FDA Enforcement Class II Terminated

LIGACLIP¿ Endoscopic Rotating Multiple Clip Applier (ER320), 10mm M/L Product Usage: The LIGACLIP Endoscopic Rotating Multiple Clip Applier is a sterile, single patient use instrument designed to provide a means of ligation through ENDOPATH¿ surgical trocars. The instrument delivers 20 titanium clips that individually advance after each firing. The shaft is made of a low glare material that minimizes reflective distortion. It is designed to rotate 360¿ in either direction. The rotating knob is located to allow for a one-handed technique.

Recall: Z-1775-2013 · Reported July 31, 2013

Enforcement

Recall Number
Z-1775-2013
Event ID
65141
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ethicon Endo-Surgery Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 31, 2013
Initiation Date
April 12, 2013
Classification Date
July 19, 2013
Termination Date
May 1, 2014
Address
4545 Creek Rd, N/A, Cincinnati, OH, 45242-2803, United States

Description

LIGACLIP¿ Endoscopic Rotating Multiple Clip Applier (ER320), 10mm M/L Product Usage: The LIGACLIP Endoscopic Rotating Multiple Clip Applier is a sterile, single patient use instrument designed to provide a means of ligation through ENDOPATH¿ surgical trocars. The instrument delivers 20 titanium clips that individually advance after each firing. The shaft is made of a low glare material that minimizes reflective distortion. It is designed to rotate 360¿ in either direction. The rotating knob is located to allow for a one-handed technique.

Reason

Ethicon Endo-Surgery is initiating a voluntary recall for LIGACLIP¿ 10mm M/L Endoscopic Rotating Multiple Clip Applier (ER320) due to potential clip formation and feeding issues which may result in improper clip formation, vessel damage and/or insufficient occlusion of the vessel or other structure

