FDA Enforcement Class II Terminated

The Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery. The system consists of a water tank, circulating pumps, heater manifolds, mercury free temperature sensors, water detectors, mixing valves and a tank divider which is provided to partition the tank into two separate channels (Left and Right). The system has the capacity to circulate water at a rate of up to 6.5 gal./min (25 L/min) with no load connected. The system is capable of heating and cooling for a single channel or for both channels. Device Name / Model Number: HX2 Temperature Management System (P/N 809810) Catalog Number: 809810

Recall: Z-1789-2021 · Reported June 9, 2021

Enforcement

Recall Number
Z-1789-2021
Event ID
87834
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Terumo Cardiovascular Systems Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 9, 2021
Initiation Date
April 16, 2021
Classification Date
June 4, 2021
Termination Date
November 29, 2023
Address
6200 Jackson Rd, Ann Arbor, MI, 48103-9586, United States

Description

The Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery. The system consists of a water tank, circulating pumps, heater manifolds, mercury free temperature sensors, water detectors, mixing valves and a tank divider which is provided to partition the tank into two separate channels (Left and Right). The system has the capacity to circulate water at a rate of up to 6.5 gal./min (25 L/min) with no load connected. The system is capable of heating and cooling for a single channel or for both channels. Device Name / Model Number: HX2 Temperature Management System (P/N 809810) Catalog Number: 809810

Reason

Terumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol will not be developed by Terumo CVS and it has been determined that the best course of action is for users to discontinue use of and dispose of HX2, TCM I and TCM II devices.

Code Info

All lot numbers distributed from 05/02/1985 thru 06/10/2015

Distribution

Domestic: Foreign: Australia, Belgium, Canada, Chile, China, Colombia, Dominican Republic, England, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Israel, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, New Zealand, Norway, Philippines, Russia, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), Vietnam

Quantity

75 devices