FDA Enforcement Class II Ongoing

SIGNA Premier systems

Recall: Z-1767-2026 · Reported April 8, 2026

Enforcement

Recall Number
Z-1767-2026
Event ID
98579
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
GE Healthcare LLC
Voluntary / Mandated
Voluntary: Firm initiated
Report Date
April 8, 2026
Initiation Date
February 27, 2026
Classification Date
April 2, 2026
Address
3001 W Radio Dr, Florence, SC, 29501-7806, United States

Description

SIGNA Premier systems

Reason

Whole-Body MR scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with ferrous (steel) fittings instead of non-ferrous (brass) fittings on the magnet rear and if system is accessed to perform service while the magnet is ramped, the magnetic field could attract the ferrous fittings potentially resulting in personnel injury.

Code Info

UDI-DI/GTIN: 00840682135269, 00195278010797. System ID: 415353UMR2, 415723SHMR19, 413253MR3T, 908725MR, 609652MR3T, M48823, 7168453T, 513584MR5, 513636EASTMR, 513636T1MR, 605333MRPRM, 832522CRMR1, 703858PRE2, 804675VA3T, 082427110162, 082427220143, 082427020147, 082427200242, 082427160461, 082427250139, 082427140382, M4216469, 353745MR01, GON5655515, KRMR56474971, KRMR56203791, KRMR55433661, KRMR56344161, KRMR55774051, KRMR55914661, NOMRI26833321, 15974045, GON1882380, ESMRI27056983, ESMRI26960129, A203MR06, 690120MR01, 11524261, VNMRI26792525, VNMRI26592248, VNMRI26809677

Distribution

Worldwide - US Nationwide distribution in the states of CA, MA, NJ, NY, OH, SD, TX, VA, PR and the countries of Sweden, Switzerland, United Kingdom, Vietnam, China, Norway, Portugal, Spain, France, Germany, Greece, Korea, Republic of

Quantity

41