FDA Enforcement Class II Terminated

Axiom Aristos MX Solid State Flat Panel Digital Imager, Model number 5895003. The device is used for diagnostic X-ray imaging.

Recall: Z-1795-2013 · Reported July 31, 2013

Enforcement

Recall Number
Z-1795-2013
Event ID
65604
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 31, 2013
Initiation Date
June 5, 2013
Classification Date
July 24, 2013
Termination Date
February 24, 2015
Address
51 Valley Stream Pkwy, N/A, Malvern, PA, 19355, United States

Description

Axiom Aristos MX Solid State Flat Panel Digital Imager, Model number 5895003. The device is used for diagnostic X-ray imaging.

Reason

Siemens issued a Field Safety Notice about the potential hazard to patients or operators when using the function "table top float" of the AXIOM Aristos MX. When repositioning a patient lying on the table top and releasing the table top, it is possible that the fingers of the patient or operator are between the table top and the detector tray and if the table top is moved, the fingers can be squee

Code Info

Model Number 5895003 with serial numbers 1052 through 2500.

Distribution

Nationwide Distribution.

Quantity

401