GE Centricity PACS-IW versions 3.6.0 through 3.7.3.9 SP1, SP2, SP3 and 3.7.3 SPa10, Model Numbers: a) 2052831-00x b) 2050578-0xx c) 2090228-002 d) 2052782-00x
Enforcement
- Recall Number
- Z-1781-2019
- Event ID
- 83002
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- GE Healthcare, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 19, 2019
- Initiation Date
- May 22, 2019
- Classification Date
- June 13, 2019
- Termination Date
- September 20, 2024
- Address
- 3000 N Grandview Blvd, N/A, Waukesha, WI, 53188-1615, United States
Description
GE Centricity PACS-IW versions 3.6.0 through 3.7.3.9 SP1, SP2, SP3 and 3.7.3 SPa10, Model Numbers: a) 2052831-00x b) 2050578-0xx c) 2090228-002 d) 2052782-00x
There is a potential that one or more image series (i.e., all images within an image set) may be missing from an exam without a warning displayed in the viewer.
a) 2052831-00x, Lot or Serial Numbers: 100839-1-Centricity PACS-IW-00558122 100893-1-Centricity PACS-IW-00558171 4320-1-Centricity PACS-IW-00558202 4474-1-Centricity PACS-IW-00558169 2455-1-Centricity PACS-IW-00558139 100863-1-Centricity PACS-IW-00558143 100794-1-Centricity PACS-IW-00558193 4346-1-Centricity PACS-IW-00558136 100909-1-Centricity PACS-IW-00558198 4355-1-Centricity PACS-IW-00558204 100809-1-Centricity PACS-IW-00558124 4300-1-Centricity PACS-IW-00558207 4484-1-Centricity PACS-IW-00558213 4479-1-Centricity PACS-IW-00558214 13240-2-Centricity PACS-IW-01225875 4270-1-Centricity PACS-IW-00558129 11059-2-Centricity PACS-IW-00558101 101598-1-Centricity PACS-IW-00667133 4273-1-Centricity PACS-IW-01225873 4357-1-Centricity PACS-IW-00558444 4293-1-Centricity PACS-IW-00558131 100835-1-Centricity PACS-IW-00558568 100900-1-Centricity PACS-IW-00558170 4344-1-Centricity PACS-IW-00558196 3836-1-Centricity PACS-IW-00558179 4428-1-Centricity PACS-IW-00558205 100609-1-Centricity PACS-IW-00558130 4297-1-Centricity PACS-IW-00558134 100975-1-Centricity PACS-IW-00558137 100667-1-Centricity PACS-IW-00558163 4316-1-Centricity PACS-IW-00558210 4251-1-Centricity PACS-IW-00558120 4254-1-Centricity PACS-IW-00558211 100981-1-Centricity PACS-IW-00558132 100871-1-Centricity PACS-IW-00558146 2553-1-Centricity PACS-IW-00558206 4283-1-Centricity PACS-IW-00558187 4469-1-Centricity PACS-IW-00558128 100020-1-Centricity PACS-IW-00558147 4443-1-Centricity PACS-IW-00558167 4317-1-Centricity PACS-IW-00558208 100775-1-Centricity PACS-IW-00558209 4323-1-Centricity PACS-IW-00558164 100825-1-Centricity PACS-IW-00558172 100830-1-Centricity PACS-IW-00558166 4304-1-Centricity PACS-IW-00558145 4721-1-Centricity PACS-IW-00558138 4104-1-Centricity PACS-IW-00558133 4292-1-Centricity PACS-IW-00558203 100980-1-Centricity PACS-IW-00558140 4343-1-Centricity PACS-IW-00558115 4298-1-Centricity PACS-IW-00558116 4252-1-Centricity PACS-IW-00558119 4287-1-Centricity PACS-IW-00558117 4291-1-Centricity PACS-IW-00558194 