CRTD COBALT XT HF MRI IS1 DF1, Model Number DTPA2D1; Implantable Cardioverter Defibrillators
Enforcement
- Recall Number
- Z-1774-2023
- Event ID
- 92328
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 28, 2023
- Initiation Date
- May 10, 2023
- Classification Date
- June 16, 2023
- Address
- 8200 Coral Sea St Ne, Mounds View, MN, 55112-4391, United States
Description
CRTD COBALT XT HF MRI IS1 DF1, Model Number DTPA2D1; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
GTIN 00763000178147, Lot Serial numbers: RTH601483S, RTH602021S, RTH602022S, RTH602302S, RTH602303S, RTH602305S, RTH602306S, RTH605169S, RTH602089S, RTH603993S, RTH601526S, RTH600467S, RTH600514S, RTH600533S, RTH600607S, RTH600608S, RTH601387S, RTH601543S, RTH601544S, RTH601545S, RTH602017S, RTH605116S, RTH600520S, RTH602108S, RTH604790S, RTH602149S, RTH605081S, RTH605082S, RTH604360S, RTH604780S, RTH604943S, RTH604734S, RTH600153S, RTH601390S, RTH601395S, RTH601849S, RTH601868S, RTH602192S, RTH602619S, RTH600554S, RTH602035S, RTH602694S, RTH600170S, RTH602206S, RTH601611S, RTH602091S, RTH602195S, RTH603628S, RTH600220S, RTH600223S, RTH600228S, RTH600288S, RTH601447S, RTH601448S, RTH601449S, RTH601450S, RTH602055S, RTH602056S, RTH602057S, RTH602058S, RTH602059S, RTH602060S, RTH602061S, RTH602064S, RTH602065S, RTH602067S, RTH602069S, RTH602071S, RTH602072S, RTH602073S, RTH602074S, RTH602077S, RTH602078S, RTH602079S, RTH602271S, RTH602272S, RTH602276S, RTH602277S, RTH602288S, RTH602625S, RTH602626S, RTH602627S, RTH602628S, RTH602629S, RTH602630S, RTH602631S, RTH602632S, RTH602633S, RTH602634S, RTH602635S, RTH602639S, RTH602640S, RTH602641S, RTH602642S, RTH602643S, RTH602645S, RTH602646S, RTH603959S, RTH603977S, RTH603986S, RTH604454S, RTH604455S, RTH604456S, RTH600592S, RTH601482S, RTH601553S, RTH601555S, RTH601556S, RTH601558S, RTH601563S, RTH601564S, RTH601567S, RTH602618S, RTH601756S, RTH601641S, RTH600460S, RTH601716S, RTH601948S, RTH602517S, RTH600294S, RTH600472S, RTH600475S, RTH601454S, RTH601541S, RTH601633S, RTH601761S, RTH601885S, RTH602042S, RTH602592S, RTH602623S, RTH601518S, RTH602085S, RTH602267S, RTH602686S, RTH605142S, RTH600428S, RTH600449S, RTH601687S, RTH602137S, RTH602144S, RTH602148S, RTH602240S, RTH602260S, RTH601936S, RTH605049S, RTH605053S, RTH600120S, RTH600359S, RTH601663S, RTH601666S, RTH601704S, RTH601793S, RTH600184S, RTH600530S, RTH602053S, RTH602682S, RTH601420S, RTH601422S, RTH602043S, RTH602046S, RTH602048S, RTH602051S, RTH602647S, RTH603950S, RTH603954S, RTH601846S, RTH604174S, RTH604927S, RTH600541S, RTH601765S, RTH602032S, RTH602221S, RTH602676S, RTH600390S, RTH602536S, RTH602201S, RTH600362S, RTH602209S, RTH600434S, RTH600555S, RTH602068S, RTH604529S, RTH600350S, RTH601653S, RTH602583S, RTH605132S, RTH600364S, RTH602019S, RTH602562S, RTH602172S, RTH602198S, RTH602220S, RTH602250S, RTH602251S, RTH602252S, RTH602256S, RTH602257S, RTH602258S, RTH602259S, RTH604476S, RTH604477S, RTH605144S, RTH605145S, RTH600115S, RTH600137S, RTH600175S, RTH600179S, RTH600239S, RTH600266S, RTH600272S, RTH600278S, RTH600281S, RTH600284S, RTH600384S, RTH600397S, RTH600413S, RTH600422S, RTH600480S, RTH600499S, RTH600509S, RTH600511S, RTH600547S, RTH600549S, RTH600565S, RTH600572S, RTH601667S, RTH601827S, RTH601902S, RTH601912S, RTH602541S, RTH602558S, RTH602559S, RTH602561S, RTH602564S, RTH602565S, RTH602593S, RTH602594S, RTH602595S, RTH602597S, RTH602599S, RTH602600S, RTH602601S, RTH602602S, RTH604528S, RTH600461S, RTH600462S, RTH600484S, RTH600503S, RTH600507S, RTH600586S, RTH600587S, RTH601319S, RTH601394S, RTH601615S, RTH602579S, RTH601554S, RTH600136S, RTH602236S, RTH600532S, RTH601333S, RTH602169S, RTH601367S, RTH600198S, RTH600200S, RTH601444S, RTH601783S, RTH600176S, RTH601913S, RTH602086S, RTH602034S, RTH600436S, RTH602566S, RTH600583S, RTH600601S, RTH602537S, RTH601661S, RTH601781S, RTH602268S, RTH602269S, RTH602538S, RTH602539S, RTH602540S, RTH602769S, RTH602772S, RTH602777S, RTH602778S, RTH601542S, RTH601726S, RTH600122S, RTH600123S, RTH600165S, RTH600167S, RTH600173S, RTH600222S, RTH600559S, RTH602159S, RTH600389S, RTH601727S, RTH602557S, RTH600420S, RTH601812S, RTH602223S, RTH600561S, RTH601511S, RTH604506S, RTH604532S, RTH602693S, RTH600544S, RTH600399S, RTH601733S, RTH602691S, RTH603652S, RTH601418S, RTH601582S, RTH601583S, RTH600506S, RTH601735S, RTH602134S, RTH602667S, RTH605047S, RTH601325S, RTH601446S, RTH601877S, RTH605167S, RTH605168S, RTH602203S, RTH600092S, RTH600107S, RTH600164S, RTH600380S, RTH600432S, RTH600452S, RTH600454S, RTH601376S, RTH601380S, RTH601391S, RTH602174S, RTH602214S, RTH602270S, RTH600491S, RTH601725S, RTH601813S, RTH601879S, RTH605051S, RTH602212S, RTH600430S, RTH601806S, RTH601714S, RTH601830S, RTH602168S, RTH602170S, RTH605055S, RTH601707S, RTH601878S, RTH600134S, RTH601486S, RTH605147S, RTH600233S, RTH600236S, RTH601638S, RTH602162S, RTH602210S, RTH600324S, RTH600366S, RTH600560S, RTH601880S, RTH600417S, RTH601710S, RTH600320S, RTH603625S, RTH601396S, RTH601432S, RTH601484S, RTH600406S, RTH600594S, RTH603194S, RTH600264S, RTH600412S, RTH601310S, RTH604804S, RTH600492S, RTH601712S, RTH605112S, RTH602075S, RTH602097S, RTH602191S, RTH600168S, RTH600469S, RTH601502S, RTH601724S, RTH602031S, RTH605162S, RTH600204S, RTH605146S, RTH600203S, RTH605001S, RTH601640S, RTH601766S, RTH601792S, RTH602139S, RTH602189S, RTH602638S, RTH602673S, RTH601844S, RTH604793S, RTH602317S, RTH602664S, RTH604035S, RTH600094S, RTH600095S, RTH600311S, RTH605073S, RTH605080S, RTH600321S, RTH600361S, RTH601428S, RTH601439S, RTH601375S, RTH601577S, RTH601644S, RTH602087S, RTH601856S, RTH601874S, RTH600336S, RTH602522S, RTH601521S, RTH604530S, RTH604531S, RTH600279S, RTH600525S, RTH600588S, RTH602229S, RTH600118S, RTH601673S, RTH600505S, RTH601694S, RTH601790S, RTH601865S, RTH602090S, RTH602227S, RTH602767S, RTH604467S, RTH605239S, RTH603932S, RTH600195S, RTH600196S, RTH600219S, RTH600337S, RTH601575S, RTH601864S, RTH601866S, RTH600443S, RTH600457S, RTH600458S, RTH601522S, RTH601559S, RTH601731S, RTH601815S, RTH601899S, RTH602190S, RTH602590S, RTH602591S, RTH600114S, RTH600400S, RTH601466S, RTH601550S, RTH601721S, RTH601750S, RTH601791S, RTH601810S, RTH602100S, RTH602678S, RTH604767S, RTH605240S, RTH601719S, RTH601437S, RTH603938S, RTH603943S, RTH603945S, RTH603947S, RTH601436S, RTH605131S, RTH604763S, RTH600392S, RTH601588S, RTH601591S, RTH601613S, RTH601829S, RTH601938S, RTH602668S, RTH601574S, RTH601568S, RTH604835S, RTH604636S, RTH601696S, RTH601838S, RTH601839S, RTH602698S, RTH600468S, RTH600540S, RTH604995S, RTH600124S, RTH601451S, RTH601751S, RTH601805S, RTH602230S, RTH600285S, RTH602211S, RTH600510S, RTH601584S, RTH601883S, RTH601745S, RTH601820S, RTH600542S, RTH600433S, RTH601393S, RTH601397S, RTH601412S, RTH604434S, RTH604435S, RTH604436S, RTH604437S, RTH604508S, RTH600437S, RTH601892S, RTH602088S, RTH602183S, RTH601629S, RTH601631S, RTH602030S, RTH602238S, RTH601434S, RTH601740S, RTH604960S, RTH600214S, RTH601400S, RTH601546S, RTH601580S, RTH601671S, RTH601753S, RTH600477S, RTH601430S, RTH600358S, RTH602023S, RTH602581S, RTH600146S, RTH600395S, RTH600494S, RTH600574S, RTH601299S, RTH601304S, RTH601517S, RTH601549S, RTH601560S, RTH601743S, RTH601787S, RTH601797S, RTH602176S, RTH602585S, RTH605069S, RTH600528S, RTH602200S, RTH604877S, RTH601401S, RTH601403S, RTH601504S, RTH601508S, RTH602018S, RTH602699S, RTH602700S, RTH605043S, RTH605241S, RTH601822S, RTH601823S, RTH601907S, RTH601908S, RTH601909S, RTH602029S, RTH602095S, RTH602136S, RTH602141S, RTH602142S, RTH602150S, RTH602185S, RTH602262S, RTH602319S, RTH602555S, RTH602556S, RTH602604S, RTH601585S, RTH602024S, RTH602226S, RTH602570S, RTH601978S, RTH604524S, RTH600351S, RTH605012S, RTH600470S, RTH600476S, RTH601872S, RTH602552S, RTH602577S, RTH600603S, RTH601817S, RTH602076S, RTH602099S, RTH602103S, RTH602532S, RTH602551S, RTH602553S, RTH602703S, RTH600206S, RTH602624S, RTH605076S, RTH601456S, RTH601457S, RTH601459S, RTH601460S, RTH601470S, RTH601475S, RTH601479S, RTH602050S, RTH602054S, RTH602062S, RTH602063S, RTH602231S, RTH602232S, RTH602233S, RTH602234S, RTH602235S, RTH602280S, RTH602282S, RTH602284S, RTH602298S, RTH602301S, RTH601322S, RTH601368S, RTH601651S, RTH601655S, RTH602049S, RTH602218S, RTH602219S, RTH602248S, RTH605157S, RTH600160S, RTH600595S, RTH601801S, RTH600295S, RTH600367S, RTH600482S, RTH601369S, RTH600315S, RTH602205S, RTH600017S, RTH600102S, RTH600292S, RTH600556S, RTH600557S, RTH600329S, RTH601910S, RTH604782S, RTH605039S, RTH602092S, RTH600226S, RTH600478S, RTH600515S, RTH601377S, RTH601589S, RTH600182S, RTH600438S, RTH602587S, RTH600401S, RTH600402S, RTH601744S, RTH601848S, RTH605171S, RTH600259S, RTH600246S, RTH601487S, RTH602094S, RTH605038S, RTH600117S, RTH600446S, RTH601443S, RTH601662S, RTH601842S, RTH601873S, RTH602047S, RTH602070S, RTH602093S, RTH602222S, RTH604998S, RTH602576S, RTH601835S, RTH602160S, RTH602692S, RTH600466S, RTH601616S, RTH600373S, RTH600522S, RTH600330S, RTH601845S, RTH601374S, RTH601379S, RTH601525S, RTH601548S, RTH601746S, RTH602533S, RTH602574S, RTH605028S, RTH604656S, RTH601427S, RTH602548S, RTH604654S, RTH605060S, RTH605062S, RTH600096S, RTH600097S, RTH600111S, RTH600119S, RTH600131S, RTH600151S, RTH600157S, RTH600521S, RTH601706S, RTH601708S, RTH601785S, RTH601861S, RTH601862S, RTH602171S, RTH602550S, RTH605152S, RTH600305S, RTH600346S, RTH601929S, RTH602215S, RTH602578S, RTH602560S, RTH604172S, RTH604994S, RTH603329S, RTH601598S, RTH601762S, RTH601782S, RTH601980S, RTH602261S, RTH602285S, RTH602320S, RTH602339S, RTH602712S, RTH602716S, RTH602726S, RTH602762S, RTH602764S, RTH602765S, RTH602785S, RTH602789S, RTH602790S, RTH602833S, RTH602835S, RTH602836S, RTH602839S, RTH602842S, RTH602844S, RTH602847S, RTH602854S, RTH602880S, RTH602922S, RTH602923S, RTH602925S, RTH602928S, RTH602932S, RTH602933S, RTH602936S, RTH602941S, RTH602963S, RTH602970S, RTH603032S, RTH603049S, RTH603075S, RTH603077S, RTH603078S, RTH603079S, RTH603080S, RTH603081S, RTH603084S, RTH603085S, RTH603086S, RTH603087S, RTH603089S, RTH603090S, RTH603093S, RTH603094S, RTH603095S, RTH603096S, RTH603098S, RTH603099S, RTH603100S, RTH603102S, RTH603103S, RTH603104S, RTH603105S, RTH603106S, RTH603107S, RTH603108S, RTH603109S, RTH603111S, RTH603112S, RTH603115S, RTH603118S, RTH603119S, RTH603121S, RTH603123S, RTH603128S, RTH603129S, RTH603130S, RTH603131S, RTH603133S, RTH603134S, RTH603136S, RTH603137S, RTH603139S, RTH603140S, RTH603148S, RTH603149S, RTH603151S, RTH603152S, RTH603169S, RTH603171S, RTH603276S, RTH603277S, RTH603278S, RTH603280S, RTH603281S, RTH603282S, RTH603283S, RTH603284S, RTH603285S, RTH603304S, RTH603305S, RTH603351S, RTH603403S, RTH603420S, RTH603431S, RTH603432S, RTH603449S, RTH603453S, RTH603469S, RTH603493S, RTH603517S, RTH603518S, RTH603519S, RTH603520S, RTH603521S, RTH603522S, RTH603525S, RTH603527S, RTH603530S, RTH603531S, RTH603532S, RTH603533S, RTH603534S, RTH603535S, RTH603537S, RTH603538S, RTH603539S, RTH603542S, RTH603554S, RTH603562S, RTH603563S, RTH603568S, RTH603569S, RTH603570S, RTH603571S, RTH603572S, RTH603613S, RTH603614S, RTH603615S, RTH603623S, RTH603624S, RTH603626S, RTH603627S, RTH603637S, RTH603639S, RTH603868S, RTH603933S, RTH603957S, RTH603965S, RTH603967S, RTH603972S, RTH603978S, RTH603981S, RTH603988S, RTH604012S, RTH604019S, RTH604020S, RTH604021S, RTH604022S, RTH604023S, RTH604024S, RTH604025S, RTH604026S, RTH604028S, RTH604034S, RTH604037S, RTH604038S, RTH604039S, RTH604041S, RTH604043S, RTH604046S, RTH604048S, RTH604049S, RTH604061S, RTH604066S, RTH604067S, RTH604074S, RTH604080S, RTH604082S, RTH604084S, RTH604085S, RTH604088S, RTH604089S, RTH604091S, RTH604093S, RTH604094S, RTH604095S, RTH604098S, RTH604099S, RTH604100S, RTH604101S, RTH604148S, RTH604149S, RTH604150S, RTH604151S, RTH604152S, RTH604153S, RTH604154S, RTH604155S, RTH604194S, RTH604196S, RTH604312S, RTH604323S, RTH604330S, RTH604332S, RTH604333S, RTH604350S, RTH604366S, RTH604367S, RTH604387S, RTH604393S, RTH604401S, RTH604416S, RTH604417S, RTH604420S, RTH604423S, RTH604424S, RTH604426S, RTH604428S, RTH604430S, RTH604433S, RTH604439S, RTH604448S, RTH604453S, RTH604457S, RTH604462S, RTH604465S, RTH604472S, RTH604478S, RTH604479S, RTH604480S, RTH604481S, RTH604482S, RTH604483S, RTH604484S, RTH604487S, RTH604488S, RTH604489S, RTH604496S, RTH604499S, RTH604502S, RTH604507S, RTH604509S, RTH604510S, RTH604511S, RTH604512S, RTH604515S, RTH604517S, RTH604523S, RTH604525S, RTH604526S, RTH604527S, RTH604533S, RTH604534S, RTH604535S, RTH604536S, RTH604537S, RTH604539S, RTH604543S, RTH604593S, RTH604608S, RTH604609S, RTH604610S, RTH604774S, RTH604794S, RTH604916S, RTH604957S, RTH605052S, RTH605058