FDA Enforcement Class II Ongoing

Clinical Chemistry Calibration Serum (Catalog number Cal2351) that is 20x5ml lyophilized human based serum covering 38 commonly used clinical chemistry tests.

Recall: Z-1759-2024 · Reported May 15, 2024

Enforcement

Recall Number
Z-1759-2024
Event ID
94358
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Randox Laboratories Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
May 15, 2024
Initiation Date
March 5, 2024
Classification Date
May 3, 2024
Address
55 Diamond Road, N/A, Crumlin (North), N/A, N/A, Ireland

Description

Clinical Chemistry Calibration Serum (Catalog number Cal2351) that is 20x5ml lyophilized human based serum covering 38 commonly used clinical chemistry tests.

Reason

Randox Laboratories has identified that Inorganic Phosphate in Calibration Serum Level 3, CAL2351 & CAL10388, is running with a negative bias on RX Series instruments compared to other methods.

Code Info

Catalogue No. CAL2351 & CAL10388; GTIN: 05055273200966; Lot No. 1260UE, 1262UE, 1295UE, 1297UE, 1298UE, 1315UE, 1325UE, 1326UE, 1295UE, & 1325UE.

Distribution

Domestic: Nationwide Distribution.

Quantity

80,366 units (US: 242 units; OUS: 80,094 units)