100978-1-Centricity PACS-IW-00558141 4448-1-Centricity PACS-IW-00558118 100979-1-Centricity PACS-IW-00558142 100865-1-Centricity PACS-IW-01225871 100807-1-Centricity PACS-IW-00558121 100807-1-Centricity PACS-IW-00558212 4424-1-Centricity PACS-IW-00558429 339142-Centricity PACS-IW-01431503 13759-Centricity PACS-IW-01431508 320090-Centricity PACS-IW-01444604 320092-Centricity PACS-IW-01444618 297958-Centricity PACS-IW-01444903 329997-Centricity PACS-IW-01444898 AiKang-Centricity PACS-IW-01824737 SiteID-Centricity PACS-IW-01572187 SiteID-Centricity PACS-IW-01797297 339734-Centricity PACS-IW-01444878 313825-Centricity PACS-IW-01444348 332562-Centricity PACS-IW-01444345 338295-Centricity PACS-IW-01443383 SiteID-Centricity PACS-IW-01444354 SiteID-Centricity PACS-IW-02003770 276233-Centricity PACS-IW-01444355 331973-Centricity PACS-IW-01434505 279961-Centricity PACS-IW-01444358 SiteID-Centricity PACS-IW-00161903 SiteID-Centricity PACS-IW-00161906 186588-Centricity PACS-IW-01461366 SiteID-Centricity PACS-IW-02095351 RU9038-Centricity PACS-IW-01273313 RU1107-Centricity PACS-IW-01273202 MPX431-Centricity PACS-IW-00044783 b) 2050578-0xx Lot or Serial Numbers: 100664-1-Centricity PACS-IW-00558437 100639-1-Centricity PACS-IW-00558450 4261-1-Centricity PACS-IW-00558480 101587-1-Centricity PACS-IW-00558369 4614-1-Centricity PACS-IW-00558393 4376-1-Centricity PACS-IW-00558399 4436-1-Centricity PACS-IW-00558417 4350-13-Centricity PACS-IW-00558296 4680-1-Centricity PACS-IW-00558360 100666-1-Centricity PACS-IW-00558465 101171-1-Centricity PACS-IW-00558310 101792-1-Centricity PACS-IW-00558370 4264-1-Centricity PACS-IW-00558254 4286-1-Centricity PACS-IW-00558361 4284-1-Centricity PACS-IW-00558453 4605-1-Centricity PACS-IW-00558350 4446-2-Centricity PACS-IW-00558106 4280-1-Centricity PACS-IW-00558240 12830-1-Centricity PACS-IW-00558312 100580-1-Centricity PACS-IW-00558339 100858-1-Centricity PACS-IW-00558426 100674-1-Centricity PACS-IW-00558315 102206-1-Centricity PACS-IW-00558447 4684-1-Centricity PACS-IW-00558245 4256-1-Centricity PACS-IW-00558250 4707-1-Centricity PACS-IW-00558434 100783-1-Centricity PACS-IW-00558255 4263-1-Centricity PACS-IW-00558279 100970-1-Centricity PACS-IW-00558303 4689-1-Centricity PACS-IW-00558242 100767-1-Centricity PACS-IW-00558377 100766-1-Centricity PACS-IW-00558355 100764-1-Centricity PACS-IW-00558452 100765-1-Centricity PACS-IW-00558331 4350-1-Centricity PACS-IW-00558301 4350-1-Centricity PACS-IW-00558307 4350-1-Centricity PACS-IW-00558318 4350-1-Centricity PACS-IW-00558432 4109-1-Centricity PACS-IW-00558389 4485-1-Centricity PACS-IW-00558479 100485-1-Centricity PACS-IW-00558258 100759-1-Centricity PACS-IW-00558268 4687-1-Centricity PACS-IW-00558334 1005-1-Centricity PACS-IW-00558282 1005-1-Centricity PACS-IW-00558290 100836-1-Centricity PACS-IW-00558256 4311-1-Centricity PACS-IW-00558321 100757-1-Centricity PACS-IW-00558261 