S, RTH602458S, RTH602653S, RTH602695S, RTH602705S, RTH602709S, RTH602714S, RTH602730S, RTH602734S, RTH602738S, RTH602747S, RTH602748S, RTH602749S, RTH602753S, RTH602756S, RTH602757S, RTH602758S, RTH602760S, RTH602776S, RTH602781S, RTH602782S, RTH602787S, RTH602788S, RTH602791S, RTH602793S, RTH602794S, RTH602795S, RTH602796S, RTH602797S, RTH602798S, RTH602799S, RTH602800S, RTH602802S, RTH602803S, RTH602804S, RTH602805S, RTH602806S, RTH602807S, RTH602809S, RTH602810S, RTH602811S, RTH602812S, RTH602813S, RTH602814S, RTH602815S, RTH602816S, RTH602817S, RTH602818S, RTH602819S, RTH602820S, RTH602821S, RTH602822S, RTH602823S, RTH602824S, RTH602825S, RTH602828S, RTH602829S, RTH602837S, RTH602838S, RTH602856S, RTH602858S, RTH602861S, RTH602862S, RTH602864S, RTH602871S, RTH602872S, RTH602875S, RTH602876S, RTH602877S, RTH602878S, RTH602885S, RTH602889S, RTH602896S, RTH602897S, RTH602900S, RTH602901S, RTH602902S, RTH602905S, RTH602907S, RTH602908S, RTH602917S, RTH602918S, RTH602920S, RTH602921S, RTH602926S, RTH602945S, RTH602946S, RTH602947S, RTH602948S, RTH602949S, RTH602950S, RTH602952S, RTH602955S, RTH602956S, RTH602957S, RTH602958S, RTH602964S, RTH602971S, RTH602976S, RTH602978S, RTH602979S, RTH602980S, RTH602981S, RTH602983S, RTH602985S, RTH602987S, RTH602988S, RTH602991S, RTH602992S, RTH602993S, RTH602994S, RTH602995S, RTH602997S, RTH602998S, RTH603000S, RTH603001S, RTH603002S, RTH603003S, RTH603004S, RTH603005S, RTH603006S, RTH603007S, RTH603013S, RTH603022S, RTH603023S, RTH603030S, RTH603035S, RTH603039S, RTH603041S, RTH603053S, RTH603054S, RTH603055S, RTH603056S, RTH603057S, RTH603059S, RTH603060S, RTH603062S, RTH603063S, RTH603064S, RTH603065S, RTH603066S, RTH603067S, RTH603068S, RTH603069S, RTH603070S, RTH603071S, RTH603072S, RTH603073S, RTH603101S, RTH603110S, RTH603120S, RTH603122S, RTH603125S, RTH603141S, RTH603144S, RTH603145S, RTH603147S, RTH603154S, RTH603158S, RTH603160S, RTH603161S, RTH603164S, RTH603166S, RTH603168S, RTH603226S, RTH603270S, RTH603272S, RTH603273S, RTH603274S, RTH603275S, RTH603279S, RTH603286S, RTH603287S, RTH603288S, RTH603289S, RTH603290S, RTH603291S, RTH603292S, RTH603293S, RTH603294S, RTH603295S, RTH603296S, RTH603297S, RTH603299S, RTH603300S, RTH603301S, RTH603302S, RTH603303S, RTH603307S, RTH603308S, RTH603309S, RTH603310S, RTH603311S, RTH603312S, RTH603313S, RTH603314S, RTH603315S, RTH603316S, RTH603317S, RTH603318S, RTH603319S, RTH603320S, RTH603321S, RTH603322S, RTH603324S, RTH603325S, RTH603326S, RTH603327S, RTH603328S, RTH603330S, RTH603331S, RTH603332S, RTH603333S, RTH603334S, RTH603335S, RTH603336S, RTH603338S, RTH603339S, RTH603340S, RTH603341S, RTH603342S, RTH603343S, RTH603344S, RTH603345S, RTH603346S, RTH603347S, RTH603356S, RTH603359S, RTH603360S, RTH603361S, RTH603362S, RTH603363S, RTH603364S, RTH603365S, RTH603366S, RTH603367S, RTH603369S, RTH603370S, RTH603371S, RTH603372S, RTH603373S, RTH603374S, RTH603375S, RTH603376S, RTH603377S, RTH603378S, RTH603379S, RTH603380S, RTH603381S, RTH603382S, RTH603383S, RTH603384S, RTH603385S, RTH603386S, RTH603387S, RTH603388S, RTH603389S, RTH603390S, RTH603391S, RTH603392S, RTH603393S, RTH603394S, RTH603395S, RTH603396S, RTH603397S, RTH603398S, RTH603399S, RTH603402S, RTH603404S, RTH603405S, RTH603406S, RTH603407S, RTH603408S, RTH603409S, RTH603412S, RTH603414S, RTH603417S, RTH603419S, RTH603422S, RTH603424S, RTH603425S, RTH603426S, RTH603427S, RTH603428S, RTH603429S, RTH603430S, RTH603433S, RTH603434S, RTH603452S, RTH603460S, RTH603461S, RTH603462S, RTH603463S, RTH603464S, RTH603465S, RTH603466S, RTH603467S, RTH603468S, RTH603470S, RTH603471S, RTH603478S, RTH603481S, RTH603482S, RTH603486S, RTH603488S, RTH603489S, RTH603490S, RTH603491S, RTH603492S, RTH603494S, RTH603495S, RTH603497S, RTH603499S, RTH603500S, RTH603501S, RTH603502S, RTH603503S, RTH603504S, RTH603505S, RTH603506S, RTH603507S, RTH603549S, RTH603550S, RTH603551S, RTH603553S, RTH603555S, RTH603556S, RTH603558S, RTH603559S, RTH603560S, RTH603561S, RTH603564S, RTH603565S, RTH603566S, RTH603567S, RTH603573S, RTH603574S, RTH603575S, RTH603576S, RTH603577S, RTH603578S, RTH603608S, RTH603609S, RTH603610S, RTH603611S, RTH603612S, RTH603616S, RTH603617S, RTH603618S, RTH603619S, RTH603620S, RTH603621S, RTH603622S, RTH603629S, RTH603630S, RTH603631S, RTH603632S, RTH603633S, RTH603634S, RTH603640S, RTH603641S, RTH603642S, RTH603643S, RTH603644S, RTH603646S, RTH603647S, RTH603648S, RTH603649S, RTH603650S, RTH603651S, RTH603655S, RTH603656S, RTH603657S, RTH603658S, RTH603659S, RTH603660S, RTH603661S, RTH603662S, RTH603663S, RTH603664S, RTH603665S, RTH603666S, RTH603667S, RTH603668S, RTH603670S, RTH603671S, RTH603672S, RTH603673S, RTH603674S, RTH603675S, RTH603677S, RTH603678S, RTH603679S, RTH603680S, RTH603681S, RTH603682S, RTH603683S, RTH603684S, RTH603685S, RTH603686S, RTH603687S, RTH603688S, RTH603689S, RTH603690S, RTH603691S, RTH603693S, RTH603694S, RTH603695S, RTH603696S, RTH603697S, RTH603699S, RTH603701S, RTH603702S, RTH603703S, RTH603705S, RTH603706S, RTH603707S, RTH603708S, RTH603709S, RTH603711S, RTH603712S, RTH603713S, RTH603714S, RTH603715S, RTH603716S, RTH603717S, RTH603718S, RTH603719S, RTH603720S, RTH603721S, RTH603722S, RTH603723S, RTH603724S, RTH603725S, RTH603726S, RTH603728S, RTH603729S, RTH603730S, RTH603731S, RTH603732S, RTH603735S, RTH603736S, RTH603737S, RTH603739S, RTH603740S, RTH603741S, RTH603742S, RTH603743S, RTH603744S, RTH603745S, RTH603746S, RTH603747S, RTH603748S, RTH603749S, RTH603750S, RTH603752S, RTH603754S, RTH603755S, RTH603756S, RTH603804S, RTH603809S, RTH603857S, RTH603859S, RTH603860S, RTH603862S, RTH603863S, RTH603928S, RTH603929S, RTH603948S, RTH603992S, RTH604016S, RTH604029S, RTH604030S, RTH604031S, RTH604032S, RTH604102S, RTH604104S, RTH604105S, RTH604106S, RTH604107S, RTH604108S, RTH604109S, RTH604110S, RTH604111S, RTH604112S, RTH604113S, RTH604114S, RTH604115S, RTH604116S, RTH604117S, RTH604118S, RTH604119S, RTH604120S, RTH604122S, RTH604123S, RTH604124S, RTH604125S, RTH604126S, RTH604127S, RTH604128S, RTH604129S, RTH604131S, RTH604132S, RTH604133S, RTH604134S, RTH604135S, RTH604136S, RTH604137S, RTH604138S, RTH604139S, RTH604140S, RTH604141S, RTH604142S, RTH604144S, RTH604146S, RTH604147S, RTH604156S, RTH604157S, RTH604158S, RTH604159S, RTH604160S, RTH604161S, RTH604163S, RTH604165S, RTH604166S, RTH604167S, RTH604