FDA Enforcement Class II Terminated

VITROS Immunodiagnostic Products NT-pro BNP Reagent Pack, REF/Catalog 680 2156, IVD --- Ortho-Clinical Diagnostics Pencoed, Bridgend CF35 5PZ, UK

Recall: Z-1765-2017 · Reported April 19, 2017

Enforcement

Recall Number
Z-1765-2017
Event ID
76665
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
ORTHO-CLINICAL DIAGNOSTICS
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
April 19, 2017
Initiation Date
March 7, 2017
Classification Date
April 8, 2017
Termination Date
June 14, 2019
Address
FELINDRE MEADOWS, N/A, Bridgend, N/A, N/A, United Kingdom

Description

VITROS Immunodiagnostic Products NT-pro BNP Reagent Pack, REF/Catalog 680 2156, IVD --- Ortho-Clinical Diagnostics Pencoed, Bridgend CF35 5PZ, UK

Reason

Increased frequency of calibration failures for VITROS Immunodiagnostic Products NT-proBNP Reagent Lots 1568, 1570, 1580 and 1590, due to background signals for these affected NT-proBNP lots that have been increasing with time since release testing, resulting in higher than expected VITROS Immunodiagnostic Products NT-proBNP level 1 calibrator signals produced, leading to calibration failures.

Code Info

Lot 1570 (expiry 22 March 2017), Lot 1580 (expiry 13 April 2017), Lot 1590 (expiry 27 April 2017)

Distribution

Worldwide: United States, Puerto Rico, Bermuda, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Poland, Singapore, Spain, and Venezuela.

Quantity

US: 6324; Foreign: 1868