4679-1-Centricity PACS-IW-00558412 100718-1-Centricity PACS-IW-00558424 13332-1-Centricity PACS-IW-00558390 4294-1-Centricity PACS-IW-00558379 13360-1-Centricity PACS-IW-00558263 12908-1-Centricity PACS-IW-00558302 4359-1-Centricity PACS-IW-00558410 13039-1-Centricity PACS-IW-00558385 100166-1-Centricity PACS-IW-00558448 9757-1-Centricity PACS-IW-00558247 1412-1-Centricity PACS-IW-00558228 4274-1-Centricity PACS-IW-00558436 100156-1-Centricity PACS-IW-00558243 100156-1-Centricity PACS-IW-00558475 4723-1-Centricity PACS-IW-00558367 4539-1-Centricity PACS-IW-00558378 101042-1-Centricity PACS-IW-00558386 3189-1-Centricity PACS-IW-00558314 4495-1-Centricity PACS-IW-00558476 4734-1-Centricity PACS-IW-00558375 100701-1-Centricity PACS-IW-00558322 100722-1-Centricity PACS-IW-00558407 2035-1-Centricity PACS-IW-00558411 101742-1-Centricity PACS-IW-00558470 104648-2-Centricity PACS-IW-00558406 4670-1-Centricity PACS-IW-00558329 100855-1-Centricity PACS-IW-00558298 4385-1-Centricity PACS-IW-00558336 4260-1-Centricity PACS-IW-00558244 100065-1-Centricity PACS-IW-00558317 4295-1-Centricity PACS-IW-00558341 13209-1-Centricity PACS-IW-00558357 4676-1-Centricity PACS-IW-00558449 13295-1-Centricity PACS-IW-00558346 9791-1-Centricity PACS-IW-00558418 4273-1-Centricity PACS-IW-01225873 4269-1-Centricity PACS-IW-00558252 100589-1-Centricity PACS-IW-00558384 13337-1-Centricity PACS-IW-00558401 13337-1-Centricity PACS-IW-00558478 100876-1-Centricity PACS-IW-00558440 11764-1-Centricity PACS-IW-00558388 101689-1-Centricity PACS-IW-00558326 9764-1-Centricity PACS-IW-00558394 100835-1-Centricity PACS-IW-00558287 100831-1-Centricity PACS-IW-00558232 4744-1-Centricity PACS-IW-00558373 100772-1-Centricity PACS-IW-00558269 4281-1-Centricity PACS-IW-00558354 12960-1-Centricity PACS-IW-00558289 9978-3-Centricity PACS-IW-00558108 1402-1-Centricity PACS-IW-00558291 4321-1-Centricity PACS-IW-00558371 4381-1-Centricity PACS-IW-00558423 1451-1-Centricity PACS-IW-00558294 13175-1-Centricity PACS-IW-00558383 10467-1-Centricity PACS-IW-00558405 10911-1-Centricity PACS-IW-00558471 100821-1-Centricity PACS-IW-00558325 4425-1-Centricity PACS-IW-00558472 4268-1-Centricity PACS-IW-00558416 9844-1-Centricity PACS-IW-00558332 9767-1-Centricity PACS-IW-00558265 4305-1-Centricity PACS-IW-00558337 4367-1-Centricity PACS-IW-00558415 4128-1-Centricity PACS-IW-00558234 101098-1-Centricity PACS-IW-00558387 4322-1-Centricity PACS-IW-00558285 4352-1-Centricity PACS-IW-00558316 4374-1-Centricity PACS-IW-00558422 100827-1-Centricity PACS-IW-00558233 4360-1-Centricity PACS-IW-00558237 4335-1-Centricity PACS-IW-00558267 4349-1-Centricity PACS-IW-00558284 4211-1-Centricity PACS-IW-00558277 3545-1-Centricity PACS-IW-00558239 10527-3-Centricity PACS-IW-00558344 4348-1-Centricity PACS-IW-00558458 4362-1-Centricity PACS-IW-00558395 4576-1-Centricity PACS-IW-00558306 10916-1-Centricity