FDA Enforcement Class II Terminated

PASCAL and PASCAL Streamline (532 nm.577 nm); Treats the patient's eye.

Recall: Z-1774-2013 · Reported July 31, 2013

Enforcement

Recall Number
Z-1774-2013
Event ID
65688
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Topcon Medical Laser Systems, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 31, 2013
Initiation Date
March 27, 2012
Classification Date
July 19, 2013
Termination Date
July 19, 2013
Address
3130 Coronado Dr, N/A, Santa Clara, CA, 95054-3205, United States

Description

PASCAL and PASCAL Streamline (532 nm.577 nm); Treats the patient's eye.

Reason

When using the "Favorites" function for saved settings, there is potential for a mismatch between displayed power and the actual power output. This may result in over or under treatment.

Code Info

PASCAL Standard 2R/G System v2.x.x and PASCAL Streamline 2R/G Software v2.x.x Serial numbers: United States 33057910 33060410 33063610 66078611 66084911 66085111 66085511 Out of United States OMC-179 11138509 11141209 11147209 11147809 11148910 11149010 11149210 11149710 . 11149810 11150110 11150210 11150410 11150610 11150810 11150910 11151010 1 151110 11151211 33054310 33059310 33059510 33062810" 66075211' 66076711 66077111 66078011 66079911 6608391'1 66084611 66084811 66085011 66085211

Distribution

Worldwide Distribution-USA and the countries of China, Hong Kong, Brazil, France, Japan, Netherlands, Pakistan, Malaysia, United Kingdom, South Korea, Thailand, Dubai, and South Africa.

Quantity

7 US, 33 foreign