FDA Enforcement Class II Ongoing

Puritan Bennett 560 Ventilator, PB560, CFN 4096600

Recall: Z-1775-2022 · Reported September 28, 2022

Enforcement

Recall Number
Z-1775-2022
Event ID
90766
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Covidien Llc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 28, 2022
Initiation Date
August 19, 2022
Classification Date
September 22, 2022
Address
15 Hampshire St, N/A, Mansfield, MA, 02048-1113, United States

Description

Puritan Bennett 560 Ventilator, PB560, CFN 4096600

Reason

Medtronic has confirmed six reports from customers outside the US where the ventilator became inoperable due to loss of gas supply caused by manufacturing error of specific turbine components.

Code Info

UDI-DI: 10884521196377 U.S. Serial Numbers: 40966N7229 40966N7231 40966N7334 40966N7335 40966N7340

Distribution

Domestic distribution to one customer in Washington, D.C. Foreign distribution worldwide.

Quantity

5 US