PACS-IW-00558299 4288-1-Centricity PACS-IW-00558235 4266-1-Centricity PACS-IW-00558304 100511-1-Centricity PACS-IW-00558549 1565-1-Centricity PACS-IW-00603058 4277-1-Centricity PACS-IW-00558474 4553-1-Centricity PACS-IW-00558327 101216-1-Centricity PACS-IW-00558356 4531-1-Centricity PACS-IW-00558376 101240-1-Centricity PACS-IW-00558425 100021-1-Centricity PACS-IW-00558260 4660-1-Centricity PACS-IW-00558288 2481-1-Centricity PACS-IW-00558381 4447-1-Centricity PACS-IW-00558398 4078-1-Centricity PACS-IW-00558462 100806-1-Centricity PACS-IW-00558441 4361-1-Centricity PACS-IW-00558343 13164-1-Centricity PACS-IW-00558439 4504-1-Centricity PACS-IW-00558438 4413-1-Centricity PACS-IW-00558433 4416-1-Centricity PACS-IW-00558292 4416-1-Centricity PACS-IW-00558293 4417-1-Centricity PACS-IW-00558364 4419-1-Centricity PACS-IW-00558483 4420-1-Centricity PACS-IW-00558404 100378-1-Centricity PACS-IW-00558238 4508-1-Centricity PACS-IW-00558271 100603-1-Centricity PACS-IW-00558266 10462-1-Centricity PACS-IW-00558272 12953-1-Centricity PACS-IW-00558443 10483-1-Centricity PACS-IW-00558098 1331-1-Centricity PACS-IW-00558451 12949-1-Centricity PACS-IW-00558442 4296-1-Centricity PACS-IW-00558477 100088-1-Centricity PACS-IW-00558320 4351-1-Centricity PACS-IW-00558485 1584-1-Centricity PACS-IW-00558297 1173-1-Centricity PACS-IW-00558473 4450-1-Centricity PACS-IW-00558319 100972-1-Centricity PACS-IW-00558278 4276-1-Centricity PACS-IW-00558274 101647-1-Centricity PACS-IW-00558515 11711-2-Centricity PACS-IW-00558262 4542-1-Centricity PACS-IW-00558430 13265-1-Centricity PACS-IW-00558482 12974-1-Centricity PACS-IW-00558300 100770-1-Centricity PACS-IW-00558097 100769-1-Centricity PACS-IW-00558427 4393-1-Centricity PACS-IW-00558328 13372-1-Centricity PACS-IW-00558309 13197-1-Centricity PACS-IW-00558251 11042-1-Centricity PACS-IW-00558368 100468-1-Centricity PACS-IW-00558330 4575-1-Centricity PACS-IW-00558342 12931-1-Centricity PACS-IW-00558419 4482-1-Centricity PACS-IW-00558281 4568-1-Centricity PACS-IW-00558280 3571-1-Centricity PACS-IW-00558468 100645-1-Centricity PACS-IW-00558408 100771-1-Centricity PACS-IW-00558249 4632-1-Centricity PACS-IW-00558311 4681-1-Centricity PACS-IW-00558431 4573-1-Centricity PACS-IW-00558420 100729-1-Centricity PACS-IW-00558455 4456-1-Centricity PACS-IW-00558295 4506-1-Centricity PACS-IW-00558345 BA4080-Centricity PACS-IW-00038769 324462-Centricity PACS-IW-01431554 293035-Centricity PACS-IW-01401532 280801-Centricity PACS-IW-01431490 335279-Centricity PACS-IW-01431638 306930-Centricity PACS-IW-01401530 296619-Centricity PACS-IW-01401506 330917-Centricity PACS-IW-01401529 SiteID-Centricity PACS-IW-01422991 313741-Centricity PACS-IW-01422998 238110-Centricity PACS-IW-01431521 246838-Centricity PACS-IW-01421153 134265-Centricity PACS-IW-01431539 307319-Centricity PACS-IW-01444609 230048-Centricity PACS-IW-01442018 SiteID-Centricity