FDA Enforcement Class II Terminated

Delta XL, Catalog Number: MS18596 in combination with Scio, Scio Four, Scio Four Oxi plus, Scio Four Oxi, Scio Four plus. Product Usage: The devices are intended to be used in the environment where patient care is provided by Healthcare Professionals, i.e. Physicians, Nurses, and Technicians, who will determine when use of the device is indicated, based upon their professional assessment of the patients medical condition. The Infinity Delta Series (Delta/Delta XL/Kappa) monitors are intended to be used on adult, pediatric, and neonatal populations, with the exception of the parameter Cardiac Output, ST Segment Analysis, and arrhythmia which are intended for use in the adult and pediatric populations only; and tcpO2, which for the neonatal population, is to only be used when the patient is not under gas anesthesia. For combination with Scio gas module: Scio gas module samples breathing gases from adults and pediatrics. The gas module continuously measure the content of CO2, N2O, O2 and one of the anesthetic agents, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane in any mixture and communicates real time and derived gas information to the In-finity monitors.

Recall: Z-1773-2017 · Reported April 19, 2017

Enforcement

Recall Number
Z-1773-2017
Event ID
76862
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Draegar Medical Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 19, 2017
Initiation Date
March 28, 2017
Classification Date
April 10, 2017
Termination Date
September 17, 2018
Address
6 Tech Dr, N/A, Andover, MA, 01810-2434, United States

Description

Delta XL, Catalog Number: MS18596 in combination with Scio, Scio Four, Scio Four Oxi plus, Scio Four Oxi, Scio Four plus. Product Usage: The devices are intended to be used in the environment where patient care is provided by Healthcare Professionals, i.e. Physicians, Nurses, and Technicians, who will determine when use of the device is indicated, based upon their professional assessment of the patients medical condition. The Infinity Delta Series (Delta/Delta XL/Kappa) monitors are intended to be used on adult, pediatric, and neonatal populations, with the exception of the parameter Cardiac Output, ST Segment Analysis, and arrhythmia which are intended for use in the adult and pediatric populations only; and tcpO2, which for the neonatal population, is to only be used when the patient is not under gas anesthesia. For combination with Scio gas module: Scio gas module samples breathing gases from adults and pediatrics. The gas module continuously measure the content of CO2, N2O, O2 and one of the anesthetic agents, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane in any mixture and communicates real time and derived gas information to the In-finity monitors.

Reason

It was reported that a set low O2 alarm does not go off although the measured O2 level is below the alarm limit.

Code Info

VF10.0 software

Distribution

Worldwide - US Nationwide distribution in the states of: AL, CA, GA, IL, LA, MA NY, OH, PA, TN, TX and VA and the countries of: AR AT BE BG BN BO BY CA CH CL CO CZ DE EG ES FI FR GB GR HR HU ID IE IL IN IT LT MX MY NL NO NP PA PE PK PL PT RO RS RU SA SE SK TH TR TW UG UZ VN ZA

Quantity

2,156 (total for all devices)