FDA Enforcement Class I Ongoing

B. Braun Hemodialysis Bloodlines, STREAMLINE FRESENIUS FOR DAVITA. Model Number: SL-2000M2095DA.

Recall: Z-1800-2026 · Reported April 22, 2026

Enforcement

Recall Number
Z-1800-2026
Event ID
98552
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
B Braun Medical Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 22, 2026
Initiation Date
March 19, 2026
Classification Date
April 15, 2026
Address
824 12th Ave, N/A, Bethlehem, PA, 18018-3524, United States

Description

B. Braun Hemodialysis Bloodlines, STREAMLINE FRESENIUS FOR DAVITA. Model Number: SL-2000M2095DA.

Reason

Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.

Code Info

Model Number: SL-2000M2095DA; UDI-DI Primary: 04046955776023; UDI-DI Unit: 04046955776016; All lots manufactured since 09JUN2025.

Distribution

Worldwide distribution - US Nationwide and the country of Canada.

Quantity

1,174,271 units