PACS-IW-01444879 209545-Centricity PACS-IW-01460552 EG1856-Centricity PACS-IW-00118541 EG1009-Centricity PACS-IW-00118535 EG1871-Centricity PACS-IW-00118532 EG1905-Centricity PACS-IW-00117725 EG1771-Centricity PACS-IW-00118550 EG1858-Centricity PACS-IW-00118544 EG1808-Centricity PACS-IW-00118538 EG1857-Centricity PACS-IW-00118547 M41678-Centricity PACS-IW-00043697 M41741-Centricity PACS-IW-00043705 M42172-Centricity PACS-IW-00165182 683198-Centricity PACS-IW-00000826 600791-Centricity PACS-IW-00000776 UHAM02-Centricity PACS-IW-00001927 RMUE01-Centricity PACS-IW-00001801 PC0125-Centricity PACS-IW-00001683 HU1025-Centricity PACS-IW-00038776 HU1469-Centricity PACS-IW-00038785 HU1038-Centricity PACS-IW-00038778 HU1093-Centricity PACS-IW-00038784 HU1397-Centricity PACS-IW-00038782 HU1185-Centricity PACS-IW-00038781 HU1433-Centricity PACS-IW-00038783 HU1008-Centricity PACS-IW-00038775 HU1036-Centricity PACS-IW-00038777 HU1041-Centricity PACS-IW-00038780 HU1039-Centricity PACS-IW-00038779 ApolloBSR-Centricity PACS-IW-01819452 StThomas-Centricity PACS-IW-01819447 HOLY-FAMILY-THODU-Centricity PACS-IW-01821493 SODANI-Centricity PACS-IW-01819445 WOCKHARDT-Centricity PACS-IW-01819449 DAKE-Centricity PACS-IW-01819448 SIMS-Centricity PACS-IW-01819453 MARTHANDAM-Centricity PACS-IW-01819446 CTSI-Centricity PACS-IW-01819444 CONTINENTAL-Centricity PACS-IW-01819443 197240-Centricity PACS-IW-01442043 KW1112-Centricity PACS-IW-00168151 LB1331-Centricity PACS-IW-00116959 REGENCY-Centricity PACS-IW-01286180 BEVERLY-Centricity PACS-IW-01288659 MAHKOTA-Centricity PACS-IW-01286179 UMUM-Centricity PACS-IW-01288666 5791-Centricity PACS-IW-01442181 324271-Centricity PACS-IW-01970711 AMS39-Centricity PACS-IW-00044104 PL1731-Centricity PACS-IW-00038835 PL1517-Centricity PACS-IW-00038837 PL1515-Centricity PACS-IW-00038839 PL1394-Centricity PACS-IW-00038841 PL1639-Centricity PACS-IW-00038843 PL1426-Centricity PACS-IW-00038845 PL1491-Centricity PACS-IW-00038847 PL2619-Centricity PACS-IW-00038814 PL1631-Centricity PACS-IW-00038849 PL1684-Centricity PACS-IW-00038851 PL1942-Centricity PACS-IW-00038853 PL1825-Centricity PACS-IW-00038819 PL1505-Centricity PACS-IW-00038855 PL3249-Centricity PACS-IW-00038865 PL3249-Centricity PACS-IW-00038868 PL3249-Centricity PACS-IW-00038870 PL3249-Centricity PACS-IW-00038871 PL1530-Centricity PACS-IW-00038822 PL2762-Centricity PACS-IW-00038828 PL1544-Centricity PACS-IW-00038859 PL1644-Centricity PACS-IW-00038861 PL1650-Centricity PACS-IW-00038857 PL1508-Centricity PACS-IW-00038863 QA1015-Centricity PACS-IW-00115165 RU3382-Centricity PACS-IW-01273247 RU3680-Centricity PACS-IW-01273260 RU3380-Centricity PACS-IW-01273246 SA2187-Centricity PACS-IW-00065875 SA2294-Centricity PACS-IW-00065843 SA1198-Centricity PACS-IW-00065868 SA1175-Centricity PACS-IW-00038187 SiteID-Centricity PACS-IW-02266272 MPX352-Centricity PACS-IW-01973993 