FDA Enforcement Class II Ongoing

INOmax DSIR -(delivery system) delivers INOMAX (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, Model Number: 10007

Recall: Z-1745-2022 · Reported September 21, 2022

Enforcement

Recall Number
Z-1745-2022
Event ID
90723
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Mallinckrodt, LLC.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
September 21, 2022
Initiation Date
June 23, 2022
Classification Date
September 15, 2022
Address
1425 Us Highway 206, Bedminster, NJ, 07921-2653, United States

Description

INOmax DSIR -(delivery system) delivers INOMAX (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, Model Number: 10007

Reason

The INOmax DSIR Delivery System, manufactured by Mallinckrodt Pharmaceuticals, is not compatible with the upgraded version of the Leoni Plus ventilator, manufactured by Lowenstein Medical. The Leoni Plus ventilator was recently upgraded in 2021 by Lowenstein Medical, with new hardware and software, and the upgraded version cannot be appropriately connected with the INOmax DSIR as instructed in the INOmax DSIR Operators Manual (OM). The INOmax DSIR OM requires the use of a one-way valve with the Leoni Plus ventilator during High Frequency Oscillatory Ventilation mode and warns that use of the device without a one-way valve may result in high Nitric Oxide (NO) delivery.

Code Info

All serial numbers

Distribution

Foreign Only: Argentina Australia Austria Beligum Canada Chile Columbia Czech Republic Denmark Ecuador Estonia Finland France Germany Great Britain Hungary Iceland Ireland Italy Latvia Mexico Netherlands Norway Poland Portugal Romania Slovakia South Africa Spain Sweden Switzerland UAE Uruguay

Quantity

4,225 units OUS