Code Info

H44Z7W, 37ER320, 01/03/12 to 11/15/16; J4A016, 37ER320, 01/05/12 to 12/15/16; J4A11G, 37ER320, 01/10/12 to 12/15/16; J4A124, 37ER320, 01/10/12 to 12/15/16; J4A145 37ER320, 01/11/12 to 12/15/16 J4A182, 37ER320, 01/13/12 to 12/15/16; J4A21T, 37ER320, 01/27/12 to 12/15/16; J4A28D, 37ER320, 01/19/12 to 12/15/16; J4A28U, 37ER320, 01/19/12 to 12/15/16; J4A335, 37ER320, 01/23/12 to 12/15/16; J4A38L, 37ER320, 01/25/12 to 12/15/16; J4A39D, 37ER320, 01/26/12 to 12/15/16; J4A406, 37ER320, 01/26/12 to 12/15/16; J4A40P, 37ER320, 02/07/12 to 12/15/16; J4A45T, 37ER320, 01/29/12 to 12/15/16; J4A461, 37ER320, 01/30/12 to 12/15/16; J4A46T, 37ER320, 01/30/12 to 12/15/16; J4A48U,), 37ER320, 02/02/12 to 12/15/16; J4A49L, 37ER320, 02/01/12 to 12/15/16; J4A525, 37ER320, 02/02/12 to 12/15/16; J4A535, 37ER320, 02/02/12 to 12/15/16; J4A559, 41XBB48B 02/02/12 to 07/15/16; J4A60K, 37ER320, 02/08/12 to 01/15/17; J4A60W, 37ER320, 02/08/12 to 01/15/17; J4A61Z, 37ER320, 02/08/12 to 01/15/17; J4A62E, 37ER320, 02/08/12 to 01/15/17; J4A62U, 37ER320, 02/09/12 to 01/15/17; J4A71W, 37ER320, 02/15/12 to 01/15/17; J4A729, 37ER320, 02/15/12 to 01/15/17; J4A73H, 37ER320, 02/15/12 to 01/15/17; J4A743, 37ER320, 02/16/12 to 01/15/17; J4A74T, 37ER320, 02/16/12 to 01/15/17; J4A755, 37ER320, 02/16/12 to 01/15/17; J4A75P, 37ER320, 02/16/12 to 01/15/17; J4A86E, 37ER320, 02/24/12 to 01/15/17; J4A871, 37ER320, 02/25/12 to 01/15/17; J4A87L, 37ER320, 02/25/12 to 01/15/17; J4A889, 37ER320, 02/25/12 to 01/15/17; J4A92X, 41XBB48B 02/28/12 to 08/15/16; J4A97H, 37ER320, 03/02/12 to 01/15/17; J4A98T, 37ER320, 03/02/12 to 01/15/17; J4A997, 37ER320, 03/03/12 to 01/15/17; J4A99V, 37ER320, 03/03/12 to 02/15/17; J4AAOM, 37ER320, 03/03/12 to 02/15/17; J4AA1H, 37ER320, 03/05/12 to 02/15/17; J4AA71, 37ER320, 03/07/12 to 02/15/17; J4AA7V, 37ER320, 03/08/12 to 02/15/17; J4AA84, 37ER320, 03/08/12 to 02/15/17; J4AA9D, 37ER320, 03/08/12 to 02/15/17; J4AC2X, 37ER320, 03/10/12 to 02/15/17; J4AD09, 37ER320, 03/15/12 to 02/15/17; J4AD1A, 37ER320, 03/15/12 to 02/15/17; J4AD2G, 37ER320, 03/16/12 to 02/15/17; J4AE3W, 37ER320, 03/22/12 to 02/15/17; J4AE49, 37ER320, 03/22/12 to 02/15/17; J4AE57, 37ER320, 03/23/12 to 02/15/17; J4AE63, 37ER320, 03/23/12 to 02/15/17; J4AE6R, 37ER320, 04/13/12 to 02/15/17; J4AE7E, 37ER320, 04/11/12 to 02/15/17; J4AE86, 37ER320, 03/24/12 to 02/15/17; J4AE8L, 37ER320, 03/24/12 to 02/15/17; J4AF09, 41XBB48B, 03/27/12 to 10/15/16 84; J4AF15, 37ER320, 03/28/12 to 02/15/17; J4AF1N, 37ER320, 03/28/12 to 02/15/17; J4AF2G, 37ER320, 03/28/12 to 02/15/17; J4AF3K, 37ER320, 03/28/12" 02/15/17; J4AF48, 37ER320, 03/29/12 to 02/15/17; J4AF4J, 37ER320, 03/29/12 to 02/15/17; J4AGOV, 37ER320, 04/03/12 to 02/15/17; J4AG4V, 37ER320, 04/05/12 to 03/15/17; J4AG91, 37ER320, 04/11/12 to 03/15/17; J4AG9G, 37ER320, 04/18/12 to 03/15/17; J4AHOV, 37ER320, 04/12/12 to 03/15/17; J4AH8N, 37ER320, 04/15/12 to 03/15/17; J4AH96, 37ER320, 04/16/12 to 03/15/17; J4AH9C, 37ER320, 04/16/12 to 03/15/17; J4AH9R, 37ER320, 04/16/12 to 03/15/17; J4AJ3A, 37ER320, 04/18/12 to 03/15/17; J4AJ3T, 37ER320, 04/18/12 to 03/15/17; J4AJ6P, 37ER320, 04/20/12 to 03/15/17; J4AJ7M, 37ER320, 04/20/12 to 03/15/17; J4AKOP, 37ER320, 04/21/12 to 03/15/17; J4AK14, 37ER320, 04/22/12 to 03/15/17; J4AK1H, 37ER320, 04/23/12 to 03/15/17; J4AK1R, 37ER320, 04/23/12 to 03/15/17; J4AK2V, 37ER320, 04/24/12 to 03/15/17; J4AK35, 37ER320, 04/24/12 to 03/15/17; J4AK40, 37ER320, 04/24/12 to 03/15/17; J4AK5U, 37ER320, 04/25/12 to 03/15/17; J4AK67, 37ER320, 04/25/12 to 03/15/17; 2136; J4AK7D, 37ER320, 04/26/12 to 03/15/17; J4AK8C, 37ER320, 04/26/12 to 03/15/17; 1152; 4AK9R, 37ER320, 04/28/12 to 03/15/17; J4ALOU, 37ER320, 04/27/12 to 03/15/17; J4AL65, 37ER320, 05/02/12 to 03/15/17; J4AL6P, 37ER320, 05/02/12 to 03/15/17; 2172; J4AL87, 37ER320, 05/11/12 to 03/15/17; 2244; J4AL90, 37ER320, 05/03/12 to 04/15/17; J4AL9H, 37ER320, 05/03/12 to 04/15/17; J4AM11, 37ER320, 05/03/12 to 04/15/17; J4AM1K, 37ER320, 05/04/12 to 04/15/17; J4AM2E, 37ER320, 05/04/12 to 04/15/17; J4AM32, 37ER320, 05/04/12 to 04/15/17; J4AM3G, 37ER320, 05/05/12 to 04/15/17; J4AM45, 37ER320, 05/05/12 to 04/15/17; J4AM7R, 37ER320, 05/08/12 to 04/15/17; J4AM7W, 37ER320, 05/08/12 to 04/15/17; J4AM8H, 37ER320, 05/09/12 to 04/15/17; J4AM9J, 37ER320, 05/09/12 to 04/15/17; J4ANOC, 37ER320, 05/09/12 to 04/15/17; J4AN14, 37ER320, 05/09/12 to 04/15/17; J4AN1P, 37ER320, 05/10/12 to 04/15/17; J4AN1X, 37ER320, 05/10/12 to 04/15/17; J4AN2K, 37ER320, 05/10/12 to 04/15/17; J4AN35, 37ER320, 05/10/12 to 04/15/17; J4AR7E, 41XBB48B 05/24/12 to 01/15/17; J4AT2G, 37ER320, 06/02/12 to 04/15/17; 1152; J4AT2H, 37ER320, 05/31/12 to 04/15/17; J4AUOH, 37ER320, 06/04/12 to 04/15/17; J4AU23, 37ER320, 06/04/12 to 04/15/17; J4AU29, 37ER320, 06/05/12 to 05/15/17; J4AU2V, 37ER320, 06/05/12 to 05/15/17; J4AV5J, 37ER320, 06/12/12 to 05/15/17; J4AV5T, 37ER320, 06/12/12 to 05/15/17; J4AW02, 37ER320, 06/13/12 to 05/15/17; J4AWOG, 37ER320, 06/14/12 to 05/15/17; J4AW7D, 37ER320, 06/18/12 to 05/15/17; J4AW7N, 37ER320, 06/18/12 to 05/15/17; J4AW86, 37ER320, 06/19/12 to 05/15/17; J4AX78, 37ER320, 06/23/12 to 05/15/17; J4AX7M, 37ER320, 06/23/12 to 05/15/17; J4AX84, 37ER320, 06/24/12 to 05/15/17; J4AX8M, 37ER320, 06/24/12 to 05/15/17; J4AY22, 37ER320, 06/26/12 to 05/15/17; J4AY23, 37ER320, 06/27/12 to 05/15/17; J4AY24, 37ER320, 06/27/12 to 05/15/17; J4AY26, 37ER320, 06/27/12 to 05/15/17; J4AY7E, 37ER320, 06/29/12 to 05/15/17; J4AY8R, 