DP7603-Centricity PACS-IW-00045307 UP506-Centricity PACS-IW-00044759 529010-Centricity PACS-IW-00000342 ONCOLOGY-Centricity PACS-IW-01288658 115-Centricity PACS-IW-01288657 c) 2090228-002 Lot or Serial Numbers: 3640-1-Centricity PACS-IW-00558511 4345-1-Centricity PACS-IW-00558548 4430-1-Centricity PACS-IW-00558500 4378-1-Centricity PACS-IW-00558498 100663-1-Centricity PACS-IW-00558514 100674-1-Centricity PACS-IW-00558486 9754-1-Centricity PACS-IW-00558490 4429-1-Centricity PACS-IW-00558499 4328-1-Centricity PACS-IW-00558518 4552-1-Centricity PACS-IW-00558510 4387-1-Centricity PACS-IW-00558493 100702-1-Centricity PACS-IW-00558539 100717-1-Centricity PACS-IW-00558517 101625-1-Centricity PACS-IW-00558531 4593-1-Centricity PACS-IW-00558488 100009-1-Centricity PACS-IW-00558560 4354-1-Centricity PACS-IW-00558519 100838-1-Centricity PACS-IW-00558532 4551-1-Centricity PACS-IW-00558508 100199-1-Centricity PACS-IW-00558503 12968-1-Centricity PACS-IW-00558506 13102-1-Centricity PACS-IW-00558512 4379-1-Centricity PACS-IW-00558496 4366-1-Centricity PACS-IW-00558542 4694-1-Centricity PACS-IW-00558556 1635-1-Centricity PACS-IW-00558557 100881-1-Centricity PACS-IW-00558546 4299-2-Centricity PACS-IW-00558545 10884-2-Centricity PACS-IW-00558100 101883-1-Centricity PACS-IW-00558487 100852-1-Centricity PACS-IW-00558494 4368-1-Centricity PACS-IW-00558513 4668-1-Centricity PACS-IW-00558529 100811-1-Centricity PACS-IW-00558535 4497-1-Centricity PACS-IW-00558497 4682-1-Centricity PACS-IW-00558270 100458-1-Centricity PACS-IW-00558538 4358-1-Centricity PACS-IW-00558564 13095-1-Centricity PACS-IW-00558534 13064-2-Centricity PACS-IW-00558102 4372-1-Centricity PACS-IW-00558523 4326-4-Centricity PACS-IW-00558105 101932-1-Centricity PACS-IW-00558495 4301-1-Centricity PACS-IW-00558525 10750-1-Centricity PACS-IW-00558536 100761-1-Centricity PACS-IW-00558527 10749-1-Centricity PACS-IW-00558509 2387-1-Centricity PACS-IW-00558489 10250-1-Centricity PACS-IW-00558504 100802-1-Centricity PACS-IW-00558540 101193-1-Centricity PACS-IW-00558551 1475-1-Centricity PACS-IW-00558558 12821-1-Centricity PACS-IW-00558526 4661-1-Centricity PACS-IW-00558528 2481-1-Centricity PACS-IW-00558524 3943-1-Centricity PACS-IW-00558501 4299-2-Centricity PACS-IW-02180909 13134-1-Centricity PACS-IW-00558492 4157-1-Centricity PACS-IW-00558516 4383-1-Centricity PACS-IW-00558533 13233-1-Centricity PACS-IW-00558554 4389-1-Centricity PACS-IW-00558563 100582-1-Centricity PACS-IW-00558530 4179-1-Centricity PACS-IW-00558543 DZ1004-Centricity PACS-IW-00116835 BH1038-Centricity PACS-IW-00162931 80083-Centricity PACS-IW-00043981 SiteID-Centricity PACS-IW-01431505 332921-Centricity PACS-IW-01442042 313689-Centricity PACS-IW-01431524 SiteID-Centricity PACS-IW-01401523 000341097-Centricity PACS-IW-01431636 191244-Centricity PACS-IW-01422989 298983-Centricity PACS-IW-01401548 293834-Centricity PACS-IW-01422975 