FDA Enforcement Class II Terminated

Brilliance CT 16 Slice Air (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.

Recall: Z-1765-2015 · Reported June 17, 2015

Enforcement

Recall Number
Z-1765-2015
Event ID
71134
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips Medical Systems (Cleveland) Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 17, 2015
Initiation Date
March 6, 2015
Classification Date
June 11, 2015
Termination Date
March 13, 2017
Address
595 Miner Rd, N/A, Cleveland, OH, 44143-2131, United States

Description

Brilliance CT 16 Slice Air (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.

Reason

The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.

Code Info

Model #: 728246; Serial #s: 3007, 3011, 3014, 3047, 3050, 3095, 3117, 3123, 3162, 3164, 3170, 3197, 3247, 3331, 3407, 3433, 3458, 3497, 3578, 3721, 3723, 4015, 4019, 4042, 5013, 5019, 5038, 5040, 5050, 5052, 5090, 5091, 5094, 5097, 5101, 5116, 5133, 5143, 5150, 5155, 5172, 5179, 5180, 5192, 5202, 5252, 5335, 5336, 5342, 5343, 5399, 5416, 5455, 5467, 5601, 5703, 5811, 5869, 5929, 6006, 6012, 6021, 6030, 6037, 6038, 6039, 6043, 6045, 6048, 6051, 6055, 6056, 6064, 6071, 6072, 6074, 6076, 6078, 6079, 6081, 6082, 6083, 6084, 6085, 6087, 6088, 6089, 6090, 6092, 6093, 6094, 6095, 6097, 6100, 6104, 6105, 6107, 6109, 6110, 6111, 6112, 6114, 6116, 6118, 6119, 6120, 6122, 6123, 6124, 6125, 6126, 6127, 6128, 6129, 6130, 6131, 6132, 6133, 6135, 6136, 6137, 6139, 6140, 6141, 6143, 6144, 6145, 6146, 6147, 6148, 6149, 6151, 6153, 6155, 6156, 6157, 6159, 6161, 6162, 6163, 6164, 6165, 6166, 6167, 6168, 6169, 6170, 6171, 6172, 6173, 6174, 6175, 6176, 6177, 6178, 6179, 6180, 6181, 6182, 6183, 6184, 6185, 6186, 6187, 6188, 6189, 6190, 6191, 6192, 6195, 6197, 6198, 6199, 6200, 6201, 6202, 6203, 6204, 6205, 6206, 6207, 6208, 6209, 6210, 6211, 6212, 6213, 6215, 6216, 6217, 6219, 6220, 6221, 6222, 6223, 6224, 6225, 6226, 6227, 6228, 6229, 6230, 6231, 6232, 6233, 6234, 6235, 6237, 6238, 6239, 6240, 6241, 6242, 6243, 6245, 6246, 6247, 6248, 6249, 6250, 6251, 6252, 6253, 6254, 6255, 6256, 6257, 6259, 6260, 6261, 6262, 6264, 6265, 6266, 6267, 6268, 6269, 6270, 6271, 6272, 6273, 6275, 6276, 6279, 6280, 6281, 6282, 6283, 6285, 6287, 6288, 6289, 6290, 6291, 6292, 6293, 6294, 6295, 6296, 6297, 6298, 6299, 6300, 6301, 6302, 6303, 6304, 6306, 6307, 6308, 6309, 6310, 6311, 6312, 6313, 6314, 6315, 6316, 6317, 6319, 6320, 6321, 6322, 6323, 6324, 6325, 6326, 6327, 6328, 6329, 6330, 6331, 6332, 6334, 6335, 6336, 6337, 6338, 6339, 6340, 6341, 6342, 6343, 6344, 6345, 6346, 6347, 6348, 6349, 6350, 6351, 6352, 6353, 6354, 6355, 6356, 6357, 6358, 6359, 6360, 6361, 6362, 6363, 6364, 6366, 6367, 6368, 6369, 6370, 6371, 6373, 6374, 6375, 6376, 6377, 6378, 6379, 6380, 6382, 6383, 6384, 6385, 6386, 6387, 6389, 6392, 6393, 6394, 6395, 6397, 6398, 6399, 6400, 6401, 6402, 6403, 6404, 6405, 6406, 6407, 6408, 6409, 6410, 6411, 6412, 6413, 6414, 6415, 6416, 6418, 6419, 6420, 6421, 6422, 6423, 6424, 6425, 6426, 6427, 6428, 6429, 6430, 6431, 6432, 6433, 6434, 6435, 6436, 6437, 6438, 6439, 6440, 6441, 6442, 6443, 6447, 6448, 6449, 6450, 6451, 6452, 6453, 6454, 6455, 6456, 6457, 6458, 6460, 6461, 6462, 6463, 6464, 6465, 6466, 6467, 6468, 6469, 6470, 6471, 6472, 6473, 6474, 6475, 6476, 6477, 6478, 6479, 6480, 6481, 6482, 6483, 6484, 6485, 6486, 6487, 6488, 6489, 6490, 6491, 6492, 6493, 6497, 6498, 6499, 6500, 6502, 6503, 6504, 6505, 6506, 6507, 6508, 6509, 