FDA Enforcement Class II Terminated

Hemodynamic monitoring system (HMS), comprised of a Patient Interface Module (Part Number 58401-000 and a User Interface Module (Part Number 58403-000). Intended for patients for whom the monitoring of continuous cardiac output is indicated for diagnostic and prognostic evaluation by a clinician in the hospital.

Recall: Z-1775-2018 · Reported May 16, 2018

Enforcement

Recall Number
Z-1775-2018
Event ID
79814
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
ICU Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 16, 2018
Initiation Date
January 5, 2018
Classification Date
May 9, 2018
Termination Date
March 18, 2019
Address
951 Calle Amanecer, N/A, San Clemente, CA, 92673-6212, United States

Description

Hemodynamic monitoring system (HMS), comprised of a Patient Interface Module (Part Number 58401-000 and a User Interface Module (Part Number 58403-000). Intended for patients for whom the monitoring of continuous cardiac output is indicated for diagnostic and prognostic evaluation by a clinician in the hospital.

Reason

When continuous cardiac output monitoring with a pulmonary artery catheter is placed in the Pause mode, the catheter thermal coil may continue while in pause mode.

Code Info

a. Part Number 58401-000 - UDI (Serial No.): (01)00840619079499(11)170615(21)1647100134 (1647100134), (01)00840619079499(11)170912(21)1647100137 (1647100137), (01)00840619079499(11)170912(21)1731100224 (1731100224), (01)00840619079499(11)170914(21)17311002301 (731100230), (01)00840619079499(11)170914(21)1731100228 (1731100228), (01)00840619079499(11)170605(21)1647100140 (1647100140), (01)00840619079499(11)170822(21)1647100135 (1647100135), (01)00840619079499(11)170825(21)1726100215 (1726100215), (01)00840619079499(11)170721(21)1726100214 (1726100214), (01)00840619079499(11)170801(21)1718100164 (1718100164), (01)00840619079499(11)170615(21)1647100139 (1647100139), (01)00840619079499(11)170822(21)1647100146 (1647100146), (01)00840619079499(11)170822(21)1647100136 (1647100136), (01)00840619079499(11)170716(21)1711100147 (1711100147), (01)00840619079499(11)170615(21)1647100143 (1647100143), (01)00840619079499(11)170824(21)1726100204 (1726100204), (01)00840619079499(11)170712(21)1718100166 (1718100166), (01)00840619079499(11)170824(21)1726100205 (1726100205), (01)00840619079499(11)170922(21)1731100241 (1731100241), (01)00840619079499(11)170915(21)1731100231 (1731100231), (01)00840619079499(11)170925(21)1731100243 (1731100243), (01)00840619079499(11)170615(21)1647100144 (1647100144), (01)00840619079499(11)170831(21)1726100209 (1726100209), (01)00840619079499(11)170830(21)1726100218 (1726100218), (01)00840619079499(11)170809(21)1726100216 (1726100216), (01)00840619079499(11)170906(21)1726100210 (1726100210), (01)00840619079499(11)170911(21)1731100220 (1731100220), (01)00840619079499(11)170913(21)1731100226 (1731100226), (01)00840619079499(11)170716(21)1711100154 (1711100154), (01)00840619079499(11)170716(21)1711100155 (1711100155), (01)00840619079499(11)170715(21)1711100151 (1711100151), (01)00840619079499(11)170715(21)1711100150 (1711100150), (01)00840619079499(11)170727(21)1718100171 (1718100171), (01)00840619079499(11)170801(21)1718100173 (1718100173), (01)00840619079499(11)170731(21)1718100159 (1718100159), (01)00840619079499(11)170801(21)1726100213 (1726100213), (01)00840619079499(11)170720(21)1711100157 (1711100157), (01)00840619079499(11)160512(21)1629100112 (1629100112), (01)00840619079499(11)170912(21)1731100222 (1731100222), (01)00840619079499(11)160523(21)1629100120 (1629100120), (01)00840619079499(11)170728(21)1718100162 (1718100162), (01)00840619079499(11)170911(21)1731100221 (1731100221), (01)00840619079499(11)170913(21)1731100227 (1731100227), (01)00840619079499(11)170914(21)1731100229 (1731100229), (01)00840619079499(11)170911(21)1731100219 (1731100219), (01)00840619079499(11)170919(21)1731100239 (1731100239), (01)00840619079499(11)170918(21)1731100238 (1731100238), (01)00840619079499(11)170920(21)1731100237 (1731100237), (01)00840619079499(11)170807(21)1647100141 (1647100141), (01)00840619079499(11)160512(21)1629100118 (1629100118), (01)00840619079499(11)160512(21)1629100116 (1629100116), (01)00840619079499(11)170717(21)1711100148 (1711100148), (01)00840619079499(11)170720(21)1711100152 (1711100152), (01)00840619079499(11)170720(21)1711100149 (1711100149), (01)00840619079499(11)170726(21)1718100160 (1718100160), (01)00840619079499(11)170728(21)1718100161 (1718100161), (01)00840619079499(11)170925(21)1731100244 (1731100244), (01)00840619079499(11)170615(21)1647100138 (1647100138), (01)00840619079499(11)170911(21)1718100172 (1718100172), (01)00840619079499(11)170716(21)1718100169 (1718100169), (01)00840619079499(11)160908(21)1629100103 (1629100103), (01)00840619079499(11)160908(21)1629100105 (1629100105), (01)00840619079499(11)160505(21)1629100111 (1629100111), (01)00840619079499(11)160520(21)1629100113 (1629100113), (01)00840619079499(11)160523(21)1629100117 (1629100117), (01)00840619079499(11)160523(21)1629100123 (1629100123), (01)00840619079499(11)160523(21)1629100124 (1629100124), (01)00840619079499(11)170922(21)1731100240 (1731100240), (01)00840619079499(11)160526(21)1629100122 (1629100122), (01)00840619079499(11)160525(21)1629100119 (1629100119); b. Part Number 58403-000 - UDI (Serial No.): (01)00840619079505(11)151126(21)140165 (140165), (01)00840619079505(11)151126(21)140153 (140153), (01)00840619079505(11)170207(21)TSH101367 (TSH101367), (01)00840619079505(11)170308(21)TSH200754 (TSH200754), (01)00840619079505(11)170308(21)TSH200753 (TSH200753), (01)00840619079505(11)151126(21)140146 (140146), (01)00840619079505(11)151126(21)140131 (140131), (01)00840619079505(11)170308(21)140260 (140260), (01)00840619079505(11)170308(21)140228 (140228), (01)00840619079505(11)170207(21)140186 (140186), (01)00840619079505(11)151126(21)140151 (140151), (01)00840619079505(11)151126(21)140133 (140133), (01)00840619079505(11)151126(21)140166 (140166), (01)00840619079505(11)170207(21)140227 (140227), (01)00840619079505(11)151126(21)140126 (140126), (01)00840619079505(11)170308(21)140259 (140259), (01)00840619079505(11)170207(21)140210 (140210), (01)00840619079505(11)170207(21)140191 (140191), (01)00840619079505(11)170308(21)TSH200773 (TSH200773), (01)00840619079505(11)170308(21)TSH200822 (TSH200822), (01)00840619079505(11)170308(21)TSH200748 (TSH200748), (01)00840619079505(11)151126(21)140130140130 (01)00840619079505(11)170207(21)140198 (140198), (01)00840619079505(11)170207(21)TSH101372 (TSH101372), (01)00840619079505(11)170207(21)140207 (140207), (01)00840619079505(11)170207(21)140220 (140220), (01)00840619079505(11)170207(21)TSH101379 (TSH101379), (01)00840619079505(11)170308(21)TSH200814 (TSH200814), (01)00840619079505(11)170207(21)140173 (140173), (01)00840619079505(11)170207(21)140176 (140176), (01)00840619079505(11)170207(21)140178 (140178), (01)00840619079505(11)170207(21)140188 (140188), (01)00840619079505(11)170207(21)140221 (140221), (01)00840619079505(11)170207(21)140223 (140223), (01)00840619079505(11)170308(21)140276 (140276), (01)00840619079505(11)170308(21)140233 (140233), (01)00840619079505(11)170308(21)140277 (140277), (01)00840619079505(11)151126(21)140144 (140144), (01)00840619079505(11)170207(21)TSH101369 (TSH101369), (01)00840619079505(11)151126(21)TSFA01666TSFA01666 (01)00840619079505(11)170207(21)TSH101375 (TSH101375), (01)00840619079505(11)170207(21)TSH101381 (TSH101381), (01)00840619079505(11)170308(21)TSH200815 (TSH200815), (01)00840619079505(11)170308(21)TSH200755 (TSH200755), (01)00840619079505(11)170207(21)TSH101380 (TSH101380), (01)00840619079505(11)170308(21)TSH200818 (TSH200818), (01)00840619079505(11)170308(21)TSH200821 (TSH200821), (01)00840619079505(11)170308(21)TSH200811 (TSH200811), (01)00840619079505(11)151126(21)TSFA01663 (TSFA01663), (01)00840619079505(11)151126(21)TSFA01667 (TSFA01667), (01)00840619079505(11)151126(21)TSFA01658 (TSFA01658), (01)00840619079505(11)170207(21)TSH101354 (TSH101354), (01)00840619079505(11)170207(21)TSH101353 (TSH101353), (01)00840619079505(11)170308(21)TSH200739 (TSH200739), (01)00840619079505(11)170207(21)TSH101386 (TSH101386), (01)00840619079505(11)170207(21)TSH101349 (TSH101349), (01)00840619079505(11)170308(21)TSH200752 (TSH200752), (01)00840619079505(11)151126(21)TSFA01695 (TSFA01695), (01)00840619079505(11)170207(21)TSH101351 (TSH101351), (01)00840619079505(11)170308(21)TSH200742 (TSH200742), (01)00840619079505(11)151126(21)140129 (140129), (01)00840619079505(11)151126(21)140137 (140137), (01)00840619079505(11)151126(21)140155 (140155), (01)00840619079505(11)151126(21)140171 (140171), (01)00840619079505(11)151126(21)140159 (140159), (01)00840619079505(11)151126(21)140147 (140147), (01)00840619079505(11)151126(21)140138 (140138), (01)00840619079505(11)170308(21)TSH200771 (TSH200771), (01)00840619079505(11)151126(21)140172 (140172), (01)00840619079505(11)151126(21)140123 (140123).

Distribution

Distributed to the states of CA, FL, NC, NE, NY, and WA.

Quantity

36 devices