37ER320, 06/29/12 to 05/15/17; J4AY9D, 37ER320, 07/09/12 to 05/15/17; J4AZOL, 37ER320, 07/02/12 to 05/15/17; J4AZ1X, 37ER320, 07/11/12 to 05/15/17; J4AZ2H, 37ER320, 07/03/12 to 06/15/17; J4AZ6R, 37ER320, 07/06/12 to 06/15/17; J4AZ7K, 37ER320, 07/06/12 to 06/15/17; J4C04N, 37ER320, 07/11/12 to 06/15/17; J4C059, 37ER320, 07/12/12 to 06/15/17; J4C08U, 37ER320, 07/13/12 to 06/15/17; J4C09E, 37ER320, 07/14/12 to 06/15/17; J4C12Y, 37ER320, 07/16/12 to 06/15/17; J4C14F, 37ER320, 07/18/12 to 06/15/17; J4C16V, 37ER320, 07/18/12 to 06/15/17; J4C216, 37ER320, 07/23/12 to 06/15/17; J4C21H, 37ER320, 07/24/12 to 06/15/17; J4C21N, 37ER320, 07/24/12 to 06/15/17; J4C22M, 37ER320, 07/24/12 to 06/15/17; J4C30T, 37ER320, 07/27/12 to 06/15/17; J4C31J, 37ER320, 07/28/12 to 06/15/17; J4C32M, 37ER320, 08/21/12 to 06/15/17; J4C332, 37ER320, 07/30/12 to 06/15/17; J4C399, 37ER320, 08/03/12 to 06/15/17; J4C41U, 37ER320, 08/06/12 to 07/15/17; J4C42C, 37ER320, 08/04/12 to 07/15/17; J4C42X, 37ER320, 08/03/12 to 07/15/17; J4C45J, 37ER320, 08/07/12 to 07/15/17; J4C46Y, 37ER320, 08/07/12 to 07/15/17; J4C47A, 37ER320, 08/07/12 to 07/15/17; J4C489, 37ER320, 08/08/12 to 07/15/17; J4C49Y, 37ER320, 08/08/12 to 07/15/17; J4C508, 37ER320, 08/09/12 to 07/15/17; J4C510, 37ER320, 08/09/12 to 07/15/17; J4C57X, 37ER320, 08/14/12 to 07/15/17; J4C59E, 37ER320, 08/14/12 to 07/15/17; J4C606, 37ER320, 08/15/12 to 07/15/17; J4C65Z, 37ER320, 08/18/12 to 07/15/17; J4C66P, 37ER320, 08/20/12 to 07/15/17; J4C678, 37ER320, 08/21/12 to 07/15/17; J4C67K, 37ER320, 08/21/12 to 07/15/17; J4C762, 37ER320, 08/24/12 to 07/15/17; J4C765, 37ER320, 08/24/12 to 07/15/17; J4C76T, 37ER320, 08/24/12 to 07/15/17; J4C84K, 37ER320, 08/30/12 to 07/15/17; J4C84Z, 37ER320, 08/30/12 to 07/15/17; J4C90P, 37ER320, 09/05/12 to 07/15/17; J4C91H, 37ER320, 09/02/12 to 07/15/17; J4C91W, 37ER320, 09/05/12 to 07/15/17; J4C92P, 37ER320, 09/05/12 to 07/15/17; J4C96L, 37ER320, 09/06/12 to 08/15/17; J4C97D, 37ER320, 09/06/12 to 08/15/17; J4C97Z, 37ER320, 09/07/12 to 08/15/17; J4C98X, 37ER320, 09/07/12 to 08/15/17; J4CA3A, 37ER320, 09/11/12 to 08/15/17; J4CA4L, 37ER320, 09/11/12 to 08/15/17; J4CA62, 37ER320, 09/12/12 to 08/15/17; J4CA6K, 37ER320, 09/12/12 to 08/15/17; J4CA7F, 37ER320, 09/13/12 to 08/15/17; J4CA7J, 37ER320, 09/13/12 to 08/15/17; J4CA96, 37ER320, 09/13/12 to 08/15/17; J4CA9V, 37ER320, 09/13/12 to 08/15/17; J4CC63, 37ER320, 09/19/12 to 08/15/17; J4CD25, 42XBB48B 09/21/12 to 02/15/17; J4CD4R, 37ER320, 09/21/12 to 08/15/17; J4CD5H, 37ER320, 09/22/12 to 08/15/17; J4CD6D, 42XBB48B 09/25/12 to 02/15/17; J4CD6R, 37ER320, 09/25/12 to 08/15/17; J4CD8F, 42XBB48B 09/27/12 to 02/15/17; J4CE05, 37ER320, 09/26/12 to 08/15/17; J4CEOA, 37ER320, 09/26/12 to 08/15/17; J4CE1U, 37ER320, 09/26/12 to 08/15/17; J4CF6M, 37ER320, 10/05/12 to 09/15/17; J4CF6P, 37ER320, 10/05/12 to 09/15/17; J4CF7E, 37ER320, 10/05/12 to 09/15/17; J4CF7R, 37ER320, 10/06/12 to 09/15/17; J4CF80, 37ER320, 10/06/12 to 09/15/17; J4CF8X, 37ER320, 10/08/12 to 09/15/17; J4CGOP, 37ER320, 10/09/12 to 09/15/17; J4CG17, 37ER320, 10/10/12 to 09/15/17; J4CG1U, 37ER320, 10/10/12 to 09/15/17; J4CG2U, 37ER320, 10/10/12 to 09/15/17; J4CG37, 37ER320, 10/10/12 to 09/15/17; J4CG4L, 37ER320, 10/11/12 to 09/15/17; J4CG66, 37ER320, 10/11/12 to 09/15/17; J4CG8R, 37ER320, 10/16/12 to 09/15/17; J4CG8X, 37ER320, 10/16/12 to 09/15/17; J4CH10, 37ER320, 10/15/12 to 09/15/17; J4CH4U, 37ER320, 10/17;/12 to 09/15/17; J4CH4V, 37ER320, 11/16/12 to 09/15/17; J4CH97, 37ER320, 10/20/12 to 09/15/17; J4CH9R, 37ER320, 10/20/12 to 09/15/17; J4CJ17, 37ER320, 10/20/12 to 09/15/17; J4CJ61, 37ER320, 10/25/12 to 09/15/17; J4CJ6L, 37ER320, 10/25/12 to 09/15/17; J4CKOV, 37ER320, 10/27/12 to 09/15/17; J4CK16, 37ER320, 10/29/12 to 09/15/17; J4CK3L, 37ER320, 10/30/12 to 09/15/17; J4CK9D, 37ER320, 11/02/12 to 09/15/17; J4CK9J, 37ER320, 11/02/12 to 09/15/17; J4CL23, 37ER320, 11/03/12 to 10/15/17; J4CL3F, 37ER320, 11/06/12 to 10/15/17; J4CL3U, 37ER320, 11/06/12 to 10/15/17; J4CL49, 37ER320, 11/06/12 to 10/15/17; J4CL8D, 37ER320, 11/08/12 to 10/15/17; J4CL9A, 37ER320, 11/08/12 to 10/15/17; J4CL9Z, 37ER320, 11/17;/12 to 10/15/17; J4CN1C, 37ER320, 11/15/12 to 10/15/17; J4CN9T, 37ER320, 11/20/12 to 10/15/17; J4CP1P, 37ER320, 11/27/12 to 10/15/17; J4CP1Y, 37ER320, 11/27/12 to 10/15/17; J4CP4C, 37ER320, 11/27/12 to 10/15/17; J4CP4L, 37ER320, 11/27/12 to 10/15/17; J4CP8U, 37ER320, 11/29/12 to 10/15/17; J4CR1C, 37ER320, 11/30/12 to 10/15/17; J4CR3L, 37ER320, 12/04/12 to 10/15/17; J4CR44, 37ER320, 12/04/12 to 10/15/17; J4CR6Y, 37ER320, 12/04/12 to 11/15/17; J4CR7X, 37ER320, 12/07/12 to 11/15/17; J4CT1U, 37ER320, 12/07/12 to 11/15/17; J4CT27, 37ER320, 12/09/12 to 11/15/17; J4CTSR, 37ER320, 12/11/12 to 11/15/17; J4CU1U, 37ER320, 12/12/12 to 11/15/17; J4CU27, 37ER320, 12/12/12 to 11/15/17; J4CU3K, 37ER320, 12/14/12 to 11/15/17; J4CU68, 37ER320, 12/15/12 to 11/15/17; J4CV2F, 37ER320, 12/20/12 to 11/15/17; J4CV48, 37ER320, 12/20/12 to 11/15/17; J4CV4J, 37ER320, 12/21/12 to 11/15/17; J4CV4U, 37ER320, 12/21/12 to 11/15/17; J4CV58, 37ER320, 12/21/12 to 11/15/17; J9002K, 41XCB57S 01/03/12 to 12/15/16; J9002L, 41XCD52S 01/03/12 to 12/15/16; J9003P, 02CK017 01/05/12 to 12/31/99; J90058, 41XCC51S 01/06/12 to 12/15/16; J90078, 41XCDS15 01/05/12 to 12/15/16; J9012T, 41KNC61XL 01/06/12 to 12/15/16; J9012U, 41XCD50S 01/06/12 to 12/15/16; J9012V, 41FNC42XL, 01/06/12 to 12/15/16; J90143, 41XCB52S 01/09/12 to 12/15/16; J90145, 41XCB57S 