Refinity Rotational Intravascular Ultrasound (IVUS) Catheter; Product number: 400-0200.289; Catalog number 89800. Cardiovascular: The REFINITY (or REFINITY ST) Rotational IVUS Catheter (REFINITY or REFINITY ST) is intended for the intravascular ultrasound examination of coronary arteries. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal interventional procedures. REFINITY (or REFINITY ST) catheters are designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures.

Recall: Z-1775-2017 · Reported April 19, 2017

Enforcement

Recall Number
Z-1775-2017
Event ID
76885
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Volcano Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 19, 2017
Initiation Date
March 29, 2017
Classification Date
April 13, 2017
Termination Date
June 21, 2017
Address
2870 Kilgore Rd, N/A, Rancho Cordova, CA, 95670-6133, United States

Description

Refinity Rotational Intravascular Ultrasound (IVUS) Catheter; Product number: 400-0200.289; Catalog number 89800. Cardiovascular: The REFINITY (or REFINITY ST) Rotational IVUS Catheter (REFINITY or REFINITY ST) is intended for the intravascular ultrasound examination of coronary arteries. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal interventional procedures. REFINITY (or REFINITY ST) catheters are designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures.

Reason

High complaint rates of kinking near the junction of the proximal shaft to telescope section of the catheter. Kinking could result in disruption of normal clinical workflow or loss of imaging.

Code Info

Lot # 50068951, Expiry date 06/30/2018; Lot # 50068952, Expiry date 06/30/2018; Lot # 50069133, Expiry date 06/30/2018; Lot # 50083387, Expiry Date 11/30/2018; Lot # 50083390, Expiry Date 11/30/2018.

Distribution

Distribution to US and Japan

Quantity

718 total units