FDA Enforcement Class I Ongoing

Boston Scientific Cardiac Resynchronization Therapy (CRT)labeled as: 1. VALITUDE CRT-P EL, Model Number U125; 2. VALITUDE CRT-P EL MRI, Model Number U128; 3. VISIONIST CRT-P EL, Model Number U225; 4. VISIONIST CRT-P EL, Model Number U226; 5. VISIONIST CRT-P EL MRI, Model Number U228.

Recall: Z-1771-2026 · Reported May 6, 2026

Enforcement

Recall Number
Z-1771-2026
Event ID
98615
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Boston Scientific Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 6, 2026
Initiation Date
March 19, 2026
Classification Date
April 24, 2026
Address
4100 Hamline Ave N, Saint Paul, MN, 55112-5700, United States

Description

Boston Scientific Cardiac Resynchronization Therapy (CRT)labeled as: 1. VALITUDE CRT-P EL, Model Number U125; 2. VALITUDE CRT-P EL MRI, Model Number U128; 3. VISIONIST CRT-P EL, Model Number U225; 4. VISIONIST CRT-P EL, Model Number U226; 5. VISIONIST CRT-P EL MRI, Model Number U228.

Reason

Updated software, Brady software maintenance release 6 (SMR6), is now available for the ACCOLADE" family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps)1 and the advisory population is expanding to include all CRT-P and dual-chamber extended life (DR-EL) devices. Dual chamber modes are specifically indicated for treatment of the following: " Conduction disorders that require restoration of AV synchrony, including varying degrees of AV block " VVI intolerance (i.e., pacemaker syndrome) in the presence of persistent sinus rhythm " Low cardiac output or congestive heart failure secondary to bradycardia

Code Info

The advisory population includes all models listed; however, the bounding differs by battery type: " All serialized DR-EL pacemakers and CRT-Ps from the ACCOLADE family are included in the advisory population. " ACCOLADE DR-SL and SR-SLs with a use-by-date (UBD) on or before 30 June 2025 are included in the advisory population. Model number alone will not precisely identify individual DR-SL or SR-SL devices in the advisory population. To determine if a device is affected, enter a model/serial into the device lookup tool at www.BostonScientific.com/lookup.

Distribution

worldwide

Quantity

685,776 units