01/09/12 to 12/15/16; J90146, 41FNC41XL 01/09/12 to 12/15/16; J9014N, 02CK017 01/06/12 to 12/31/99; J9016G, 41XCB52S 01/09/12 to 12/15/16; J9016H, 41FNC43XL 01/09/12 to 12/15/16; J9016J 41TNC49XL 01/10/12 to 12/15/16; J9016K 41KNC60XL 01/10/12 to 12/15/16; J9020X 41XCD51S 01/10/12 to 12/15/16; J9022W 02CK017 01/10/12 to 12/31/99; J90249 41KBC17XL 01/12/12 to 12/15/16; J9026C 41XCD51S 01/11/12 to 12/15/16; J9026D 41XCC51S 01/12/12 to 12/15/16; J9026H 02CK017 01/11/12 to 12/31/99; J9028G 41XCB57S 01/13/12 to 12/15/16; J9032A 41XCD50S 01/13/12 to 12/15/16; J9032D 41KBC17XL 01/13/12 to 12/15/16; J9033P 02CK017 01/13/12 to 12/31/99; J90379 41XCC51S 01/16/12 to 12/15/16; J9037A 41FNC42XL, 01/16/12 to 12/15/16; J9038A 41KBC17XL 01/17;/12 to 12/15/16; J90390 41KNC61XL 01/17;/12 to 12/15/16; J9042C 02CK017 01/17;/12 to 12/31/99; J9044R 41XCD51S 01/18/12 to 12/15/16; J9044T 41FNC42XL, 01/18/12 to 12/15/16; J9047G 41XCD51S 01/19/12 to 12/15/16; J9047H 41FNC41XL 01/19/12 to 12/15/16; J9051P 41XCD51S 01/20/12 to 12/15/16; J9051R 41XCC51S 01/20/12 to 12/15/16; J9051T 41XCC50S 01/20/12 to 12/15/16; J9051U 41FNC41XLD 01/20/12 to 12/15/16; J90574, 41XCB57S 01/23/12 to 12/15/16; J90575, 41TNC49XL 01/24/12 to 12/15/16; J90598, 41FNC42XL, 01/24/12 to 12/15/16; J9059A, 41XCD51S 01/25/12 to 12/15/16; J90643, 41XCD52S 01/26/12 to 12/15/16; J90644, 41KNC61XL 01/26/12 to 12/15/16; J90646, 41XCC53S 01/27/12 to 12/15/16; J9066Z, 41XCD51S 01/27/12 to 12/15/16; J90671, 41KNC60XL 01/27/12 to 12/15/16; J9068K, 02CK017 01/26/12 to 12/31/99; J90743, 41XCD51S 01/30/12 to 12/15/16; J9074T, 41TNC69XL 01/31/12 to 12/15/16; J9074X, 41FNC41XL 02/01/12 to 12/15/16; J90771, 41XCB52S, 01/31/12 to 12/15/16; J9080P, 41XCD51S, 02/01/12 to 12/15/16; J9080R, 41FNC42XL, 02/01/12 to 12/15/16; J90823, 41XCD52S, 02/01/12 to 12/15/16; J90824, 41XCC51S, 02/01/12 to 12/15/16; J90825, 41XCB57S, 02/01/12 to 12/15/16; J9082G, 02CK017, 02/01/12 to 12/31/99; J9084F, 41FNC43XL, 02/02/12 to 12/15/16; J90867, 41FNC41XL, 02/03/12 to 12/15/16; J90881, 41KNC09XL, 02/03/12 to 12/15/16; J90883, 41KNC60XL, 02/03/12 to 12/15/16; J9090A, 41XCD50S, 02/03/12 to 12/15/16; J9091Z, 02CK017, 02/03/12 to 12/31/99; J90931, 41XCD51S, 02/07/12 to 01/15/17; J90948, 41XCC53S, 02/07/12 to 01/15/17; J90949, 41TNC49XL, 02/07/12 to 01/15/17; J90968, 41FNC41XLD, 02/08/12 to 01/15/17; J90969, 41KNC07XL, 02/09/12 to 01/15/17; J9098C, 41TNC36XL, 02/09/12 to 01/15/17; J90993, 02CK017, 02/08/12 to 12/31/99; J90C1W, 41XCC51S, 02/14/12 to 01/15/17; J90C1X, 41XCD51S, 02/14/12 to 01/15/17; J90C1Y, 41XCD52S, 02/14/12 to 01/15/17; J90C1Z, 41XCC50S, 02/14/12 to 01/15/17; J90C67, 41XCD51S, 02/15/12 to 01/15/17; J90C8C, 02CK017, 02/15/12 to 12/31/99; J90DOL, 41XCD51S, 02/16/12 to 01/15/17; J9003N, 02CK017, 02/17/12 to 12/31/99; J9004Z, 41FNC43XL, 02/17/12 to 01/15/17; J90050, 41XCC51S, 02/17/12 to 01/15/17; J90051, 41FNC41XL, 02/17/12 to 01/15/17; J9007A, 41XCD51S, 02/17/12 to 01/15/17; J9008J, 41XCD51S, 02/21/12 to 01/15/17; J9008K, 41TNC36XL, 02/21/12 to 01/15/17; J9008L, 41KNC60XL, 02/21/12 to 01/15/17; J9009H, 41FNC41XLD, 02/20/12 to 01/15/17; J9009K, 02CK017, 02/20/12 to 12/31/99; J90E2R, 41XCD51S, 02/21/12 to 01/15/17; J90E2T, 41XCB57S, 02/21/12 to 01/15/17; J90E40, 41FNC42XL, 02/22/12 to 01/15/17; J90E5P, 41XCD51S, 02/22/12 to 01/15/17; J90E5T, 41XCB52S, 02/22/12 to 01/15/17; J90E7G, 41XCC53S, 02/22/12 to 01/15/17; J90E7R, 02CK017, 02/22/12 to 12/31/99; J90F03, 41XCD51S, 02/23/12 to 01/15/17; J90F06, 41XCD50S, 02/23/12 to 01/15/17; J90F2P, 41XCD515, 02/24/12 to 01/15/17; J90F2R, 41KBC17XL, 02/24/12 to 01/15/17; J90F2T, 41TNC69XL, 02/24/12 to 01/15/17; J90F2U, 41XCD525, 02/27/12 to 01/15/17; J90F4T, 41XCC515, 02/27/12 to 01/15/17; J90F68, 41XCD515, 02/27/12 to 01/15/17; J90F6A, 41KNC07XL, 02/27/12 to 01/15/17; J90F6C, 41FNC42XL, 02/27/12 to 01/15/17; J90F6U, 41FNC41XL, 02/27/12 to 01/15/17; J90F90, 41FNC41XLD, 02/28/12 to 01/15/17; J90F91, 41XCD515, 02/28/12 to 01/15/17; J90F9Z, 41KNC60XL, 02/28/12 to 01/15/17; J90GOO, 41KNC61XL, 02/28/12 to 01/15/17; J90GOF, 02CK017, 02/28/12 to 12/31/99; J90G23, 41XCD515, 02/29/12 to 01/15/17; J90G25, 41TNC36XL, 02/29/12 to 01/15/17; J90G26, 41XCC515, 02/29/12 to 01/15/17; J90G3A, 41KNC60XL, 02/29/12 to 01/15/17; J90G5F, 41XCD515, 03/01/12 to 01/15/17; J90G5H, 41FNC42XL, 03/01/12 to 01/15/17; J90G5J, 41XCD525, 03/01/12 to 01/15/17; J90G9G, 41XCB515, 03/02/12 to 01/15/17; J90G9H, 41KBC17XL, 03/02/12 to 01/15/17; J90G9J, 41XCC515, 03/02/12 to 01/15/17; J90G9K, 41TNC36XL, 03/02/12 to 01/15/17; J90HOP, 41TNC69XL, 03/05/12 to 01/15/17; J90H1R, 41XCD51S, 03/05/12 to 02/15/17; J90H6E, 41FNC42XL, 03/07/12 to 02/15/17; J90H6K, 41FNC41XL, 03/07/12 to 02/15/17; J90H6L, 41KBC17XL, 03/07/12 to 02/15/17; J90H7J, 41KNC09XL, 03/07/12 to 02/15/17; J90H7K, 41XCD525, 03/07/12 to 02/15/17; J90H7L, 41XCB575, 03/07/12 to 02/15/17; J90J1A, 02CK017, 03/07/12 to 12/31/99; J90J36, 41XCD51S, 03/09/12 to 02/15/17; J90J7X, 41XCC515, 03/09/12 to 02/15/17; J90J7Y, 41XCD525, 03/09/12 to 02/15/17; J90J7Z, 41FNC42XL, 03/09/12 to 02/15/17; J90J8P, 41FNC43XL, 03/12/12 to 02/15/17; J90KOH, 41XCD515, 03/12/12 to 02/15/17; J90KOJ, 41TNC20XL, 03/12/12 to 02/15/17; J90KOK, 41XCB575, 03/12/12 to 02/15/17; J90K1V, 41KNC60XL, 03/13/12 to 02/15/17; J90K4E, 41XCC505, 03/13/12 to 02/15/17; J90K4G, 41FNC41XL, 03/13/12 to 02/15/17; J90K4H, 41TNC49XL, 03/13/12 to 02/15/17; J90K4J, 41KNC07XL, 03/13/12 to 02/15/17; J90K5E, 41KNC61XL, 03/13/12 