168S, RTH604170S, RTH604171S, RTH604173S, RTH604175S, RTH604188S, RTH604197S, RTH604198S, RTH604200S, RTH604202S, RTH604204S, RTH604205S, RTH604206S, RTH604207S, RTH604208S, RTH604209S, RTH604210S, RTH604211S, RTH604212S, RTH604214S, RTH604215S, RTH604216S, RTH604243S, RTH604245S, RTH604294S, RTH604295S, RTH604296S, RTH604297S, RTH604298S, RTH604299S, RTH604300S, RTH604301S, RTH604302S, RTH604303S, RTH604304S, RTH604305S, RTH604306S, RTH604307S, RTH604308S, RTH604309S, RTH604310S, RTH604311S, RTH604313S, RTH604314S, RTH604315S, RTH604317S, RTH604318S, RTH604319S, RTH604320S, RTH604321S, RTH604322S, RTH604325S, RTH604334S, RTH604337S, RTH604338S, RTH604341S, RTH604342S, RTH604343S, RTH604345S, RTH604346S, RTH604349S, RTH604351S, RTH604352S, RTH604353S, RTH604354S, RTH604355S, RTH604358S, RTH604359S, RTH604361S, RTH604362S, RTH604369S, RTH604370S, RTH604371S, RTH604372S, RTH604373S, RTH604374S, RTH604375S, RTH604376S, RTH604377S, RTH604378S, RTH604379S, RTH604381S, RTH604383S, RTH604384S, RTH604385S, RTH604390S, RTH604391S, RTH604396S, RTH604397S, RTH604398S, RTH604402S, RTH604403S, RTH604408S, RTH604409S, RTH604410S, RTH604411S, RTH604412S, RTH604413S, RTH604421S, RTH604422S, RTH604425S, RTH604431S, RTH604444S, RTH604445S, RTH604446S, RTH604447S, RTH604449S, RTH604450S, RTH604451S, RTH604452S, RTH604458S, RTH604459S, RTH604460S, RTH604461S, RTH604468S, RTH604469S, RTH604473S, RTH604474S, RTH604475S, RTH604485S, RTH604486S, RTH604490S, RTH604491S, RTH604492S, RTH604493S, RTH604494S, RTH604495S, RTH604497S, RTH604498S, RTH604503S, RTH604504S, RTH604505S, RTH604558S, RTH604559S, RTH604560S, RTH604594S, RTH604596S, RTH604635S, RTH604647S, RTH604648S, RTH604649S, RTH604652S, RTH604653S, RTH604657S, RTH604662S, RTH604663S, RTH604686S, RTH604693S, RTH604699S, RTH604707S, RTH604709S, RTH604716S, RTH604726S, RTH604727S, RTH604733S, RTH604740S, RTH604741S, RTH604768S, RTH604770S, RTH604776S, RTH604779S, RTH604783S, RTH604797S, RTH604807S, RTH604824S, RTH604846S, RTH604849S, RTH604853S, RTH604854S, RTH604855S, RTH604862S, RTH604864S, RTH604871S, RTH604874S, RTH604885S, RTH604893S, RTH604901S, RTH604903S, RTH604905S, RTH604906S, RTH604907S, RTH604909S, RTH604910S, RTH604911S, RTH604912S, RTH604913S, RTH604914S, RTH604917S, RTH604921S, RTH604924S, RTH604926S, RTH604930S, RTH604931S, RTH604940S, RTH604941S, RTH604945S, RTH604950S, RTH604951S, RTH604958S, RTH604963S, RTH604964S, RTH604965S, RTH604966S, RTH604969S, RTH604977S, RTH604978S, RTH604981S, RTH604986S, RTH604987S, RTH604988S, RTH604996S, RTH604997S, RTH605002S, RTH605003S, RTH605014S, RTH605015S, RTH605020S, RTH605023S, RTH605026S, RTH605029S, RTH605035S, RTH605042S, RTH605044S, RTH605045S, RTH605063S, RTH605070S, RTH605074S, RTH605101S, RTH605104S, RTH605127S, RTH605170S, RTH600347S, RTH600133S, RTH600265S, RTH600372S, RTH600408S, RTH601534S, RTH601686S, RTH605064S, RTH605114S, RTH601749S, RTH600189S, RTH604065S, RTH604516S, RTH600139S, RTH600141S, RTH601370S, RTH601538S, RTH601425S, RTH601445S, RTH601688S, RTH601695S, RTH600194S, RTH601586S, RTH602181S, RTH602527S, RTH600191S, RTH600385S, RTH601713S, RTH601965S, RTH601934S, RTH604936S, RTH605013S, RTH600227S, RTH601679S, RTH604162S, RTH600377S, RTH600504S, RTH600576S, RTH600493S, RTH601569S, RTH601800S, RTH600371S, RTH604850S, RTH601410S, RTH601715S, RTH601798S, RTH602696S, RTH600291S, RTH601656S, RTH601657S, RTH604801S, RTH604800S, RTH600573S, RTH601858S, RTH602672S, RTH601732S, RTH601897S, RTH601905S, RTH602083S, RTH600582S, RTH600545S, RTH601622S, RTH602621S, RTH600126S, RTH600128S, RTH600474S, RTH601573S, RTH601675S, RTH601863S, RTH602571S, RTH600142S, RTH601780S, RTH601896S, RTH601949S, RTH602027S, RTH602028S, RTH602603S, RTH601364S, RTH601419S, RTH600302S, RTH601440S, RTH601677S, RTH601952S, RTH600562S, RTH601378S, RTH601402S, RTH601416S, RTH601701S, RTH601738S, RTH601852S, RTH602207S, RTH605172S, RTH600104S, RTH600174S, RTH600177S, RTH600178S, RTH600186S, RTH600197S, RTH600207S, RTH600237S, RTH600376S, RTH600456S, RTH600501S, RTH600534S, RTH601431S, RTH601453S, RTH602081S, RTH602164S, RTH602165S, RTH602175S, RTH602193S, RTH602217S, RTH602255S, RTH602580S, RTH602620S, RTH602701S, RTH600421S, RTH600429S, RTH600563S, RTH600564S, RTH600580S, RTH601405S, RTH601406S, RTH602020S, RTH600352S, RTH600496S, RTH600585S, RTH601759S, RTH602609S, RTH600387S, RTH604637S, RTH604638S, RTH604645S, RTH604646S, RTH605068S, RTH605267S, RTH602278S, RTH602279S, RTH604036S, RTH604554S, RTH604555S, RTH604556S, RTH604557S, RTH601514S, RTH601799S, RTH602026S, RTH602677S, RTH601802S, RTH601895S, RTH605134S, RTH600193S, RTH600495S, RTH601323S, RTH601398S, RTH601426S, RTH601533S, RTH601551S, RTH601702S, RTH601728S, RTH601809S, RTH602066S, RTH602182S, RTH604961S, RTH605133S, RTH601789S, RTH601442S, RTH601764S, RTH602572S, RTH601951S, RTH600334S, RTH600516S, RTH602588S, RTH601718S, RTH600488S, RTH600569S, RTH600578S, RTH600579S, RTH601658S, RTH601664S, RTH600201S, RTH601649S, RTH601888S, RTH600332S, RTH601438S, RTH601784S, RTH601760S, RTH601942S, RTH601832S, RTH601811S, RTH601904S, RTH601916S, RTH600156S, RTH600523S, RTH602025S, RTH604956S, RTH605083S, RTH600464S, RTH600497S, RTH600439S, RTH600453S, RTH600465S, RTH601736S, RTH600225S, RTH604954S, RTH600322S, RTH601915S, RTH604470S, RTH604471S, RTH600138S, RTH601579S, RTH602208S, RTH600150S, RTH600354S, RTH600355S, RTH601652S, RTH601755S, RTH601944S, RTH605164S, RTH602224S, RTH604399S, RTH601565S, RTH600210S, RTH600213S, RTH600312S, RTH600339S, RTH600345S, RTH600388S, RTH600486S, RTH600546S, RTH600558S, RTH601350S, RTH601570S, RTH602084S, RTH602180S, RTH602216S, RTH602549S, RTH602704S, RTH600393S, RTH600551S, RTH601409S, RTH601413S, RTH601414S, RTH601821S, RTH602143S, RTH602146S, RTH602263S, RTH602264S, RTH604363S, RTH604442S, RTH604443S, RTH605209S, RTH605210S, RTH605211S, RTH601331S, RTH601758S, RTH600110S, RTH600144S, RTH600230S, RTH601353S, RTH601619S, RTH601906S, RTH602204S, RTH602575S, RTH600101S, RTH600135S, RTH600147S, RTH601366S, RTH601670S, RTH602156S, RTH602202S, RTH600448S, RTH600543S, RTH602237S, RTH603874S, RTH604033S, RTH604575S, RTH604576S, RTH604577S, RTH604578S, RTH604579S, RTH604580S, RTH604581S, RTH604582S, RTH604583S, RTH604584S, RTH605219S, RTH605220S, RTH605221S, RTH605222S, RTH605223S, RTH605224S, RTH605226S, RTH605227S, RTH605229S, RTH605231S, RTH605234S, RTH605236S, RTH602702S, RTH604521S, RTH604522S, RTH602052S, RTH600208S, RTH601316S, RTH600163S, RTH601415S, RTH601875S, RTH601911S, RTH602178S, RTH600318S, RTH600568S, RTH601981S, RTH602239S, RTH600286S, RTH600365S, RTH600414S, RTH601408S, RTH601734S, RTH601833S, RTH600103S, RTH600338S, RTH600411S, RTH602179S, RTH602584S, RTH600450S, RTH601825S, RTH602225S, RTH602679S, RTH602598S, RTH604368S, RTH605266S, RTH600161S, RTH601480S, RTH600235S, RTH601455S, RTH602199S, RTH603998S, RTH604000S, RTH604006S, RTH604007S, RTH604009S, RTH601340S, RTH601342S, RTH601343S, RTH601561S, RTH601562S, RTH601566S, RTH601741S, RTH602606S, RTH600567S, RTH601429S, RTH601711S, RTH602101S, RTH602567S, RTH602582S, RTH602138S, RTH604639S, RTH604640S, RTH604641S, RTH604642S, RTH604643S, RTH604644S, RTH601313S, RTH602589S, RTH604335S, RTH604336S, RTH604597S, RTH604598S, RTH604599S, RTH604600S, RTH604601S, RTH604602S, RTH604604S, RTH604605S, RTH604606S, RTH604607S, RTH603153S, RTH603353S, RTH603355S, RTH603540S, RTH603541S, RTH603543S, RTH603544S, RTH603545S, RTH603546S, RTH604051S, RTH604052S, RTH604053S, RTH604055S, RTH604057S, RTH604060S, RTH604063S, RTH604064S, RTH604611S, RTH604612S, RTH604613S, RTH604614S, RTH604615S, RTH604616S, RTH604617S, RTH604618S, RTH604619S, RTH604620S, RTH604621S, RTH604622S, RTH604623S, RTH604624S, RTH604626S, RTH604392S, RTH604394S, RTH604395S, RTH604595S, RTH604811S, RTH604429S, RTH601590S, RTH602568S, RTH604571S, RTH604572S, RTH604573S, RTH604574S, RTH600590S, RTH602147S, RTH600404S, RTH601742S, RTH602197S, RTH604389S, RTH604440S, RTH604441S, RTH601814S, RTH602687S, RTH602666S, RTH604513S, RTH604514S, RTH601386S, RTH601388S, RTH601421S, RTH601520S, RTH601795S, RTH601803S, RTH601804S, RTH601854S, RTH601859S, RTH602196S, RTH603306S, RTH603523S, RTH603526S, RTH603733S, RTH603734S, RTH604326S, RTH604327S, RTH604328S, RTH604329S, RTH604331S, RTH604628S, RTH604629S, RTH604630S, RTH604631S, RTH604632S, RTH604633S, RTH604634S, RTH601826S, RTH600327S, RTH601788S, RTH601828S, RTH601399S, RTH600100S, RTH601317S, RTH601318S, RTH601510S, RTH605005S, RTH601620S, RTH601383S, RTH600105S, RTH600212S, RTH600455S, RTH601441S, RTH602080S, RTH602228S, RTH601423S, RTH602167S, RTH600190S, RTH600199S, RTH601404S, RTH602188S, RTH604967S, RTH600192S, RTH601433S, RTH601578S, RTH600202S, RTH601703S, RTH601843S, RTH601485S, RTH601876S, RTH600316S, RTH600317S, RTH600410S, RTH601614S, RTH602213S, RTH600369S, RTH605124S, RTH604847S, RTH600517S, RTH600419S, RTH601665S, RTH602033S, RTH600512S, RTH601337S, RTH601851S, RTH601857S, RTH602102S, RTH604130S, RTH602111S, RTH602586S, RTH600262S, RTH600263S, RTH601357S, RTH601730S, RTH601757S, RTH601950S, RTH601954S, RTH601321S, RTH600162S, RTH605165S, RTH600254S, RTH600255S, RTH600368S, RTH600529S, RTH601424S, RTH601763S, RTH601807S, RTH601819S, RTH601853S, RTH601871S, RTH602596S, RTH601587S, RTH604627S, RTH600479S, RTH601729S, RTH600447S, RTH600524S, RTH601571S, RTH602096S, RTH600577S, RTH601959S, RTH601796S, RTH601818S, RTH601836S, RTH600231S, RTH600232S, RTH600341S, RTH600356S, RTH600357S, RTH601576S, RTH601618S, RTH602535S, RTH600308S, RTH600379S, RTH600280S, RTH600409S, RTH602177S, RTH600180S, RTH600490S, RTH600485S, RTH600188S, RTH600244S, RTH600489S, RTH605213S, RTH605214S, RTH605217S, RTH604518S, RTH604519S, RTH604520S, RTH604585S, RTH604586S, RTH604587S, RTH604588S, RTH604589S, RTH604590S, RTH604591S, RTH604592S, RTH602919S, RTH601365S, RTH604027S, RTH604386S, RTH604388S, RTH604500S, RTH604501S, RTH604742S, RTH604744S, RTH604745S, RTH604746S, RTH602554S, RTH600342S, RTH604561S, RTH604562S, RTH604564S, RTH604565S, RTH604566S, RTH604567S, RTH604568S, RTH604569S, RTH600596S, RTH600597S, RTH600598S, RTH600599S, RTH600600S, RTH601294S, RTH601295S, RTH601296S, RTH601297S, RTH601300S, RTH601301S, RTH601303S, RTH601306S, RTH601307S, RTH601617S, RTH601623S, RTH601624S, RTH601625S, RTH601626S, RTH601627S, RTH601632S, RTH601634S, RTH601635S, RTH601636S, RTH601637S, RTH601642S, RTH601643S, RTH601646S, RTH601647S, RTH601648S, RTH601669S, RTH601889S, RTH601890S, RTH601891S, RTH601894S, RTH601975S, RTH602005S, RTH602014S, RTH602015S, RTH602016S, RTH602036S, RTH602037S, RTH602038S, RTH602040S, RTH602041S, RTH602104S, RTH602105S, RTH602106S, RTH602107S, RTH602109S, RTH602110S, RTH602112S, RTH602133S, RTH602151S, RTH602152S, RTH602153S, RTH602154S, RTH602155S, RTH602243S, RTH602244S, RTH602245S, RTH602246S, RTH602247S, RTH602249S, RTH602266S, RTH602510S, RTH602515S, RTH602520S, RTH602523S, RTH602524S, RTH602525S, RTH602526S, RTH602528S, RTH602529S, RTH602530S, RTH602531S, RTH602545S, RTH602546S, RTH602547S, RTH602607S, RTH602608S, RTH602611S, RTH602613S, RTH602614S, RTH602615S, RTH602648S, RTH602649S, RTH602650S, RTH602651S, RTH602652S, RTH602654S, RTH602655S, RTH602656S, RTH602657S, RTH602658S, RTH602659S, RTH602660S, RTH602661S, RTH602662S, RTH602663S, RTH602688S, RTH602689S, RTH602690S, RTH603076S, RTH603082S, RTH603528S, RTH603529S, RTH603536S, RTH604087S, RTH604092S, RTH604096S, RTH604097S, RTH604176S, RTH604187S, RTH604189S, RTH604190S, RTH604191S, RTH604193S, RTH604218S, RTH604220S, RTH604221S, RTH604222S, RTH604239S, RTH604365S, RTH604414S, RTH604415S, RTH604418S, RTH604419S, RTH604463S, RTH604464S, RTH604466S, RTH604540S, RTH604541S, RTH604542S, RTH604544S, RTH604545S, RTH604546S, RTH604547S, RTH604548S, RTH604549S, RTH604550S, RTH604551S, RTH604552S, RTH604553S, RTH605158S, RTH605159S, RTH605160S, RTH605161S, RTH601786S, RTH604772S, RTH600340S, RTH601344S, RTH602186S, RTH602187S, RTH602241S, RTH602242S, RTH602321S, RTH602671S, RTH603508S, RTH603509S, RTH603510S, RTH603511S, RTH603514S, RTH603515S, RTH603516S, RTH602826S, RTH602827S, RTH602830S, RTH602832S, RTH601309S, RTH600431S, RTH600360S, RTH604400S, RTH604538S, RTH600221S, RTH600229S, RTH600241S, RTH600247S, RTH600296S, RTH600297S, RTH600306S, RTH600307S, RTH600313S, RTH601767S, RTH601768S, RTH601769S, RTH601770S, RTH601771S, RTH601772S, RTH601774S, RTH601776S, RTH601777S, RTH601778S, RTH601779S; GTIN 00763000178161, Lot Serial Numbers: RTH600974S, RTH601847S, RTH604279S, RTH604723S, RTH604947S, RTH603847S, RTH601928S, RTH605331S, RTH605120S, RTH603234S, RTH605362S, RTH603799S, RTH604680S, RTH603976S, RTH603183S, RTH605111S, RTH605293S, RTH605354S, RTH602743S, RTH603964S, RTH604813S, RTH604754S, RTH602314S, RTH601930S, RTH603999S, RTH603815S, RTH600036S, RTH604758S, RTH601468S, RTH603245S, RTH600083S, RTH601285S, RTH602334S, RTH600328S, RTH601015S, RTH601191S, RTH601359S, RTH602322S, RTH601172S, RTH601869S, RTH603235S, RTH601013S, RTH601073S, RTH601074S, RTH602124S, RTH602323S, RTH602331S, RTH601068S, RTH601189S, RTH601472S, RTH601489S, RTH601604S, RTH601887S, RTH602286S, RTH602337S, RTH602462S, RTH603196S, RTH603773S, RTH603774S, RTH603951S, RTH603990S, RTH604277S, RTH600071S, RTH600072S, RTH601069S, RTH601292S, RTH601994S, RTH602010S, RTH602328S, RTH602330S, RTH602335S, RTH602403S, RTH603930S, RTH601593S, RTH600061S, RTH600062S, RTH600063S, RTH601048S, RTH601867S, RTH601898S, RTH601924S, RTH600695S, RTH601194S, RTH601995S, RTH600046S, RTH600575S, RTH600750S, RTH601012S, RTH601076S, RTH601850S, RTH602333S, RTH602486S, RTH602729S, RTH603995S, RTH600539S, RTH600080S, RTH601008S, RTH601291S, RTH601122S, RTH604812S, RTH605351S, RTH603775S, RTH604014S, RTH603176S, RTH603052S, RTH602318S, RTH603496S, RTH601171S, RTH602935S, RTH600035S, RTH600041S, RTH603191S, RTH603231S, RTH603020S, RTH603019S, RTH603979S, RTH603817S, RTH601932S, RTH602132S, RTH601903S, RTH602356S, RTH604201S, RTH605323S, RTH605352S, RTH601195S, RTH601672S, RTH602296S, RTH600892S, RTH602355S, RTH602421S, RTH600044S, RTH601183S, RTH602423S, RTH602431S, RTH601933S, RTH601963S, RTH601966S, RTH601977S, RTH601992S, RTH602474S, RTH602475S, RTH603255S, RTH603261S, RTH600971S, RTH601692S, RTH601882S, RTH602338S, RTH602424S, RTH602471S, RTH603199S, RTH603237S, RTH6023
54S, RTH602428S, RTH602725S, RTH602727S, RTH602962S, RTH603017S, RTH603142S, RTH603181S, RTH604237S, RTH604698S, RTH601096S, RTH601200S, RTH601277S, RTH601280S, RTH601503S, RTH601953S, RTH602461S, RTH603036S, RTH603241S, RTH604003S, RTH601038S, RTH601046S, RTH601946S, RTH602125S, RTH603058S, RTH602440S, RTH600058S, RTH603192S, RTH601278S, RTH601498S, RTH601499S, RTH601698S, RTH601941S, RTH602610S, RTH603202S, RTH605056S, RTH604199S, RTH600013S, RTH600014S, RTH600015S, RTH601463S, RTH601495S, RTH601960S, RTH602299S, RTH603213S, RTH601493S, RTH602674S, RTH603177S, RTH603193S, RTH601290S, RTH601501S, RTH601668S, RTH601775S, RTH601945S, RTH601972S, RTH602307S, RTH602959S, RTH603512S, RTH603822S, RTH603864S, RTH603939S, RTH602120S, RTH601115S, RTH600031S, RTH600042S, RTH600442S, RTH601116S, RTH601328S, RTH601971S, RTH603217S, RTH603178S, RTH601124S, RTH601284S, RTH600806S, RTH601488S, RTH602416S, RTH602427S, RTH600086S, RTH600087S, RTH601179S, RTH601709S, RTH602003S, RTH602012S, RTH604226S, RTH601984S, RTH601985S, RTH600016S, RTH600299S, RTH601041S, RTH601043S, RTH601059S, RTH601065S, RTH601314S, RTH601682S, RTH601700S, RTH601947S, RTH602430S, RTH602723S, RTH602841S, RTH602972S, RTH603238S, RTH603760S, RTH603856S, RTH603961S, RTH600030S, RTH600775S, RTH600845S, RTH600917S, RTH600929S, RTH602400S, RTH602477S, RTH603984S, RTH602253S, RTH602378S, RTH602404S, RTH603018S, RTH604755S, RTH600085S, RTH600571S, RTH601075S, RTH602425S, RTH603204S, RTH604264S, RTH604275S, RTH602890S, RTH602834S, RTH600759S, RTH601188S, RTH603143S, RTH605290S, RTH600994S, RTH602465S, RTH600763S, RTH601870S, RTH602490S, RTH602622S, RTH603247S, RTH602383S, RTH602414S, RTH603205S, RTH603215S, RTH601345S, RTH600915S, RTH601931S, RTH601600S, RTH600774S, RTH601094S, RTH602001S, RTH602398S, RTH602684S, RTH602741S, RTH604044S, RTH604086S, RTH602438S, RTH600861S, RTH601026S, RTH603778S, RTH603186S, RTH605092S, RTH605338S, RTH601286S, RTH604886S, RTH600932S, RTH602000S, RTH602292S, RTH602459S, RTH604232S, RTH605119S, RTH605306S, RTH601722S, RTH602605S, RTH600866S, RTH602373S, RTH601112S, RTH601062S, RTH600847S, RTH600865S, RTH600946S, RTH601477S, RTH601983S, RTH602464S, RTH603244S, RTH603758S, RTH605110S, RTH602290S, RTH603135S, RTH603156S, RTH603214S, RTH602329S, RTH600886S, RTH600978S, RTH601140S, RTH602384S, RTH602437S, RTH602735S, RTH601919S, RTH601927S, RTH603197S, RTH601184S, RTH603239S, RTH601349S, RTH601351S, RTH601491S, RTH602287S, RTH602422S, RTH602736S, RTH603216S, RTH602336S, RTH604004S, RTH601190S, RTH600967S, RTH601135S, RTH601467S, RTH601203S, RTH602480S, RTH600914S, RTH601121S, RTH603170S, RTH603974S, RTH603200S, RTH600790S, RTH601281S, RTH603771S, RTH602315S, RTH602316S, RTH602324S, RTH602385S, RTH603146S, RTH603987S, RTH602118S, RTH601592S, RTH601594S, RTH604017S, RTH604759S, RTH603223S, RTH603770S, RTH603776S, RTH604230S, RTH600066S, RTH600733S, RTH602006S, RTH602359S, RTH602407S, RTH604739S, RTH603155S, RTH601474S, RTH601935S, RTH603766S, RTH604013S, RTH600531S, RTH600762S, RTH601690S, RTH602481S, RTH600986S, RTH600984S, RTH601006S, RTH601197S, RTH601005S, RTH601955S, RTH600907S, RTH601111S, RTH600771S, RTH600943S, RTH601071S, RTH600374S, RTH600871S, RTH602184S, RTH600059S, RTH600745S, RTH600970S, RTH601036S, RTH601169S, RTH601199S, RTH601923S, RTH602485S, RTH603638S, RTH603780S, RTH601107S, RTH605100S, RTH603165S, RTH601066S, RTH601137S, RTH600968S, RTH600975S, RTH601007S, RTH601040S, RTH601336S, RTH601384S, RTH601609S, RTH601660S, RTH601834S, RTH603126S, RTH603172S, RTH603794S, RTH603835S, RTH604213S, RTH604247S, RTH604773S, RTH605096S, RTH600867S, RTH600973S, RTH601360S, RTH602418S, RTH603175S, RTH604002S, RTH601174S, RTH603982S, RTH604229S, RTH600989S, RTH602370S, RTH604884S, RTH604269S, RTH604273S, RTH600950S, RTH601196S, RTH601138S, RTH602367S, RTH602478S, RTH603190S, RTH603941S, RTH604203S, RTH604769S, RTH600764S, RTH601105S, RTH602426S, RTH602432S, RTH603219S, RTH604749S, RTH600952S, RTH600963S, RTH601362S, RTH601659S, RTH601881S, RTH602273S, RTH602326S, RTH603185S, RTH603248S, RTH601358S, RTH602393S, RTH602469S, RTH604817S, RTH601067S, RTH601070S, RTH600721S, RTH600791S, RTH601141S, RTH605097S, RTH601925S, RTH600998S, RTH601098S, RTH601166S, RTH601315S, RTH602325S, RTH602388S, RTH604266S, RTH601100S, RTH602472S, RTH600720S, RTH601039S, RTH600526S, RTH603777S, RTH601101S, RTH601186S, RTH600903S, RTH601289S, RTH600078S, RTH601293S, RTH601605S, RTH601860S, RTH602410S, RTH602745S, RTH601855S, RTH602405S, RTH605093S, RTH603784S, RTH603789S, RTH603203S, RTH600993S, RTH601117S, RTH602484S, RTH603025S, RTH600079S, RTH601178S, RTH601187S, RTH602456S, RTH603548S, RTH603791S, RTH604691S, RTH604735S, RTH600780S, RTH600051S, RTH600052S, RTH600779S, RTH601063S, RTH601164S, RTH601683S, RTH602013S, RTH602308S, RTH602309S, RTH602386S, RTH602406S, RTH603415S, RTH603966S, RTH604238S, RTH602460S, RTH602340S, RTH601051S, RTH602683S, RTH603232S, RTH604223S, RTH605137S, RTH603189S, RTH604265S, RTH601123S, RTH602681S, RTH600772S, RTH600032S, RTH600033S, RTH600034S, RTH600090S, RTH600276S, RTH600919S, RTH600921S, RTH601118S, RTH601119S, RTH601120S, RTH602389S, RTH603162S, RTH603163S, RTH603819S, RTH603820S, RTH605260S, RTH605292S, RTH603127S, RTH600073S, RTH600075S, RTH600344S, RTH601108S, RTH601109S, RTH600761S, RTH601476S, RTH601492S, RTH601691S, RTH601720S, RTH602468S, RTH602718S, RTH602722S, RTH603206S, RTH601940S, RTH600048S, RTH600064S, RTH600049S, RTH600813S, RTH601198S, RTH602669S, RTH600054S, RTH600055S, RTH600056S, RTH600789S, RTH600895S, RTH601033S, RTH601052S, RTH601058S, RTH601127S, RTH601481S, RTH601681S, RTH601922S, RTH601997S, RTH602311S, RTH601990S, RTH603772S, RTH603813S, RTH603797S, RTH601490S, RTH601684S, RTH603173S, RTH601335S, RTH601287S, RTH603818S, RTH603787S, RTH602433S, RTH604233S, RTH600037S, RTH601004S, RTH604438S, RTH600782S, RTH600976S, RTH603962S, RTH601338S, RTH601469S, RTH602353S, RTH601497S, RTH602300S, RTH603792S, RTH603795S, RTH601180S, RTH601014S, RTH601181S, RTH601204S, RTH601914S, RTH601991S, RTH603124S, RTH603179S, RTH603182S, RTH603969S, RTH603246S, RTH603980S, RTH604236S, RTH601917S, RTH602341S, RTH600065S, RTH601473S, RTH601957S, RTH602342S, RTH600912S, RTH603762S, RTH600964S, RTH601060S, RTH601110S, RTH601114S, RTH602708S, RTH602863S, RTH602881S, RTH603807S, RTH603808S, RTH601085S, RTH600913S, RTH601034S, RTH601081S, RTH601606S, RTH602784S, RTH600858S, RTH601678S, RTH602685S, RTH600778S, RTH600955S, RTH600957S, RTH603268S, RTH603850S, RTH601044S, RTH601535S, RTH602938S, RTH601045S, RTH601674S, RTH600948S, RTH600980S, RTH601126S, RTH601524S, RTH601530S, RTH603251S, RTH600930S, RTH602904S, RTH603042S, RTH603074S, RTH601103S, RTH604944S, RTH603230S, RTH601089S, RTH603829S, RTH600927S, RTH601018S, RTH601023S, RTH601102S, RTH601532S, RTH601599S, RTH602350S, RTH602915S, RTH603830S, RTH604263S, RTH601202S, RTH603044S, RTH603028S, RTH603031S, RTH600890S, RTH603257S, RTH603269S, RTH601106S, RTH601531S, RTH601536S, RTH601537S, RTH602916S, RTH604949S, RTH600741S, RTH600795S, RTH600899S, RTH600925S, RTH600939S, RTH601092S, RTH601176S, RTH601539S, RTH604005S, RTH600938S, RTH601035S, RTH601918S, RTH602706S, RTH603228S, RTH600910S, RTH601099S, RTH601527S, RTH601529S, RTH604256S, RTH600951S, RTH602869S, RTH602953S, RTH600960S, RTH601020S, RTH601086S, RTH601597S, RTH601607S, RTH602344S, RTH602347S, RTH602349S, RTH602351S, RTH602911S, RTH602912S, RTH603008S, RTH603242S, RTH600931S, RTH600739S, RTH600773S, RTH602774S, RTH600977S, RTH602996S, RTH602899S, RTH601047S, RTH601080S, RTH601097S, RTH602783S, RTH602859S, RTH602984S, RTH603254S, RTH601505S, RTH600928S, RTH602345S, RTH603843S, RTH604010S, RTH604011S, RTH604260S, RTH602867S, RTH603250S, RTH601028S, RTH601173S, RTH601540S, RTH601595S, RTH603802S, RTH603812S, RTH603854S, RTH602853S, RTH600934S, RTH600936S, RTH600942S, RTH603266S, RTH601087S, RTH603267S, RTH603824S, RTH601507S, RTH603015S, RTH601689S, RTH602482S, RTH604071S, RTH604939S, RTH602868S, RTH602860S, RTH603855S, RTH601011S, RTH601022S, RTH601596S, RTH601608S, RTH600864S, RTH600909S, RTH600911S, RTH600988S, RTH600996S, RTH601002S, RTH601049S, RTH601050S, RTH601077S, RTH601506S, RTH601512S, RTH601515S, RTH601519S, RTH601601S, RTH601602S, RTH601612S, RTH601630S, RTH602343S, RTH602851S, RTH602852S, RTH602895S, RTH602898S, RTH603225S, RTH603258S, RTH603264S, RTH603849S, RTH603852S, RTH601082S, RTH601654S, RTH603826S, RTH602846S, RTH600924S, RTH601017S, RTH601031S, RTH602848S, RTH600920S, RTH601143S, RTH600740S, RTH601091S, RTH603844S, RTH602903S, RTH603845S, RTH600693S, RTH602914S, RTH603227S, RTH600959S, RTH601016S, RTH601019S, RTH601054S, RTH601084S, RTH601165S, RTH601628S, RTH602888S, RTH602891S, RTH603846S, RTH604072S, RTH604261S, RTH604937S, RTH604946S, RTH600958S, RTH600933S, RTH600900S, RTH602906S, RTH603259S, RTH600880S, RTH601032S, RTH601513S, RTH602892S, RTH604938S, RTH600848S, RTH600941S, RTH600962S, RTH604083S, RTH602707S, RTH603263S, RTH603811S, RTH600785S, RTH601603S, RTH601685S, RTH602346S, RTH603010S, RTH603014S, RTH603243S, RTH604050S, RTH604258S, RTH600904S, RTH601042S, RTH601523S, RTH603260S, RTH600896S, RTH602874S, RTH602759S, RTH603256S, RTH602457S, RTH604259S, RTH600765S, RTH602002S, RTH603061S, RTH601021S, RTH603814S, RTH600760S, RTH601989S, RTH603034S, RTH603045S, RTH603046S, RTH603092S, RTH603097S, RTH603113S, RTH603157S, RTH601276S, RTH603800S, RTH600283S, RTH600727S, RTH600981S, RTH601288S, RTH601465S, RTH601471S, RTH602411S, RTH603222S, RTH604224S, RTH601136S, RTH600902S, RTH600908S, RTH601090S, RTH601528S, RTH602894S, RTH600916S, RTH602943S, RTH603853S, RTH602304S, RTH602348S, RTH604227S, RTH604948S, RTH603271S, RTH600788S, RTH601088S, RTH602909S, RTH603249S, RTH603825S, RTH601000S, RTH602116S, RTH603953S, RTH600039S, RTH601113S, RTH604288S, RTH600796S, RTH600797S, RTH604291S, RTH604292S, RTH604293S, RTH604833S, RTH604858S, RTH604863S, RTH604881S, RTH605263S, RTH603233S, RTH600730S, RTH601215S, RTH603759S, RTH600691S, RTH602873S, RTH603016S, RTH603047S, RTH603048S, RTH603088S, RTH603091S, RTH603485S, RTH604008S, RTH604287S, RTH604786S, RTH604787S, RTH604788S, RTH604789S, RTH600874S, RTH600901S, RTH602855S, RTH602942S, RTH602951S, RTH602974S, RTH602975S, RTH603159S, RTH602675S, RTH600792S, RTH601639S, RTH601901S, RTH601987S, RTH602007S, RTH603208S, RTH603210S, RTH603952S, RTH601920S, RTH603117S, RTH602004S, RTH601131S, RTH601133S, RTH600021S, RTH600793S, RTH601083S, RTH601650S, RTH604671S, RTH602680S, RTH602840S, RTH602882S, RTH602893S, RTH603033S, RTH603043S, RTH603114S, RTH603132S, RTH603138S, RTH603184S, RTH603198S, RTH603207S, RTH603209S, RTH603218S, RTH603224S, RTH603236S, RTH603421S, RTH603458S, RTH603761S, RTH603764S, RTH603765S, RTH603768S, RTH603769S, RTH603781S, RTH603782S, RTH603786S, RTH603788S, RTH603816S, RTH603851S, RTH603858S, RTH603870S, RTH603879S, RTH603889S, RTH603907S, RTH603910S, RTH603920S, RTH603923S, RTH603931S, RTH603934S, RTH603935S, RTH603940S, RTH603975S, RTH603985S, RTH604073S, RTH604075S, RTH604076S, RTH604217S, RTH604228S, RTH604234S, RTH604251S, RTH604267S, RTH604282S, RTH604603S, RTH604674S, RTH604676S, RTH604685S, RTH604690S, RTH604719S, RTH604724S, RTH604725S, RTH604736S, RTH604751S, RTH604775S, RTH604777S, RTH604784S, RTH604821S, RTH604915S, RTH604932S, RTH605054S, RTH605057S, RTH605090S, RTH605098S, RTH605108S, RTH605115S, RTH605122S, RTH605123S, RTH605126S, RTH605130S, RTH605154S, RTH605156S, RTH605177S, RTH605233S, RTH605235S, RTH605291S, RTH605308S, RTH605310S, RTH605313S, RTH605333S, RTH605336S, RTH605337S, RTH605360S, RTH602163S, RTH602376S, RTH602145S, RTH602157S, RTH602166S, RTH602274S, RTH602291S, RTH602293S, RTH602294S, RTH602295S, RTH602297S, RTH602362S, RTH602364S, RTH602379S, RTH602969S, RTH603757S, RTH604248S, RTH605246S; GTIN 