6510, 6511, 6512, 6513, 6514, 6515, 6516, 6517, 6518, 6519, 6521, 6522, 6523, 6524, 6525, 6526, 6527, 6528, 6529, 6530, 6531, 6532, 6534, 6535, 6536, 6537, 6538, 6539, 6540, 6541, 6542, 6543, 6544, 6545, 6546, 6547, 6548, 6549, 6550, 6551, 6552, 6553, 6554, 6555, 6556, 6557, 6558, 6559, 6560, 6561, 6562, 6563, 6564, 6565, 6566, 6567, 6568, 6569, 6570, 6571, 6572, 6573, 6574, 6575, 6576, 6577, 6580, 6581, 6582, 6583, 6584, 6585, 6586, 6587, 6588, 6589, 6590, 6592, 6593, 6594, 6595, 6596, 6597, 6598, 6599, 6600, 6601, 6602, 6603, 6604, 6605, 6606, 6607, 6608, 6609, 6610, 6611, 6612, 6613, 6615, 6616, 6618, 6619, 6620, 6621, 6622, 6623, 6624, 6625, 6626, 6627, 6628, 6629, 6630, 6631, 6632, 6633, 6634, 6635, 6636, 6638, 6639, 6640, 6641, 6643, 6644, 6645, 6646, 6647, 6648, 6649, 6650, 6651, 6652, 6653, 6654, 6655, 6656, 6657, 6658, 6660, 6661, 6662, 6663, 6664, 6665, 6666, 6667, 6668, 6669, 6670, 6671, 6672, 6673, 6674, 6675, 6676, 6677, 6678, 6679, 6680, 6681, 6682, 6683, 6684, 6685, 6686, 6687, 6688, 6689, 6690, 6691, 6692, 6693, 6694, 6695, 6696, 6697, 6698, 6699, 6700, 6701, 6702, 6703, 6704, 6705, 6706, 6707, 6708, 6709, 6710, 6711, 6712, 6713, 6714, 6715, 6716, 6717, 6718, 6719, 6720, 6721, 6722, 6723, 6724, 6725, 6726, 6727, 6728, 6729, 6730, 6731, 6732, 6733, 6734, 6735, 6736, 6737, 6738, 6740, 6741, 6742, 6743, 6744, 6745, 6746, 6747, 6748, 6749, 6750, 6751, 6752, 6753, 6754, 6755, 6756, 6757, 6758, 6759, 6760, 6761, 6762, 6763, 6764, 6765, 6766, 6767, 6768, 6769, 6770, 6771, 6772, 6775, 6777, 6778, 6779, 6782, 6783, 6784, 9744, 30028, 30049, 30051, 30063, 30076, 30100, 30103, 30109, 30110, 30115, 30122, 40018, 40020, 40025, 40032, 50011, 50023, 50028, 50031, 50041, 50049, 50051, 50053, 50143, 50144, 50145, 50146, 50147, 50148, 50149, 50150, 50151, 50152, 50153, 50154, 50156, 50157, 50158, 50159, 50160, 50161, 50162, 50163, 50164, 50165, 50166, 50168, 50170, 50171, 50172, 50175, 50176, 50177, 50178, 50179, 50180, 50181, 50182, 50184, 50185, 50186, 50187, 50188, 50189, 50190, 50191, 50192, 50193, 50194, 50195, 50197, 50198, 50199, 50200, 50201, 50202, 50204, 50223, 50224, 50225, 60007, 60013, 60018, 60020, 260005, 260006, 260007, 260009, 260010, 260011, 260013, 260015, 260016, 260017, 260018, 260019, 260020, 260021, 260022, 260023, 260024, 260025, 260027, 260028, 260029, 260030, 260032, 260035, 260036, 260038, 260039, 260040, 260041, 260042, 260043, 260044, 260045, 260046, 260047, 260048, 260049, 260050, 260051, 260052, 260053, 260056, 260057, 260063, 260064, 260065, 260066, 260067, 260068, 260069, 260070, 260071, 260073, 260074, 260075, 260076, 260077, 260078, 260084 & 260086.

Distribution

Worldwide Distribution-US (nationwide) including DC & Puerto Rico and the following countries: Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Cambodia, Cameroon, Cayman Islands, China, Colombia, Democratic Congo, Croatia, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Ethiopia, Finland, France, Gabon, Georgia, Germany, Ghana, Greece, Guatemala, Hong Kong, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Kuwait, Kyrgyzstan, LAO People's Dem. Rep., Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malawi, Malaysia, Malta, Mauritius, Monaco, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Philippines, Poland, Portugal, R¿union, Romania, Russia, Rwanda, Saudi Arabia, Senegal, Serbia & Montenegro, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam, Yemen, and Zambia.

Quantity

847 units