to 02/15/17; J90K5F, 41KNC09XL, 03/13/12 to 02/15/17; J90K6K, 02CK017, 03/13/12 to 12/31/99; J90K7N, 41FNC41XLD, 04/13/12 to 02/15/17; J90K7P, 41KNC09XL, 03/14/12 to 02/15/17; J90K7R, 41TNC49XL, 03/14/12 to 02/15/17; J90K7T, 41KNC07XL, 03/14/12 to 02/15/17; J90K9D, 41XCC51S, 03/14/12 to 02/15/17; J90L1A, 41FNC42XL, 03/15/12 to 02/15/17; J90L1D, 41XCD50S, 03/15/12 to 02/15/17; J90L1E, 41XCD52S, 03/15/12 to 02/15/17; J90L2L, 41FNC41XL, 03/15/12 to 02/15/17; J90L4E, 41KNC03XL, 03/16/12 to 02/15/17; J90L4F, 41KNC60XL, 03/16/12 to 02/15/17; J90L62, 41FNC43XL, 03/16/12 to 02/15/17; J90L6G, 02CK017, 03/16/12 to 12/31/99; J90M16, 02CK017, 03/19/12 to 12/31/99; J90M1T, 41FNC43XLD, 03/20/12 to 02/15/17; J90M1U, 41XCC51S, 03/20/12 to 02/15/17; J90M3G, 41FNC41XLD, 03/21/12 to 02/15/17; J90M3H, 41KNC19XL, 03/22/12 to 02/15/17; J90M3K, 41FNC42XL, 03/22/12 to 02/15/17; J90M4H, 41FNC41XL, 03/23/12 to 02/15/17; J90M4J, 41KNC60XL, 03/23/12 to 02/15/17; J90M50, 02CK017, 03/20/12 to 12/31/99; J90NOJ, 41KNC03XL, 03/23/12 to 02/15/17; J90NOK, 41FNC42XL, 03/23/12 to 02/15/17; J90N33, 41XCC51S 03/22/12 to 02/15/17; J90N4Y, 41XCC51S 03/27/12 to 02/15/17; J90P15, 41XCD51S 03/27/12 to 02/15/17; J90P16, 41FNC43XLD 03/27/12 to 02/15/17; J90P44, 41XCC51S 03/28/12 to 02/15/17; J90P5U, 02CK017 03/27/12 to 12/31/99; J90P6G, 02CK017 03/28/12 to 12/31/99; J90P7T, 41FNC42XL, 03/28/12 to 02/15/17; J90R22, 41KNC60XL 03/29/12 to 02/15/17; J90R23, 41FNC43XL 03/29/12 to 02/15/17; J90R24, 41XCC51S 03/29/12 to 02/15/17; J90R25, 41XCB52S 03/29/12 to 02/15/17; J90R3H, 41KNC07XL 03/29/12 to 02/15/17; J90R3J, 41FNC41XL 03/29/12 to 02/15/17; J90R45, 02CK017 03/29/12 to 12/31/99; J90T34, 41XCC51S 04/02/12 to 02/15/17; J90T35, 41KNC61XL 04/03/12 to 02/15/17; J90T3Z, 41XCB57S 04/02/12 to 02/15/17; J90T5D, 41XCC51S 04/04/12 to 03/15/17; J90T5E, 41XCB57S 04/04/12 to 03/15/17; J90T5F, 41FNC42XL, 04/04/12 to 03/15/17; J90T5G, 41FNC41XL 04/04/12 to 03/15/17; J90T90, 41KNC07XL 04/09/12 to 03/15/17; J90T91, 41XCB52S 04/09/12 to 03/15/17; J90T92, 41FNC43XL 04/09/12 to 03/15/17; J90T93, 41KNC61XL 04/09/12 to 03/15/17; J90UOT, 02CK017 04/04/12 to 12/31/99; J90U41, 41FNC41XLD 04/10/12 to 03/15/17; J90U50, 41FNC42XL, 04/10/12 to 03/15/17; J90U51, 41XCC51S 04/10/12 to 03/15/17; J90U52, 41KNC07XL 04/10/12 to 03/15/17; J90U53, 41FNC43XL 04/10/12 to 03/15/17; J90U54, 41TNC20XL 04/10/12 to 03/15/17; J90W98, 02CK017 04/10/12 to 12/31/99; J90X3A, 02CK017 04/11/12 to 12/31/99; J90X6K, 02CK017 04/12/12 to 12/31/99; J90X6X, 41FNC42XL, 04/12/12 to 03/15/17; J90X6Y, 41XCC51S 04/12/12 to 03/15/17; J90X7D, 41FNC41XL 04/12/12 to 03/15/17; J90X7K, 41KNC07XL 04/13/12 to 03/15/17; J90X9A, 41FNC43XL 04/13/12 to 03/15/17; J90VOO, 02CK017 04/13/12 to 12/31/99; J90V2U, 41XCD51S 04/17;/12 to 03/15/17; J90V41, 02CK017 04/16/12 to 12/31/99; J90V82, to 02CK017 04/17;/12 to 12/31/99; J90Z32, 41FNC42XL, 04/18/12 to 03/15/17; J90Z33, 41XCC51S 04/18/12 to 03/15/17; J90Z43, 41KBC17XL 04/19/12 to 03/15/17; J90Z44, 41XCD52S 04/18/12 to 03/15/17; J90Z5Z, 41FNC41XL, 04/19/12 to 03/15/17; J91005, 02CK017, 04/20/12 to 12/31/99; J9107X, 41FNC42XL, 04/24/12 to 03/15/17; J9110D, 41XCC51S, 04/25/12 to 03/15/17; J9110E, 41KBC17XL, 04/25/12 to 03/15/17; J9112K, 02CK017, 04/25/12 to 12/31/99; J91167, 41TNC49XL, 04/27/12 to 03/15/17; J91168, 41FNC41XL, 04/27/12 to 03/15/17; J91179, 02CK017, 04/26/12 to 12/31/99; J91195, 41XCD52S, 04/27/12 to 03/15/17; J9125D, 02CK017, 04/28/12 to 12/31/99; J9128G, 41FNC42XL, 05/02/12 to 03/15/17; J9128H, 41XCC51S, 05/02/12 to 03/15/17; J91306, 02CK017, 05/02/12 to 12/31/99; J9133G, 02CK017, 05/03/12 to 12/31/99; J9140G, 02CK017, 05/07/12 to 12/31/99; J91484, 02CK017, 05/09/12 to 12/31/99; J9167F, 02CK017, 05/17;/12 to 12/31/99; J91685, 41FNC41XL, 05/17/12 to 04/15/17; J9171F, 02CK017, 05/18/12 to 12/31/99; J9175D, 41XCC51S, 05/21/12 to 04/15/17; J9175F, 41KBC17XL, 05/21/12 to 04/15/17; J9180F, 41FNC42XL, 05/23/12 to 04/15/17; J9180G, 41KNC07XL, 05/23/12 to 04/15/17; J9181K, 02CK017, 05/22/12 to 12/31/99; J9183T, 41XCB52S, 05/23/12 to 04/15/17; J9183U, 41FNC43XL, 05/23/12 to 04/15/17; J9184V, 02CK017, 05/23/12 to 12/31/99; J9187R, 41KNC60XL, 05/25/12 to 04/15/17; J9196V, 41FNC43XLD, 05/28/12 to 04/15/17; J9196W, 41XCC51S, 06/15/12 to 04/15/17; J9196X, 41KNC07XL, 05/28/12 to 04/15/17; J9199X, 02CK017, 05/28/12 to 12/31/99; J91A1C, 41KBC17XL, 05/30/12 to 04/15/17; J91A1D, 41FNC42XL, 05/30/12 to 04/15/17; J91A1E, 41FNC43XL, 05/29/12 to 04/15/17; J91A47, 41FNC41XL, 05/30/12 to 04/15/17; J91A57, 41XCB52S, 05/30/12 to 04/15/17; J91A58, 41FNC41XLD, 05/30/12 to 04/15/17; J91A59, 41XCB57S, 05/30/12 to 04/15/17; J91A62, 41XCD52S, 05/30/12 -04/15/17; J91A63, 41TNC49XL, 05/31/12 to 04/15/17; J91A6V, 02CK017, 05/30/12 to 12/31/99; J91A9P, 02CK017, 05/31/12 to 12/31/99; J91C2H, 41KNC07XL, 06/01/12 to 04/15/17; J91C2Z, 41XCC51S, 06/04/12 to 04/15/17; J91C30, 41XCC51S, 06/01/12 to 04/15/17; J91C31, 41FNC42XL, 06/01/12 to 04/15/17; J91C3J, 41FNC43XL, 06/04/12 to 04/15/17; J91C3K, 41KBC17XL, 06/04/12 to 04/15/17; J91C60, 41FNC41XL, 06/05/12 to 04/15/17; J91C6Z, 02CK017, 06/04/12 to 12/31/99; J91D05, 41XCD51S, 06/05/12 to 05/15/17; J91D1D, 02CK017, , 06/05/12 to 12/31/99; J91D3V, 41KNC07XL, 06/06/12 to 05/15/17; J91D7X, 41XCB57S, 06/07/12 to 05/15/17; J91D7Y, 41FNC42XL, 06/07/12 to 05/15/17; J91D7Z, 41KNC07XL, 06/07/12 to 05/15/17; J91D8H, 41XCC51S, 06/08/12 to 05/15/17; J91D8J, 41TNC69XL, 06/07/12 to 05/15/17; J91E4M, 