FDA Enforcement Class II Terminated

Siemens N Latex CDT Kit- IVD quantitative determination of carbohydrate-deficient transferrin (CDT) in human serum. Siemens Material Number (SMN): 10445997

Recall: Z-1790-2020 · Reported April 29, 2020

Enforcement

Recall Number
Z-1790-2020
Event ID
85282
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
CSL Behring GmbH
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
April 29, 2020
Initiation Date
March 2, 2020
Classification Date
April 23, 2020
Termination Date
October 5, 2020
Address
Emil-Von-Behring-Str. 76, N/A, Marburg, N/A, N/A, Germany

Description

Siemens N Latex CDT Kit- IVD quantitative determination of carbohydrate-deficient transferrin (CDT) in human serum. Siemens Material Number (SMN): 10445997

Reason

Negative bias for the carbohydrate-deficient transferrin (CDT) measurement in comparison to the HPLC method, observed to occur after 15 months based on shelf life monitoring

Code Info

US Lot number: 49822 UDI: 842768018534 OUS: Lot numbers: 49729, 49849, 49985, 50084 Exp. Date: 2020-04-11

Distribution

IN, KS, MO, OH, VA Foreign: Australia, Austria, Bangladesh, Belgium, Canada, China Czech Republic, Egypt, Finland, France, Germany, Great Britain, Hungary India, Italy, Japan, Mexico, Netherlands, Norway, Poland, Portugal, Republic Korea, Russia Federation, Slovakia, Slovenia, South Africa, Spain Sweden, Switzerland, Taiwan

Quantity

297 PCK US; 1247 PCK OUS