FDA Enforcement Class I Terminated

GE Healthcare Varicam/VG and VG Hawkeye devices Product Usage: The intended use of the Quasar system is to perform general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition typed and optional imaging features designed to enhance image quality in Oncology, Cardiology, Neurology and other clinical diagnostic imaging applications. The scanning modes include planar (Static, Multi-gated, Dynamic, Whole body scanning) and tomographic (SPECT, Gated SPECT, Whole body SPECT, Camera based PET). Acquisition types include single and multi-isotope/multi-peak frame/list mode single-photon and positron imaging. Optional imaging-enhancement features include assortment of collimators, gating by physiological signals, real-time automatic body contouring, and CT-based attenuation correction and functional anatomic mapping.

Recall: Z-1788-2013 · Reported July 31, 2013

Enforcement

Recall Number
Z-1788-2013
Event ID
65579
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
GE Healthcare, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 31, 2013
Initiation Date
June 13, 2013
Classification Date
July 23, 2013
Termination Date
June 19, 2014
Address
3000 N Grandview Blvd, N/A, Waukesha, WI, 53188-1615, United States

Description

GE Healthcare Varicam/VG and VG Hawkeye devices Product Usage: The intended use of the Quasar system is to perform general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition typed and optional imaging features designed to enhance image quality in Oncology, Cardiology, Neurology and other clinical diagnostic imaging applications. The scanning modes include planar (Static, Multi-gated, Dynamic, Whole body scanning) and tomographic (SPECT, Gated SPECT, Whole body SPECT, Camera based PET). Acquisition types include single and multi-isotope/multi-peak frame/list mode single-photon and positron imaging. Optional imaging-enhancement features include assortment of collimators, gating by physiological signals, real-time automatic body contouring, and CT-based attenuation correction and functional anatomic mapping.

Reason

GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System.