41XCD51S, 06/12/12 to 05/15/17; J91E4N, 41FNC41XLD, 06/12/12 to 05/15/17; J91E9R, 41FNC43XL, 06/13/12 to 05/15/17; J91E9T, 41FNC42XL, 06/12/12 to 05/15/17; J91E9V, 41KNC07XL, 06/13/12 to 05/15/17; J91F23, 02CK017, , 06/12/12 to 12/31/99; J91F45, 41XCB52S, 06/14/12 to 05/15/17; J91F47, 41TNC49XL, 06/14/12 to 05/15/17; J91F77, 41XBB79S, 06/14/12 to 05/15/17; J91F9F, 02CK017, , 06/14/12 to 12/31/99; J91G24, 41XCD51S, 06/14/12 to 05/15/17; J91G2D, 41KNC61XL, 06/20/12 to 05/15/17; J91G2Z, 02CK017, , 06/15/12 to 12/31/99; J91G5K, 41XCD51S, 06/15/12 to 05/15/17; J91G70, 41XCC51S, 06/18/12 to 05/15/17; J91H03, 41FNC42XL, 06/19/12 to 05/15/17; J91H04, 41KNC07XL, 06/21/12 to 05/15/17; J91H05, 41KBC17XL, 06/22/12 to 05/15/17; J91H15, 02CK017, , 06/19/12 to 12/31/99; J91H35, 41FNC41XL, 06/19/12 to 05/15/17; J91H36, 41XCD51S, 06/21/12 to 05/15/17; J91H9N, 02CK017, , 06/21/12 to 12/31/99; J91J39, 41KNC60XL, 06/22/12 to 05/15/17; J91K3F, 41XBB79S, 06/26/12 to 05/15/17; J91K5G, 41XBB79S, 06/26/12 to 05/15/17; J91K6V, 41XCC51S, 06/28/12 to 05/15/17; J91K6Z, 41KNC61XL, 07/04/12 to 05/15/17; J91K71, 41FNC42XL, 06/28/12 to 05/15/17; J91K73, 41XCC50S, 06/28/12 to 05/15/17; J91K75, 41KBC17XL, 06/29/12 to 05/15/17; J91L08, 41KNC61XL, 06/29/12 to 05/15/17; J91L09, 41FNC41XL, 06/29/12 to 05/15/17; J91LOA, 41XCB52S, 06/29/12 to 05/15/17; J91LOD, 41FNC43XL, 07/04/12 to 05/15/17; J91L68, 02CK017, , 06/29/12 to 12/31/99; J91MOH, 02CK017, , 07/04/12 to 12/31/99; J91M5Z, 41XBB79S, 07/05/12 to 06/15/17; J91M62, 41FNC43XLD, 07/05/12 to 06/15/17; J91M63, 41KBC17XL, 07/05/12 to 06/15/17; J91M6U, 02CK017, , 07/05/12 to 12/31/99; J91M9U, 41TNC49XL, 07/06/12 to 06/15/17; J91M9W, 41FNC41XL, 07/06/12 to 06/15/17; J91N3A, 41FNC41XLD, 07/09/12 to 06/15/17; J91N3C, 41XCB57S, 07/09/12 to 06/15/17; J91N3D, 41XCD52S, 07/09/12 to 06/15/17; J91N4K, 02CK017, , 07/09/12 to 12/31/99; J91N9L, 41FNC43XL, 07/10/12 to 06/15/17; J91N9M, 41XCB52S, 07/10/12 to 06/15/17;J91N9N, 41XCD51S, 07/10/12 to 06/15/17; J91P1K, 02CK017, , 07/10/12 to 12/31/99; J91P2X, 41KBC17Xl, 07/11/12 to 06/15/17; J91P2Y, 41XCC50S, 07/11/12 to 06/15/17; J91P2Z, 41XCC51S, 07/12/12 to 06/15/17; J91P47, 02CK017, , 07/11/12 to 12/31/99; J91P79, 41XCB52S, 07/12/12 to 06/15/17; J91P7A, 41FNC41Xl, 07/12/12 to 06/15/17; J91ROH, 41TNC36Xl, 07/13/12 to 06/15/17; J91ROK, 41FNC42Xl, 07/13/12 to 06/15/17; J91R4P, 41XCC53S, 07/17;/12 to 06/15/17; J91R4R, 41XCD51S, 07/17;/12 to 06/15/17; J91R9C, 41TNC49Xl, 07/17;/12 to 06/15/17; J91R9D, 41XCB57S, 07/18/12 to 06/15/17; J91R9F, 41FNC41Xl, 07/17;/12 to 06/15/17; J91T22, to 02CK017, , 07/17;/12 to 12/31/99; J91T42, 41XCD52S, 07/18/12 to 06/15/17; J91T67, 02CK017, , 07/18/12 to 12/31/99; J91T98, 41XCD50S, 07/19/12 to 06/15/17; J91T99, 41FNC42Xl, 07/19/12 to 06/15/17; J91T9A, 41XCC51S, 07/19/12 to 06/15/17; J91U23, 41XCD51S, 07/19/12 to 06/15/17; J91U36, 41KNC60Xl, 07/23/12 to 06/15/17; J91U37, 41KBC17Xl, 07/20/12 to 06/15/17; J91U38, 41XCC51S, 07/23/12 to 06/15/17; J91U46, 41TNC36Xl, 07/20/12 to 06/15/17; J91U7U, 41KNC09Xl, 07/24/12 to 06/15/17; J91U7V, 41XCC53S, 07/24/12 to 06/15/17; J91V2E, 41KNC03Xl, 07/25/12 to 06/15/17; J91V2F, 41XCD51S, 07/25/12 to 06/15/17; J91V30, 02CK017, , 07/24/12 to 12/31/99; J91V60, 41KBC17Xl, 07/25/12 to 06/15/17; J91V6U, 41XCC51S, 07/25/12 to 06/15/17; J91V6Z, 02CK017, , 07/25/12 to 12/31/99; J91W06, 41XCD51S, 07/26/12 to 06/15/17; J91W07, 41KNC60Xl, 07/27/12 to 06/15/17; J91W42, 41FNC43Xl, 07/27/12 to 06/15/17; J91W43, 41FNC41Xl, 07/27/12 to 06/15/17; J91W44, 41XCD52S, 07/27/12 to 06/15/17; J91W45, 41XCD51S, 07/27/12 to 06/15/17; J91W7J, 41XCC51S, 07/31/12 to 06/15/17; J91W7l, 41KBC17Xl, 07/30/12 to 06/15/17; J91W7M, 41KNC03Xl, 08/01/12 to 06/15/17; J91X1l, 02CK017, , 07/31/12 to 12/31/99; J91X3U, 41FNC42Xl, 08/01/12 to 06/15/17; J91X3V, 41FNC43XL, 08/01/12 to 06/15/17; J91X3W, 41FNC41XL, 08/02/12 to 06/15/17; J91X3X, 41FNC41XLD, 08/02/12 to 06/15/17; J91X7K, 41KNC60XL, 08/02/12 to 06/15/17; J91Y10, 41XCD51S, 08/03/12 to 07/15/17; J91Y11, 41XCC50S, 08/06/12 to 07/15/17; J91Y49, 41XCD51S, 08/06/12 to 07/15/17; J91Y5G, 02CK017, , 08/06/12 to 12/31/99; J91Y8A, 41TNC49XL, 08/07/12 to 07/15/17; J91Y8C, 41FNC42XL, 08/07/12 to 07/15/17; J91Y9C, 02CK017, , 08/08/12 to 12/31/99; J91Z1H, 41XCB57S, 08/08/12 to 07/15/17; J91Z1J, 41FNC41XL, 08/08/12 to 07/15/17; J91Z26, 41KBC17XL, 08/08/12 to 07/15/17; J91Z5J, 41FNC43XL, 08/09/12 to 07/15/17; J91Z9E, 41XCD51S, 08/13/12 to 07/15/17; J91Z9F, 41XCC51S, 08/14/12 to 07/15/17; J9206A, 41XCD51S, 08/15/12 to 07/15/17; J92079, 41XCD51S, 08/15/12 to 07/15/17; J9208E, 02CK017, , 08/15/12 to 12/31/99; J9210Y, 41XCC50S, 08/16/12 to 07/15/17; J9210Z, 41XCD50S, 08/16/12 to 07/15/17; J92110, 41XCD51S, 08/16/12 to 07/15/17; J9214G, 41XCC51S, 08/17;/12 to 07/15/17; J9214H, 41FNC41XL, 08/17;/12 to 07/15/17; J9214J, 41FNC42XL, 08/17;/12 to 07/15/17; J9214K, 41XCD51S, 08/17;/12 to 07/15/17; J9219P, 41FNC41XLD, 08/20/12 to 07/15/17; J92239, 41FNC43XL, 08/21/12 to 07/15/17; J9223A, 41KBC17XL, 08/21/12 to 07/15/17; J9223C, 41FNC43XLD, 08/21/12 to 07/15/17; J9223D, 41XCB52S, 08/21/12 to 07/15/17; J92245, 02CK017, , 08/21/12 to 12/31/99; J92273, 41XCD51S, 08/23/12 to 07/15/17; J92274, 41XCC51S, 08/22/12 to 07/15/17; J92275, 41FNC42XL, 08/23/12 to 07/15/17; J9230G, 41XCD51S, 08/24/12 to 07/15/17; J9231U, 02CK017, , 08/23/12 to 12/31/99; J9236P, 02CK017, , 08/24/12 to 12/31/99; J9237F, 41XCD51S, 08/27/12 to 07/15/17; J9243Y, 41FNC43XL, 08/28/12 