FDA Enforcement Class II Terminated

FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2), Laboratory Automation System

Recall: Z-1798-2018 · Reported May 16, 2018

Enforcement

Recall Number
Z-1798-2018
Event ID
79645
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Inpeco S.A.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
May 16, 2018
Initiation Date
February 13, 2018
Classification Date
May 10, 2018
Termination Date
April 3, 2019
Address
Via San Gottardo 10, N/A, Lugano, N/A, N/A, Switzerland

Description

FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2), Laboratory Automation System

Reason

Module may freeze without generating user warning. There is a potential risk in delay of sample processing, leading to delayed delivery of test results to patients.

Code Info

Serial Numbers: Accelerator a3600 - ACP.150 ACP.185 ACP.186 ACP.192 ACP.202 ACP.214 FlexLab - FLX.0165 FLX.0127 FLX.0128 FLX.0133 FLX.0157 FLX.0160 FLX.0181 Aptio Automation - AP2.0054 AP2.0056 AP2.0060 AP2.0078 AP2.0095 Part Numbers of impacted sample tube carriers: 721501001 for 43 mm carriers, and 0A00006957 for 40 mm carriers

Distribution

Distributors in 2 states: NY and IL.

Quantity

18