
- Completed
- Product Type
- Devices
- Firm
- Boston Scientific Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 6, 2020
- Initiation Date
- September 18, 2019
- Classification Date
- April 27, 2020
- Address
- 300 Boston Scientific Way, N/A, Marlborough, MA, 01752-1291, United States
Description
iLab Polaris Multi-Modality Guidance System; iLab Ultrasound Imaging System
Certain iLab Polaris Systems which were upgraded with a previous 2.10 software version (disk batch 629774) contained a bug in which only the True Color Secondary Capture (SC) format will archive Diastolic Hyperemia-Free Ratio data as intended. If either Multi-Frame UltraSound (US) or SC Image Storage formats are used, the DFR data will not be archived and the user will not be notified of this unexpected behavior.
Polaris 2.10 software version (disk batch 629774); iLab Polaris Systems affected product listing (serial numbers): 200204, 200209, 200217, 200308, 200310, 200448, 200450, 200456, 200457, 200458, 200461,200480, 200481, 200103, 200145, 200044, 200118, 200360, 200498, 200170, 8361, 10312, 10443, 10548, 7402, 8397, 8497, 8498, 8508, 8516, 8522, 8523, 8530, 10044, 10053, 10275, 10278,10362, 10365, 10496, 10500, 10565, 10566, 10604, 10651, 10667, 10694, 10753, 10770, 10867, 10872, 10873, 10875, 10876, 10952, 8475, 7402, 8397, 8497, 8498, 8508, 8516, 8522, 8523, 8530, 10044, 10053, 10275, 10278, 10362, 10365, 10496, 10500, 10565, 10566, 10604, 10651, 10667, 10694, 10753, 10770, 10867, 10872, 10873, 10875, 10876, 10952, 8475, 7402, 8397, 8497, 8498, 8508, 8516, 8522, 8523, 8530, 10044, 10053, 10275, 10278, 10362, 10365, 10496, 10500, 10565, 10566, 10604, 10651, 10667, 10694, 10753, 10770, 10867, 10872, 10873, 10875, 10876, 10952, 8475, 6582 , 7023, 7857, 5637, 5716, 6650, 6894, 4031, 5003, 5043, 5060, 5109, 5232, 5239, 5393, 5465, 5651, 5678, 5860, 6193, 6350, 8264, 6819, 6921, 6934, 7332, 7341, 7362, 7733, 7828, 7858, 7933, 7934, 8011, 8155, 8290, 6383, 5182, 5219, 5296, 5398, 8157, 6760, 7723, 10108, 8365, 8367, 10504, 10657, 10501, 4274, 8046, 8048, 5892, 7818, 4773, 5585, 8113, 5165, 4402, 6468, 6319, 5313, 7511, 4952, 5277, 5457, 4116, 5403, 5078, 5664, 5666, 6327, 6733, 6994, 6136, 6206, 6172, 7604, 7602, 7353, 6273, 6993, 4045, 10221, 10261, and 8540.
OUS: Japan, Belgium, Denmark, France, Great Britain, and Spain. No U.S. consignees.
150 units