to 07/15/17; J9243Z, 41FNC41XL, 08/28/12 to 07/15/17; J92440, 41XCC51S, 08/28/12 to 07/15/17; J92441, 41FNC41XLD, 08/28/12 to 07/15/17; J9244Y, 02CK017, , 08/28/12 to 12/31/99; J9247V, 41XCB55S, 08/29/12 to 07/15/17; J9247W, 41KBC17XL, 08/29/12 to 07/15/17; J9247X, 41XCC51S, 08/29/12 to 07/15/17; J9247Y, 41FNC41XL, 08/29/12 to 07/15/17; J9247Z, 41FNC43XL, 08/29/12 to 07/15/17; J9248W, 02CK017, , 08/30/12 to 12/31/99; J92578, 02CK017, , 08/31/12 to 12/31/99; J9260L, 02CK017, , 09/03/12 to 12/31/99; J9262H, 41FNC41XLD, 09/05/12 to 08/15/17; J9262J, 41XCC53S, 09/05/12 to 08/15/17; J9262K, 41FNC42XL, 09/04/12 to 08/15/17; J92639, 41XCC51S, 09/05/12 to 08/15/17; J9263A, 41XCB57S, 09/04/12 to 08/15/17; J92700, 41KBC17XL, 09/06/12 to 08/15/17; J92701, 41FNC43XL, 09/06/12 to 08/15/17; J92702, 41FNC42XL, 09/06/12 to 08/15/17; J9276X, 02CK017, , 09/07/12 to 12/31/99; J9277T, 41XCC51S, 09/10/12 to 08/15/17; J9277V, 41FNC43XL, 09/10/12 to 08/15/17; J9284L, 02CK017, , 09/11/12 to 12/31/99; J9284M, 41KNC60XL, 09/11/12 to 08/15/17; J92866, 41XCC50S, 09/12/12 to 08/15/17; J92868, 41FNC41XL, 09/12/12 to 08/15/17; J9289Y, 41TNC49XL, 09/13/12 to 08/15/17; J9289Z, 41FNC42XL, 09/13/12 to 08/15/17; J92900, 41FNC43XL, 09/13/12 to 08/15/17; J92902, 41XCC51S, 09/14/12 to 08/15/17; J9297F, 41XCC51S, 09/18/12 to 08/15/17; J9297H, 41XCB52S, 09/18/12 to 08/15/17; J92A2C, 41FNC41XL, 09/19/12 to 08/15/17; J92A2D, 41XCB52S, 09/19/12 to 08/15/17; J92A2E, 41XCC51S, 09/19/12 to 08/15/17; J92A2N, 02CK017, , 09/19/12 to 12/31/99; J92A50, 41FNC43XL, 09/20/12 to 08/15/17; J92A51, 41XCC53S, 09/20/12 to 08/15/17; J92A52, 41FNC42XL, 09/20/12 to 08/15/17; J92A53, 41FNC41XLD, 09/20/12 to 08/15/17; J92A7N, 02CK017, , 09/20/12 to 12/31/99; J92C1E, 02CK017, , 09/21/12 to 12/31/99; J92C5E, 41FNC43XLD, 09/27/12 to 08/15/17; J92C75, 02CK017, , 09/25/12 to 12/31/99; J92D7A, 41XCD52S, 09/28/12 to 08/15/17; J92EOO, 02CK017, , 09/28/12 to 12/31/99; J92E15, 41KBC17XL, 10/01/12 to 08/15/17; J92E42, to 02CK017, , 10/01/12 to 12/31/99; J92E5N, 41KNC60XL, 10/02/12 to 08/15/17; J92E5P, 41KNC61XL, 10/02/12 to 08/15/17; J92E5R, 41XCD51S, 10/02/12 to 08/15/17; J92E7J, 02CK017, , 10/02/12 to 12/31/99; J92F04, 41XCB57S, 10/04/12 to 09/15/17; J92F05, 41FNC43XL, 10/03/12 to 09/15/17; J92F06, 41FNC42XL, 10/04/12 to 09/15/17; J92F3G, 41KNC09XL, 10/04/12 to 09/15/17; J92F3H, 41FNC41XLD, 10/04/12 to 09/15/17; J92F73, 41FNC40XL, 10/08/12 to 09/15/17; J92F74, 41XCB57S, 10/08/12 to 09/15/17; J92F76, 41XCD51S, 10/08/12 to 09/15/17; J92GOO, 02CK017, , 10/05/12 to 12/31/99; J92G48, 02CK017, , 10/08/12 to 12/31/99; J92G5T, 41XCC50S, 10/09/12 to 09/15/17; J92G6G, 41KNC60XL, 10/09/12 to 09/15/17; J92G90, 41XCD51S, 10/12/12 to 09/15/17; J92G92, 41TNC49XL, 10/10/12 to 09/15/17; J92H03, 41FNC43XLD, 10/11/12 to 09/15/17; J92H20, 02CK017, , 10/10/12 to 12/31/99; J92H3C, 41XCD52S, 10/10/12 to 09/15/17; J92H3E, 41XCD50S, 10/12/12 to 09/15/17; J92H4E, 42KNC07XL, 10/11/12 to 09/15/17; J92H4Z, 42KNC35XL, 10/11/12 to 09/15/17; J92H50, 42XCB55S, 11/05/12 to 09/15/17; J92H6E, 02CK017, , 10/11/12 to 12/31/99; J92H6L, 41XCD50S, 10/11/12 to 09/15/17; J92H8J, 42XCB51S, 10/15/12 to 09/15/17; J92H8N, 41KBC17XL, 10/15/12 to 09/15/17; J92H8P, 41FNC41XL, 10/12/12 to 09/15/17; J92H8R, 41FNC42XL, 10/15/12 to 09/15/17; J92J2D, 41FNC41XLD, 10/16/12 to 09/15/17; J92J2F, 41TNC49XL, 10/16/12 to 09/15/17; J92J2H, 41FNC43XLD, 10/16/12 to 09/15/17; J92J32, 42KNC03XL, 10/16/12 to 09/15/17; J92J6N, 41XCD50S, 10/16/12 to 09/15/17; J92J6P, 41FNC41XL, 10/17;/12 to 09/15/17; J92J8W, 02CK017, , 10/16/12 to 12/31/99; J92K9A, 42KNC07XL, 11/05/12 to 09/15/17; J92K9C, 42XCB55S, 11/05/12 to 09/15/17; J92L9N, 02CK017, , 10/23/12 to 12/31/99; J92M19, 41XCC53S, 10/25/12 to 09/15/17; J92M4E, 02CK017, , 10/24/12 to 12/31/99; J92M6C, 41FNC41XLD, 10/25/12 to 09/15/17; J92M8K, 02CK017, , 10/25/12 to 12/31/99; J92N4R, 41FNC43XL, 10/30/12 to 09/15/17; J92N7Y, 42KNC35XL, 11/02/12 to 09/15/17; J92N7Z, 42KNC03XL, 11/05/12 to 09/15/17; J92N8Z, 02CK017, , 10/30/12 to 12/31/99; J92R3Y, 02CK017, , 11/02/12 to 12/31/99; J92T1A, 41FNC42XL, 11/06/12 to 10/15/17; J92T1C, 41KBC17XL, 11/06/12 to 10/15/17; J92T1D, 42KNC35XL, 11/06/12 to 10/15/17; J92T3A, 02CK017, , 11/06/12 to 12/31/99; J92T4G, 42KNC07XL, 11/07/12 to 10/15/17; J92T6M, 02CK017, , 11/07/12 to 12/31/99; J92T7E, 41XCC51S, 11/08/12 to 10/15/17; J92T7F, 41XCD52S, 11/08/12 to 10/15/17; J92T7G, 41XCB57S, 11/08/12 to 10/15/17; J92U06, 43TNC20XL, 11/12/12 to 10/15/17; J92U07, 43TNC69XL, 11/12/12 to 10/15/17; J92U26, 41KNC61XL, 11/14/12 to 10/15/17; J92U62, 42XCB55S, 11/14/12 to 10/15/17; J92U63, 42KNC03XL, 11/14/12 to 10/15/17; J92U88, 41KNC60XL, 11/14/12 to 10/15/17; J92V5U, 02CK017, , 11/14/12 to 12/31/99; J92V63, 41XCC51S, 11/15/12 to 10/15/17; J92V64, 41FNC41XLD, 11/15/12 to 10/15/17; J92V7H, 41FNC43XL, 11/14/12 to 10/15/17; J92V7K, 42XCB51S, 11/15/12 to 10/15/17; J92V8M, 43TNC20XL, 11/15/12 to 10/15/17; J92V8N, 43TNC69XL, 11/16/12 to 10/15/17; J92WOT, 02CK017, , 11/15/12 to 12/31/99; J92W2E, 43TNC69XL, 11/21/12 to 10/15/17; J92W2F, 43TNC20XL, 11/16/12 to 10/15/17; J92W2G, 43KNC03XL, 11/16/12 to 10/15/17; J92X07, 42KNC07XL, 11/21/12 to 10/15/17; J92XOA, 41FNC41XL, 11/22/12 to 10/15/17; J92XOC, 41XCC50S, 11/22/12 to 10/15/17; J92XOZ, 41FNC41XL, 11/21/12 to 10/15/17; J92X3P, 02CK017, , 11/22/12 to 12/31/99; J92X8C, 02CK017, , 11/24/12 to 12/31/99; J92X8D, 42KNC35XL, 11/23/12 to 10/15/17; J92X8E, 42XCB55S, 11/23/12

