FDA Enforcement Class I Terminated

Advance Enforcer 35 Focal Force PTA Balloon Catheter 6mm x 4cm, 5FR/50cm, Catalog Number ASB5-35-50-6-4, REF Number G35248 Product Usage: The Advance Enforcer 35 Focal-Force PTA Balloon Catheter is intended for percutaneous transluminal angioplasty (PTA) of lesions in peripheral arteries, including iliac, renal, popliteal, infrapopliteal, femoral and iliofemoral, as well as obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the cerebral or coronary vasculature.

Recall: Z-1756-2019 · Reported June 26, 2019

Enforcement

Recall Number
Z-1756-2019
Event ID
82888
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
Cook Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 26, 2019
Initiation Date
May 24, 2019
Classification Date
June 17, 2019
Termination Date
October 31, 2019
Address
750 N Daniels Way, N/A, Bloomington, IN, 47404-9120, United States

Description

Advance Enforcer 35 Focal Force PTA Balloon Catheter 6mm x 4cm, 5FR/50cm, Catalog Number ASB5-35-50-6-4, REF Number G35248 Product Usage: The Advance Enforcer 35 Focal-Force PTA Balloon Catheter is intended for percutaneous transluminal angioplasty (PTA) of lesions in peripheral arteries, including iliac, renal, popliteal, infrapopliteal, femoral and iliofemoral, as well as obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the cerebral or coronary vasculature.

Reason

Multiple complaints were received for balloons bursting below the rated burst pressure. Potential adverse events that may occur if an affected product is used include a delay in the procedure, additional intervention, vessel injury, and balloon fragmentation in the patient.

Code Info

Lot Numbers: 9234424, 9331618

Distribution

Worldwide Distribution - US Nationwide The affected product was distributed to the following states: FL, OH, WA. The products were distributed to the following foreign countries: Belgium, Denmark, Finland, Germany, Italy, Netherlands, Norway, Poland, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.

Quantity

34