302144-Centricity PACS-IW-01421157 32125-Centricity PACS-IW-01444621 13628-Centricity PACS-IW-01422992 213537-Centricity PACS-IW-01431510 000282202-Centricity PACS-IW-01422973 319674-Centricity PACS-IW-01431535 369946-Centricity PACS-IW-01264977 218314-Centricity PACS-IW-01434509 248553-Centricity PACS-IW-01442041 EG1017-Centricity PACS-IW-00117723 X2575258-Centricity PACS-IW-00043952 M43994-Centricity PACS-IW-00043773 M60364-Centricity PACS-IW-00043899 X1413896-Centricity PACS-IW-00043945 TESTSITE-Centricity PACS-IW-01986381 714235-Centricity PACS-IW-00000842 972439-Centricity PACS-IW-00000859 363691-Centricity PACS-IW-00000520 424881-Centricity PACS-IW-00000556 A2513980-Centricity PACS-IW-00040355 A56306-Centricity PACS-IW-00040439 PUSANPAIK-Centricity PACS-IW-01819440 CHILGOK-Centricity PACS-IW-01819441 KANGBUK-Centricity PACS-IW-01819442 1971565-YUHSSC-RAD-Centricity PACS-IW-01819454 KW1003-Centricity PACS-IW-01798078 KW1019-Centricity PACS-IW-00115173 KW1018-Centricity PACS-IW-01422962 KW1001-Centricity PACS-IW-00164288 KW1096-Centricity PACS-IW-00166324 SSH-Centricity PACS-IW-01288660 NG1043-IW w UW Base SW PVP-00158638 MYHEALTH-Centricity PACS-IW-01819534 StLukes-Centricity PACS-IW-01286216 PL2059-Centricity PACS-IW-00038789 PL2997-Centricity PACS-IW-00038793 PL1803-Centricity PACS-IW-00038812 PL2823-Centricity PACS-IW-00038813 PL2342-Centricity PACS-IW-00038815 PL1666-Centricity PACS-IW-00038816 PL2257-Centricity PACS-IW-00038817 PL1961-Centricity PACS-IW-00038818 PL3092-Centricity PACS-IW-00038830 PL1006-Centricity PACS-IW-00038821 PL1822-Centricity PACS-IW-00038825 PL3362-Centricity PACS-IW-00038833 PL1951-Centricity PACS-IW-00038827 RU1016-Centricity PACS-IW-01273193 RU1241-Centricity PACS-IW-01273216 RU1493-Centricity PACS-IW-01273220 RU2967-Centricity PACS-IW-01273244 RU5302-Centricity PACS-IW-01273274 RU2420-Centricity PACS-IW-01273234 RU3910-Centricity PACS-IW-01273270 RU3262-Centricity PACS-IW-01273245 RU6248-Centricity PACS-IW-01273279 RU1648-Centricity PACS-IW-01273222 RU8071-Centricity PACS-IW-01273295 RU6340-Centricity PACS-IW-01273280 RU6340-Centricity PACS-IW-01273282 RU7576-Centricity PACS-IW-01273287 RU3880-Centricity PACS-IW-01273266 RU3894-Centricity PACS-IW-01273269 RU3564-Centricity PACS-IW-01273252 RU8162-Centricity PACS-IW-01273299 RU2527-Centricity PACS-IW-01273235 RU5310-Centricity PACS-IW-01273276 RU3959-Centricity PACS-IW-01273273 RU2878-Centricity PACS-IW-01273243 RU8370-Centricity PACS-IW-01273306 RU3572-Centricity PACS-IW-01273255 600160-Centricity PACS-IW-01273191 RU3625-Centricity PACS-IW-01273258 RU5499-Centricity PACS-IW-01273278 RU1469-Centricity PACS-IW-01273219 RU1199-Centricity PACS-IW-01273215 RU1717-Centricity PACS-IW-01273225 RU9000-Centricity PACS-IW-01273312 RU8041-Centricity PACS-IW-01273294 RU1857-Centricity PACS-IW-01273229 RU8219-Centricity PACS-IW-01273302 RU1279-Centricity