Code Info

System ID: 2998, 6970, 12865, 21414, 38070, 53263, 55461, 56948, 58230, 58378, 58379, 58483, 58510, 58582, 58622, 58630, 58711, 58785, 58797, 58808, 58810, 58871, 58883, 58919, 58958, 58993, 59012, 61061, 63422, 66515, 75074, 79722, 79966, 84636, 92635, 99691, 0850260033, 0850260034, 0850260035, 0850260038, 0850260040, 0850260041, 0850260042, 0850260046, 0850260048, 0850260054, 0850260057, 0850260062, 0853940029, 0853940068, 0853940069, 0853940077, 0853940078, 0853940080, 0853940082, 0856260014, 0856260019, 0910265011, 0910453006, 082426010009, 082426010010, 082426010011, 082426020002, 082426030009, 082426030010, 082426030015, 082426030016, 082426030018, 082426040016, 082426040020, 082426040021, 082426040024, 082426040025, 082426040026, 082426040029, 082426050001, 082426050002, 082426050003, 082426070014, 082426070015, 082426070016, 082426080006, 082426080007, 082426080009, 082426090013, 082426100017, 082426100019, 082426100020, 082426100023, 082426110002, 082426120003, 082426120004, 082426130012, 082426130013, 082426130014, 082426130015, 082426130016, 082426130019, 082426140005, 082426140007, 082426140008, 082426140009, 082426140011, 082426140012, 082426150003, 082426150006, 082426160012, 082426160014, 082426160016, 082426160017, 082426160019, 082426160020, 082426160022, 082426160023, 082426180003, 082426190008, 082426190010, 082426200002, 082426210002, 082426210003, 082426210010, 082426220006, 082426230008, 082426230010, 082426250002, 082426280001, 082426290004, 082426310005, 082426310007, 082426310008, 08302603017002, 08302603135243, 08302606123002, 08302621006334, 083026031062021, 083026242100654, 083026590040004, 083026590040010, 083026598000594, 083026826753913, 0382E00013, 00023NUC06, 0009D, 00132NUC02, 00216NUC09, 00243NUC11, 00243NUC15, 00243NUC17, 00269NUC07, 00323NUC02, 00357NUC02, 00555NUC01, 00584NUC01, 00584NUC04, 0112B, 0277A, 030010NU02, 030028NU01, 030036NU06, 030082NU10, 030151NU05, 030220NU02, 030355NU02, 030388NU01, 0306A, 0343A, 0360A, 06007NUC16, 080041NU01, 082426040029P, 082426050001P, 082426140005P, 0850EL0011, 0850EL0012, 10433NUC01, 110012NU02, 110965NU05, 130083NU07, 13228MILLEN, 16003NUC03, 160070NU01, 190226NU17, 190282NU04, 203694MIL, 203852VG, 203932VG, 204837VG, 205823NMVG, 205933VH, 206288VG, 206598VG, 207795VH, 208529VG, 209334LNVG1, 209526MILVG, 212562VG, 215893VG, 220166NU01, 223410MILVG, 225763MVG, 228818MILLVG, 248828H, 248828VG, 248937VG, 256235OPNM1, 262785VG, 2894872VG2, 2894872VG3, 301896VG, 304256VG, 304487VG, 309662SJVG, 310231VGH1, 310289VG1, 312864VG, 313299VG, 334794MILVG, 361575VG, 402449SJVG, 404605PVGH, 404712VG, 406782SNM, 407277FVG2, 407896FVG1, 407933KVG, 408947VG, 409291HVG, 410605NM1, 410955VG1, 410955VH, 412359VG, 413781VG1, 413781VG2, 414219VG, 414328VG, 414384VG, 414447VG, 414544HEYE, 414544VG, 414777MIL, 414777VG2, 414961VG, 416284VG, 416284VG2, 416290VG, 416444VG, 416469DVH, 416534VG1, 416534VG2, 416586VG, 416656VG, 416747VARICAM, 416747VG, 416756VG, 418404NU01, 418862VARICAM, 419383VG, 440329VG, 443643VG, 450359VG1, 450359VG2, 480981NM1, 501686VG, 508997VG, 509575PYVGH, 510486VG1, 512753VG, 514495VG, 515263LVG, 516256VG, 518271SVG, 519421VG, 520757NM1, 530529VG, 540186NU01, 540596N3, 540953VG, 541276AVG, 559688NM1, 570791HAWK, 580355VGACQ, 589907MVG, 600062NM01, 601426VG, 602406VG, 602870NM1, 605333HAWKEYE, 608263VG, 608364VG, 609267VG, 610954VG, 613548VG, 613548VG2, 615322NM2, 615791VG, 630801VG, 70050NUC06, 705324VG, 705327VARICAM, 705743VG, 706845VG, 707554VG, 708422MIL, 708783VG, 714774VG, 715361VG, 716694VG, 716859VG, 717657VG, 718759VG, 718780GENIEVG, 718780VG, 73240-1, 781278VG, 781744VG2, 781744VH, 786090NU03, 787782VARICAM-1, 804281VG1, 807343BMILLVG, 807343CMILLVG, 809758VARICAM, 812537VG, 813725VG, 813788VG, 815842SJVG, 819563CVG, 819822VG, 828212VG, 828213VG, 843792NUCH, 845437VG, 847695MILMVG, 847842VG, 860343VGOPC, 860344VG, 863494VG, 864573MBVG1, 904202VG, 905472VG, 905527HAWKEYE, 905683VG, 905684VG, 905684VG2, 905845VG, 905848VG, 905855VG, 905883VG, 905895VG, 912691SCVG, 914734VG, 919684VG1, 920459VG, 920729VG, 920887VG, 949760VG1, 956632MVG, 956687HKI, 970491VG1, 973754VG, 978466VG, 978741VG, 979764VGH1, 989269VG, A004NU10, A2846901, A5103321, A5107736, A5111021, A5114718, A5116033, A5128734, A5129932, A5132115, A5177338, A5177340, A5177344, A5177347, A5188524, A5215912, A5236208, A5306504, A5333009, A5353816, A5632411, AH0101NM02, AH0101NM03, AH3701NM04, AH4307NM04, AH4660NM05, ALK0130, ALK0131, APL0236, B4155407, B5119109, B5131812, B5315325, B5365411, B5483426, C5123638, CC0618NU13, CY1088MN04, D007NU07, DE282013, DE619831, DEL26415, DHG0536, DK1026NM05, DLF0237, E001NU09, EE1020MN01, EEF51528, FE416038, FE816127, FEB52048, FHALTVGH1, FHORLVGH1, G001NU07, GEC21815, GEE34020, H1404NUC06, H2103NUC06, H5174NUC02, HBER01NU02, HC2250NU01, HC2250NU02, HC2250NU03, HC3314NU01, HC4387NU01, HC5036NU01, HJJT01NU02, HOR0250, HR4106NM01, HT2106NU02, HT2106NU05, ID9320, IL1020NM01, IL1028NM03, IL1028NM13, IL1059NM01, IL1061NM01, IL1061NM05, IL1062NM01, IL1066NM03, IL1067NM03, IL1072NM01, JHOCVG10464, M001NU10, M001NU14, M2846904, M2881704, M4024091, M4134620, M4143939, M4144609, M4144845, M4145121, M4160707, M4164912, M4211044, M5400305, M5460632, MAYO1VG4, MAYO1VGH1, MAYO4VG3, MC6946NU02, ME260802, ME504345, ME612431, MPX24714, NM10374, NM73049, NM73050, NM81394, NO1012MN04, NO1015MN03, NO1016MN27, PC0173NU06, PC0396NU03, PC2306NU01, PC4178NU01, PER02544, PER10924, PER10925, PL1003NM01, PL1008NM01, PL1143NM01, PMR12503, PPR12206, RADNETNM1724, RU1100NM01, RU1457NM01, RU1591NM01, SK1071NM01, SK1082NM01, U001NU12, UC1129NU01, UC1769NU07, UC4092NU01, UC4109NU01, WH231NM1, YE645816, YE645817, YN0438, YN0440, YN0444, YN0448, YN0450, YN0454, YN0455, YN0468, YN0475, YN0477, YN0479, YN0482, YN0486, YN0493, YN0498, YN0502, YN0506, YN0542, YN0543, YN0544, YN0638, YN0639, YN0640, YN0641, YN0642, YN0643, YN0844, YN0845, YN0846, YN0847, YN0848, YN0849, YN0850, YN0851, YN0852, YN0853, YN0854, YN0907, YN0909, YN0910, YN0911, YN0912, YN0913, YN0914, YN0915, YN0917, YN0918, YN0919, YN0920, YN0921, YN0922, YN0923, YN0924, YN0966, YN0979, YN0980, YN0981, YN0982, YN0983, YN0984, YN0985, YN0986, YN0987, YN0988, YN1000, YN1009, ZA1807NM08, ZWL0431,

Distribution

Worldwide Distribution - US Nationwide including Guam, Puerto Rico and the countries of : Afghanistan, Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Korea (Republic of), Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Malta, Martinique, Mexico, Morocco, Namibia, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Kingdom, Uruguay, Venezuela, Vietnam.

Quantity

571