Philips/Children's Medical Ventures Smart Monitor 2 Professional Series infant apnea monitor.
Enforcement
- Recall Number
- Z-1789-2014
- Event ID
- 68226
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Childrens Medical Ventures
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 25, 2014
- Initiation Date
- April 29, 2014
- Classification Date
- June 13, 2014
- Termination Date
- April 6, 2015
- Address
- 191 Wyngate Dr, N/A, Monroeville, PA, 15146-3045, United States
Description
Philips/Children's Medical Ventures Smart Monitor 2 Professional Series infant apnea monitor.
The battery pack wire harness is improperly assembled rendering the device inoperable. The wiring issue causes an error code to be displayed and results in the device sounding an alarm repeatedly at power up. The presence of the error code and the devices alarm prevents the device from functioning until the issue is addressed.
989805603611 / 1014557, 989805603631 / 1014559, 989805604511 / 1028969, 989805604521 / 1028970, 989805604531 / 1028971, 989805604631 / 1030271, 989805605051 / 1038140, 989805605061 / 1038141, 989805608801 / U1014557, 989805608811 / U1014559, 989805608831 / U1028969, 989805608841 / U1028970, 989805608851 / U1038140, 989805604231 / 1023384
Worldwide Distribution - USA including the states of MI, WI, SC, FL, TX, KS, MA, AZ, UT, WV, IN, MN, NV and VA., and the countries of Australia, Belgium and Canada.