FDA Enforcement Class II Ongoing

Johnson & Johnson VISION Reusable LAMINAR High Flow Irrigation Sleeve and Test Chamber 21 Gauge REF OPOHF21L The irrigation sleeve is a device intended to direct irrigation solution across the shaft of a phacoemulsification tip, allowing the solution to enter the eye during ocular surgery.

Recall: Z-1755-2024 · Reported May 8, 2024

Enforcement

Recall Number
Z-1755-2024
Event ID
94231
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Johnson & Johnson Surgical Vision, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 8, 2024
Initiation Date
March 14, 2024
Classification Date
May 2, 2024
Address
31 Technology Dr Ste 200, Irvine, CA, 92618-2302, United States

Description

Johnson & Johnson VISION Reusable LAMINAR High Flow Irrigation Sleeve and Test Chamber 21 Gauge REF OPOHF21L The irrigation sleeve is a device intended to direct irrigation solution across the shaft of a phacoemulsification tip, allowing the solution to enter the eye during ocular surgery.

Reason

Due to a manufacturing variation of the irrigation sleeves which could result in a missing port hole, this non-conformity could lead to insufficient flow to cool the ultrasonic phaco tip, potentially leading to an unstable anterior chamber which could cause possible harms.

Code Info

Part Number: OPOHF21L UDI-DI/GTIN code: 05050474573376 Lot Numbers: 60477723 60479361 60505085 60508179 60508923

Distribution

U.S. Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, LA, MA, MD, ME, MN, MO, NC, NJ, NY, OR, PA, SC, TX, VA, WA, and WI.

Quantity

3,428 units