More Code Info

to 10/15/17; J92X9R, 41FNC43XL, 11/23/12 to 10/15/17; J92X9T, 41FNC42XL, 11/23/12 to 10/15/17; J92Y2P, 41KBC17XL, 11/26/12 to 10/15/17; J92Y2R, 41XCC50S, 11/26/12 to 10/15/17; J92Y2T, 41KNC61XL, 11/26/12 to 10/15/17; J92Y2U, 41XCC51S, 11/26/12 to 10/15/17; J92Y6R, 41XCD51S, 11/27/12 to 10/15/17; J92Y71, 41XCC51S, 11/28/12 to 10/15/17; J92ZOJ, 41FNC43XLD, 11/28/12 to 10/15/17; J92ZOK, 41KNC09XL, 11/28/12 to 10/15/17; J92Z14, 41XCD52S, 11/28/12 to 10/15/17; J92Z3Z, 02CK017, , 11/28/12 to 12/31/99; J92Z4R, 41XCB57S, 11/29/12 to 10/15/17; J92Z4T, 42XCB55S, 11/29/12 to 10/15/17; J92Z4U, 41XCC53S, 11/29/12 to 10/15/17; J92Z92, 41FNC41XL, 11/30/12 to 10/15/17; J92Z93, 41XCD51S, 12/01/12 to 10/15/17; J9302F, 41FNC42XL, 12/03/12 to 10/15/17; J9302G, 41KBC17XL, 12/03/12 to 10/15/17; J9302H, 41FNC43XL, 12/03/12 to 10/15/17; J9303L, 41XCC50S, 12/03/12 to 10/15/17; J93105, 41XCD51S, 12/05/12 to 11/15/17; J9313T, 41XCD52S, 12/05/12 to 11/15/17; J9313Z, 41XCC51S, 12/05/12 to 11/15/17; J93140, 41XCD50S, 12/07/12 to 11/15/17; J93168, 41XCD51S, 12/07/12 to 11/15/17; J9316C, 41FNC43XL, 12/10/12 to 11/15/17; J9316D, 41FNC42XL, 12/10/12 to 11/15/17; J9318Z, 02CK017, , 12/07/12 to 12/31/99; J93231, 02CK017, , 12/10/12 to 12/31/99; J9324T, 41KBC17XL, 12/10/12 to 11/15/17; J9324U, 41FNC40XL, 12/10/12 to 11/15/17; J9324V, 41XCC50S, 12/11/12 to 11/15/17; J9325M, 41FNC41XL, 12/11/12 to 11/15/17; K4C03G, 37ER320, 01/04/13 12/15/17; K4C03N, 37ER320, 01/04/13 12/15/17; K4C07P, 37ER320, 01/08/13 12/15/17; K4C09V, 37ER320, 01/09/13 12/15/17; K4C136, 37ER320, 01/10/13 12/15/17; K4C15P, 37ER320, 01/12/13 12/15/17; K4C16E, 37ER320, 01/12/13 12/15/17; K4C185, 37ER320, 01/13/13 12/15/17; K4C18P, 37ER320, 01/14/13 12/15/17; K4C18Y, 37ER320, 01/13/13 12/15/17; K4C19W, 37ER320, 01/15/13 12/15/17; K4C206, 37ER320, 01/16/13 12/15/17; K4C22M, 37ER320, 01/17;/13 12/15/17; K4C26W, 37ER320, 01/18/13 12/15/17; K4C27N, 37ER320, 01/21/13 12/15/17; K4C28T, 37ER320, 01/22/13 12/15/17; K4C293, 37ER320, 01/22/13 12/15/17; K4C31W, 37ER320, 01/22/13 12/15/17; K4C37N, 37ER320, 01/26/13 12/15/17; K4C384, 37ER320, 01/26/13 12/15/17; K4C39J, 37ER320, 01/25/13 12/15/17; K4C412, 37ER320, 01/26/13 12/15/17; K4C41Z, 37ER320, 01/26/13 12/15/17; K4C42N, 37ER320, 01/26/13 12/15/17; K4C504, 37ER320, 02/01/13 12/15/17; K4C51A, 37ER320, 02/02/13 12/15/17; K4C523, 37ER320, 02/02/13 12/15/17; K4C573, 37ER320, 02/06/13 01/15/18; K4C57K, 37ER320, 02/05/13 01/15/18; K4C581, 37ER320, 02/06/13 01/15/18; K4C58G, 37ER320, 02/06/13 01/15/18; K4C591, 37ER320, 02/06/13 01/15/18; K4C63E, 37ER320, 02/08/13 01/15/18; K4C645, 37ER320, 02/09/13 01/15/18; K4C64P, 37ER320, 02/12/13 01/15/18; K4C69N, 37ER320, 02/13/13 01/15/18; K4C709, 37ER320, 02/13/13 01/15/18; K4C71H, 37ER320, 02/13/13 01/15/18; K4C74T, 37ER320, 02/15/13 01/15/18; K4C791, 37ER320, 02/19/13 01/15/18; K4C82W, 37ER320, 02/20/13 01/15/18; K4C833, 37ER320, 02/20/13 01/15/18; K4C848, 37ER320, 02/21/13 01/15/18; K4C88J, 37ER320, 02/23/13 01/15/18; K4C88Y, 37ER320, 02/24/13 01/15/18; K4C89U, 37ER320, 02/24/13 01/15/18; K4C98D, 37ER320, 03/02/13 01/15/18; K4C98K, 37ER320, 03/01/13 01/15/18; K4CA38, 37ER320, 03/11/13 01/15/18; K4CA3V, 37ER320, 03/08/13 02/15/18; K4CC3F, 37ER320, 03/08/13 02/15/18; K4CC3H, 37ER320, 03/09/13 02/15/18; K4CC4R, 37ER320, 03/09/13 02/15/18; K4CC5J, 37ER320, 03/09/13 02/15/18; K4CC5K, 37ER320, 03/11/13 02/15/18; K4CD16, 37ER320, 03/13/13 02/15/18; K4CD1Y, 37ER320, 03/14/13 02/15/18; K4CD24, 37ER320, 03/14/13 02/15/18; K4CD2M, 37ER320, 03/14/13 02/15/18; K4CD3F, 37ER320, 03/15/13 02/15/18; K4CD3Y, 37ER320, 03/15/13 02/15/18; K4CD45, 37ER320, 03/15/13 02/15/18; K4CD7W, 37ER320, 03/18/13 02/15/18; K4CEOW, 37ER320, 03/20/13 02/15/18; K4CE2N, 37ER320, 03/20/13 02/15/18; K4CE9K, 37ER320, 03/23/13 02/15/18; K4CF07, 37ER320, 03/23/13 02/15/18; K4CFOJ, 37ER320, 03/26/13 02/15/18; K4CF43, 37ER320, 03/27/13 02/15/18; K4CF49, 37ER320, 03/27/13 02/15/18; K4CF60, 37ER320, 03/28/13 02/15/18; K4CF72, 37ER320, 03/28/13 02/15/18; K4CF7R, 37ER320, 03/28/13 02/15/18; K4CF8Z, 37ER320, 04/03/13 02/15/18; K4CG2H, 37ER320, 04/04/13 03/15/18; K4CG3H, 37ER320, 04/04/13 03/15/18; K90393, 41TNC49XL, 01/17;/13 12/15/17; K90394, 41FNC43XLD, 01/17;/13 12/15/17; K90395, 41FNC43XL, 01/17;/13 12/15/17; K9042A, 02CK017, , 01/18/13 12/31/99; K90441, 41FNC41XL, 01/21/13 12/15/17; K90442, 41KBC17XL, 01/21/13 12/15/17; K90443, 41FNC41XLD, 01/21/13 12/15/17; K90486, 41XCC51S, 01/22/13 12/15/17; K90487, 41XCD51S, 01/23/13 12/15/17; K90488, 41XCD52S, 01/23/13 12/15/17; K90489, 41XCB57S, 01/23/13 12/15/17; K90519, 02CK017, , 01/23/13 12/31/99; K9057D, 41TNC49XL, 02/01/13 12/15/17; K9059K, 02CK017, , 01/24/13 12/31/99; K9063J, 41FNC41XLD, 01/28/13 12/15/17; K9063K, 41FNC41XL, 01/28/13 12/15/17; K9083C, 41FNC43XL, 02/05/13 12/15/17; K9083D, 41XCC51S, 02/05/13 12/15/17; K9083E, 02CK017, , 02/01/13 12/31/99; K9084H, 41FNC42XL, 02/05/13 12/15/17; K9084L, 41XCB57S, 02/06/13 12/15/17; K9088P, 41FNC40XL 02/05/13 01/15/18; K9088R 41XCD52S 02/06/13 01/15/18; K90894 41XCB52S 02/06/13 01/15/18; K90895 41KBC17XL 02/06/13 01/15/18; K9093F 41XCD51S 02/07/13 01/15/18; K90947 43TNC69XL 02/08/13 01/15/18; K9094P 41FNC43XLD 02/11/13 01/15/18; K9097G 41XCD50S 02/12/13 01/15/18; K9097H 41XCD51S 02/11/13 01/15/18; K90A5Z, 43TNC69XL 02/13/13 01/15/18; K90A9V, 41FNC41XLD 02/13/13 01/15/18; K90A9W, 41FNC41XL, 02/13/13 01/15/18; K90A9X, 41XCC51S, 02/13/13 01/15/18; K90C2J, 02CK017, , 02/14/13 12/31/99; K90C3A, 41FNC43XL, 02/13/13 01/15/18; K90C3K, 41XCB57S, 02/14/13 01/15/18; K90C5C, 41XCD51S, 02/15/13 01/15/18; K90C89, 41XCD51S, 02/16/13 01/15/18; K90C8L, 41FNC43XLD, 02/15/13 01/15/18; K90C8N, 41XCD52S, 02/18/13 01/15/18; K90C9W, 02CK017, , 02/15/13 12/31/99; K90D1X, 41FNC42XL, 02/19/13 01/15/18; K90D1Y, 41KBC17XL, 02/19/13 01/15/18; K90D4U, 02CK01,7 02/18/13 12/31/99; K90D7N, 41XCC50S, 02/19/13 01/15/18; K90E1A, 41XCB57S, 02/20/13 01/15/18; K90E1D, 43TNC69XL, 02/20/13 01/15/18; K90E29, 41XCD50S, 02/20/13 01/15/18; K90E2A, 41XCB52S, 02/20/13 01/15/18; K90E4G, 02CK017, , 02/20/13 12/31/99; K90E5Y, 41XCD51S, 02/21/13 01/15/18; K90F02, 41FNC43XL, 02/22/13 01/15/18; K90F04, 41XCC51S, 02/22/13 01/15/18; K90F30, 02CK017, 02/23/13 to 12/31/99; K90F3X, 41XCD51S 02/22/13 01/15/18; K90F4E, 41XCD51S 02/25/13 01/15/18; K90F4H, 41FNC42XL, 02/25/13 01/15/18; K90F90 41FNC43XL 02/26/13 01/15/18; K90F91 41XCC51S 02/26/13 01/15/18; K90G1G 02CK017, 02/26/13 12/31/99; K90G2V 41FNC43XL 02/27/13 01/15/18; K90G2W 41FNC42XL, 02/27/13 01/15/18; K90G2X 41XCC51S 02/27/13 01/15/18; K90G4X 02CK017, 02/27/13 12/31/99; K90H1P 41KBC17XL 03/04/13 01/15/18; K90H1R 41XCD52S 03/04/13 01/15/18; K90H1T 41XCD51S 03/01/13 01/15/18; K90H5V 41FNC41XLD 03/04/13 01/15/18; K90JOZ 41KNC60XL 03/11/13 02/15/18; K90J10 41KNC61XL 03/07/13 02/15/18; K90J11 41TNC36XL 03/07/13 02/15/18; K90J93, 41XCC53S 03/07/13 02/15/18; K90K3, A 02CK017, 03/08/13 12/31/99; K90K55, 41KNC09XL 03/13/13 02/15/18; K90K56, 41FNC43XLD 03/12/13 02/15/18; K90K57, 41XCD51S 03/12/13 02/15/18; K90L08, 41FNC43XL 03/14/13 02/15/18; K90L09, 41KNC60XL 03/14/13 02/15/18; K90L17, 41FNC41XL 03/13/13 02/15/18; K90L1V, 02CK017, 03/12/13 12/31/99; K90L5A, 41KBC17XL 03/14/13 02/15/18; K90M20, 41XCD51S 03/14/13 02/15/18; K90M22, 41XCC53S 03/22/13 02/15/18; K90M3N, 41XCC50S 03/15/13 02/15/18; K90M3P, 41XCD50S 03/15/13 02/15/18; K90M6M, 41XCC51S 03/18/13 02/15/18; K90M6N, 41FNC43XL 03/18/13 02/15/18; K90M6P, 41XCD51S 03/18/13 02/15/18; K90N1Z, 41TNC49XL 03/19/13 02/15/18; K90N6A, 41XCB52S 03/20/13 02/15/18; K90N6C, 41FNC41XL 03/20/13 02/15/18; K90N9F, 41XCC51S 03/21/13 02/15/18; K90N9G, 41FNC43XL 03/21/13 02/15/18; K90N9H, 41FNC42XL, 03/21/13 02/15/18; K90P3G, 41TNC49XL 03/22/13 02/15/18; K90P3H, 41FNC41XLD 03/22/13 02/15/18; K90P4F, 41KNC61XL 03/22/13 02/15/18; K90R2P, 02CK017, 03/26/13 12/31/99; K90R8R, 41KBC17XL 04/01/13 02/15/18; K90R8T, 41FNC41XLD 04/01/13 02/15/18; K90U1H, 41TNC36XL 04/04/13 03/15/18; and K90U1J, 41KNC09XL 04/04/13 03/15/18.

Distribution

Worldwide Distribution - USA Nationwide to all fifty (50) states, and internationally including the following countries: ARGENTINA, ARGENTINA, BAHRAIN, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CYPRUS, ECUADOR, EGYPT, EL SALVADOR, GREECE, HONDURAS, INDIA, INDONESIA, ISRAEL, JAMAICA, JAPAN, JORDAN, KOREA, REPUBLIC OF, KUWAIT, LEBANON, MEXICO, NETHERLANDS ANTILLES, NICARAGUA, PAKISTAN, PANAMA, PARAGUAY, PERU, PHILIPPINES, PUERTO RICO, QATAR, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, TAIWAN, PROVINCE OF CHINA, THAILAND, TRINIDAD AND TOBAGO, TURKEY, UNITED ARAB EMIRATES, URUGUAY, VENEZUELA and YEMEN.

Quantity

738,356