FDA Enforcement Class II Terminated

GE Healthcare Definium 8000 Digital Radiographic Systems : GE Healthcare, 3000 North Grandview , Waukesha, WI, 53188 Model Number: 5131070. The Revolution XR/d with Tomosynthesis is intended for use in generating Tomographic images of human anatomy. It is not intended for mammographic use.

Recall: Z-1789-2012 · Reported June 20, 2012

Enforcement

Recall Number
Z-1789-2012
Event ID
37654
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
General Electric Med Systems LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 20, 2012
Initiation Date
May 31, 2006
Classification Date
June 8, 2012
Termination Date
July 31, 2012
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615, United States

Description

GE Healthcare Definium 8000 Digital Radiographic Systems : GE Healthcare, 3000 North Grandview , Waukesha, WI, 53188 Model Number: 5131070. The Revolution XR/d with Tomosynthesis is intended for use in generating Tomographic images of human anatomy. It is not intended for mammographic use.

Reason

An incident was reported that an operator's finger was pinched between the fixed roller and the gliding tabletop of a Proteus XR/a Table while moving the tabletop with the thumb placed over its edge and fingers underneath

Code Info

All serial numbers

Distribution

Worldwide Distribution.

Quantity

3,796 units