00763000402204, Lot Serial Numbers: RTH600799S, RTH601327S, RTH602128S, RTH602497S, RTH602883S, RTH603895S, RTH601236S, RTH603411S, RTH601982S, RTH601970S, RTH602009S, RTH600923S, RTH600937S, RTH603456S, RTH601245S, RTH601346S, RTH603896S, RTH603913S, RTH600816S, RTH600819S, RTH600897S, RTH600898S, RTH601247S, RTH601348S, RTH602507S, RTH603027S, RTH603866S, RTH601230S, RTH600743S, RTH600751S, RTH600798S, RTH600800S, RTH600804S, RTH600873S, RTH600877S, RTH600966S, RTH601142S, RTH601252S, RTH601308S, RTH602360S, RTH602506S, RTH602960S, RTH603448S, RTH603883S, RTH604185S, RTH604281S, RTH602763S, RTH603038S, RTH602391S, RTH600737S, RTH600794S, RTH602944S, RTH603454S, RTH603455S, RTH603899S, RTH604257S, RTH604270S, RTH604704S, RTH604714S, RTH602511S, RTH601974S, RTH602491S, RTH603936S, RTH604186S, RTH603867S, RTH601347S, RTH604715S, RTH601209S, RTH602503S, RTH604675S, RTH604712S, RTH602115S, RTH603401S, RTH602466S, RTH600731S, RTH600869S, RTH602742S, RTH604280S, RTH600843S, RTH601385S, RTH603989S, RTH603368S, RTH600855S, RTH600841S, RTH601976S, RTH602534S, RTH600893S, RTH603441S, RTH600828S, RTH602792S, RTH603487S, RTH600744S, RTH601225S, RTH600995S, RTH601221S, RTH603881S, RTH601212S, RTH603894S, RTH604695S, RTH600742S, RTH600769S, RTH601037S, RTH603915S, RTH604278S, RTH600829S, RTH601207S, RTH601329S, RTH601339S, RTH602728S, RTH600830S, RTH603037S, RTH600878S, RTH602369S, RTH603901S, RTH600889S, RTH603884S, RTH600875S, RTH602750S, RTH603476S, RTH604678S, RTH603451S, RTH603480S, RTH600822S, RTH602884S, RTH603413S, RTH603450S, RTH601250S, RTH602381S, RTH602857S, RTH603897S, RTH602352S, RTH602392S, RTH602887S, RTH600735S, RTH600817S, RTH600833S, RTH603444S, RTH603473S, RTH603887S, RTH602886S, RTH600777S, RTH600849S, RTH602365S, RTH602361S, RTH600818S, RTH602366S, RTH602720S, RTH602937S, RTH600824S, RTH602724S, RTH603438S, RTH600757S, RTH602498S, RTH602483S, RTH603437S, RTH603479S, RTH604682S, RTH601218S, RTH603902S, RTH604001S, RTH604090S, RTH600767S, RTH600881S, RTH600953S, RTH600985S, RTH603011S, RTH604285S, RTH601139S, RTH603869S, RTH603873S, RTH600803S, RTH600807S, RTH602415S, RTH603919S, RTH600729S, RTH600859S, RTH600926S, RTH601027S, RTH601239S, RTH601332S, RTH602408S, RTH602737S, RTH603410S, RTH602117S, RTH602967S, RTH600918S, RTH602731S, RTH600956S, RTH600802S, RTH600876S, RTH601224S, RTH600825S, RTH600990S, RTH601029S, RTH601053S, RTH601232S, RTH602910S, RTH603459S, RTH600842S, RTH600944S, RTH600945S, RTH601243S, RTH602363S, RTH602377S, RTH602504S, RTH602514S, RTH602744S, RTH603348S, RTH600872S, RTH601352S, RTH600827S, RTH602754S, RTH603472S, RTH603416S, RTH603872S, RTH600884S, RTH600801S, RTH600922S, RTH603457S, RTH600863S, RTH602395S, RTH600879S, RTH602494S, RTH602849S, RTH603921S, RTH602521S, RTH602766S, RTH602131S, RTH602502S, RTH602740S, RTH604284S, RTH600821S, RTH602512S, RTH603436S, RTH600969S, RTH601326S, RTH603354S, RTH601235S, RTH602127S, RTH602399S, RTH601132S, RTH601129S, RTH601240S, RTH602119S, RTH602751S, RTH603547S, RTH600732S, RTH602612S, RTH600857S, RTH602500S, RTH600786S, RTH603352S, RTH600888S, RTH600961S, RTH600809S, RTH600965S, RTH602394S, RTH603477S, RTH601246S, RTH602780S, RTH600810S, RTH602499S, RTH601233S, RTH601234S, RTH603876S, RTH602761S, RTH602771S, RTH603024S, RTH601226S, RTH601238S, RTH600766S, RTH601214S, RTH601219S, RTH601958S, RTH602501S, RTH602711S, RTH602755S, RTH602368S, RTH601241S, RTH604721S, RTH602770S, RTH600724S, RTH600725S, RTH602801S, RTH603423S, RTH604728S, RTH600826S, RTH600838S, RTH600854S, RTH602965S, RTH601095S, RTH603350S, RTH600832S, RTH603888S, RTH600940S, RTH602375S, RTH602516S, RTH602931S, RTH600736S, RTH600749S, RTH600840S, RTH602934S, RTH604563S, RTH600947S, RTH602930S, RTH603914S, RTH603916S, RTH603918S, RTH600837S, RTH600752S, RTH600860S, RTH602380S, RTH603917S, RTH602717S, RTH602721S, RTH602493S, RTH601206S, RTH602870S, RTH601248S, RTH602496S, RTH603418S, RTH600812S, RTH601211S, RTH601244S, RTH601964S, RTH602467S, RTH602768S, RTH600839S, RTH602927S, RTH603358S, RTH600846S, RTH602419S, RTH602977S, RTH600783S, RTH600868S, RTH602121S, RTH603357S, RTH602786S, RTH602402S, RTH603440S, RTH601223S, RTH601220S, RTH602519S, RTH600814S, RTH600887S, RTH602845S, RTH603439S, RTH600992S, RTH601205S, RTH602289S, RTH602488S, RTH603909S, RTH600844S, RTH601312S, RTH600726S, RTH601986S, RTH600746S, RTH600905S, RTH600906S, RTH603483S, RTH601251S, RTH603484S, RTH601381S, RTH602752S, RTH602130S, RTH602487S, RTH603865S, RTH603891S, RTH600850S, RTH601355S, RTH603029S, RTH602114S, RTH602713S, RTH600851S, RTH601389S, RTH602865S, RTH600755S, RTH600758S, RTH601979S, RTH601001S, RTH604720S, RTH602544S, RTH603447S, RTH603971S, RTH602973S, RTH603882S, RTH601237S, RTH602492S, RTH603912S, RTH600747S, RTH601231S, RTH602710S, RTH602999S, RTH600853S, RTH603026S, RTH601305S, RTH602372S, RTH602374S, RTH602954S, RTH601208S, RTH603908S, RTH600738S, RTH600820S, RTH601222S, RTH601227S, RTH601242S, RTH603443S, RTH603890S, RTH604717S, RTH602508S, RTH603906S, RTH602417S, RTH602509S, RTH603905S, RTH603435S, RTH602008S, RTH602773S, RTH602387S, RTH602616S, RTH603445S, RTH603040S, RTH600815S, RTH600831S, RTH601228S, RTH602866S, RTH600870S, RTH601996S, RTH601993S, RTH602739S, RTH603446S, RTH602123S, RTH602733S, RTH602968S, RTH600756S, RTH602518S, RTH601311S, RTH604667S, RTH600862S, RTH600805S, RTH601217S, RTH600834S, RTH601061S, RTH601249S, RTH601962S, RTH602126S, RTH603880S, RTH601229S, RTH602505S, RTH602831S, RTH603475S, RTH603922S, RTH604661S, RTH602924S, RTH601030S, RTH602775S, RTH603898S, RTH600808S, RTH602513S, RTH602961S, RTH602966S, RTH603893S, RTH601210S, RTH601216S, RTH602719S, RTH604665S
US Nationwide - Worldwide Distribution
4151 units