FDA Enforcement
Class II
Terminated
Centricity PACS Web Diagnostic (WebDX) by GE Healthcare Dynamic Imaging Solutions is a device that receives medical images (including mammograms) and data from various imaging sources within the system or across computer networks at distributed locations.
Recall: Z-1782-2017
·
Reported April 19, 2017
Enforcement
- Recall Number
- Z-1782-2017
- Event ID
- 64545
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Ge Medical Systems Information T
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 19, 2017
- Initiation Date
- February 18, 2013
- Classification Date
- April 11, 2017
- Termination Date
- July 31, 2017
- Address
- 540 W Northwest Hwy, N/A, Barrington, IL, 60010-3051, United States
Description
Centricity PACS Web Diagnostic (WebDX) by GE Healthcare Dynamic Imaging Solutions is a device that receives medical images (including mammograms) and data from various imaging sources within the system or across computer networks at distributed locations.
Reason
It was reported from a complaint that study category labels of "Prior" and "Baseline" and hanging position appeared incorrectly swapped when the hanging protocol for three comparison studies is manually applied from the patient folder.
Code Info
Versions 1.0, 1.1, 2.0, 2.1
Distribution
Nationwide, Worldwide
Quantity
118