PACS-IW-01273217 RU1095-Centricity PACS-IW-01273200 RU7000-Centricity PACS-IW-01273281 RU8321-Centricity PACS-IW-01273303 RU7563-Centricity PACS-IW-01273285 RU1130-Centricity PACS-IW-01273210 RU2371-Centricity PACS-IW-01273233 RU7426-Centricity PACS-IW-01273284 RU2167-Centricity PACS-IW-01273231 RU3890-Centricity PACS-IW-01273268 RU3828-Centricity PACS-IW-01273262 RU1669-Centricity PACS-IW-01273224 RU5308-Centricity PACS-IW-01273275 RU8953-Centricity PACS-IW-01273311 RU2367-Centricity PACS-IW-01273232 RU1026-Centricity PACS-IW-01273195 RU3476-Centricity PACS-IW-01273251 RU1050-Centricity PACS-IW-01273198 RU1050-Centricity PACS-IW-00167296 RU7564-Centricity PACS-IW-01273286 RU8036-Centricity PACS-IW-01273293 RU3570-Centricity PACS-IW-01273254 RU3610-Centricity PACS-IW-01273257 RU3794-Centricity PACS-IW-01273261 RU8192-Centricity PACS-IW-01273300 RU8192-Centricity PACS-IW-01273301 RU3832-Centricity PACS-IW-01273263 RU1144-Centricity PACS-IW-01273211 600100-Centricity PACS-IW-01273188 RU3878-Centricity PACS-IW-01273265 RU1028-Centricity PACS-IW-01273197 RU3925-Centricity PACS-IW-01273272 RU1055-Centricity PACS-IW-01273199 RU3919-Centricity PACS-IW-01273271 RU1022-Centricity PACS-IW-01273194 RU5442-Centricity PACS-IW-01273277 RU3883-Centricity PACS-IW-01273267 RU3467-Centricity PACS-IW-01273250 RU3663-Centricity PACS-IW-01273259 RU7905-Centricity PACS-IW-01273291 RU3593-Centricity PACS-IW-01273256 RU1298-Centricity PACS-IW-01273218 RU8151-Centricity PACS-IW-01273296 RU7838-Centricity PACS-IW-01273289 RU7838-Centricity PACS-IW-01273290 RU1152-Centricity PACS-IW-01273212 RU8607-Centricity PACS-IW-01273308 RU3569-Centricity PACS-IW-01273253 SA2128-Centricity PACS-IW-00065800 SA1126-Centricity PACS-IW-00065887 SA2235-Centricity PACS-IW-00038164 SA2221-Centricity PACS-IW-00065837 ZA2408-Centricity PACS-IW-00116590 ZA2339-Centricity PACS-IW-00116589 ZA2356-Centricity PACS-IW-00116576 ZA2355-Centricity PACS-IW-00116575 ZA2338-Centricity PACS-IW-00116588 ZA2331-Centricity PACS-IW-00116443 ZA2335-Centricity PACS-IW-00116447 ZA2330-Centricity PACS-IW-00116442 ZA2334-Centricity PACS-IW-00116446 ZA2409-Centricity PACS-IW-00116591 ZA2410-Centricity PACS-IW-00116592 ZA2336-Centricity PACS-IW-00116448 ZA2470-Centricity PACS-IW-00116572 ZA2352-Centricity PACS-IW-00116570 ZA2351-Centricity PACS-IW-00116569 ZA2474-Centricity PACS-IW-00116763 ZA2471-Centricity PACS-IW-00116573 YP5673-Centricity PACS-IW-00045220 941804-Centricity PACS-IW-00000474 d) 2052782-00x Lot or Serial Numbers: 100884-1-Centricity PACS-IW-00558222 100832-1-Centricity PACS-IW-00558224 100828-1-Centricity PACS-IW-00558231 100954-1-Centricity PACS-IW-00558225 100826-1-Centricity PACS-IW-00558229 4595-1-Centricity PACS-IW-00558226 4290-1-Centricity PACS-IW-00558217 4364-1-Centricity PACS-IW-00558219 4524-1-Centricity PACS-IW-00558218 4536-1-Centricity PACS-IW-00558216
Worldwide
620 units