ZAP-X Radiosurgery System
Recall: Z-1742-2022
·
Reported September 21, 2022
Enforcement
- Recall Number
- Z-1742-2022
- Event ID
- 90716
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Zap Surgical Systems
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 21, 2022
- Initiation Date
- July 5, 2022
- Classification Date
- September 14, 2022
- Address
- 590 Taylor Way, San Carlos, CA, 94070-6307, United States
Description
ZAP-X Radiosurgery System
Reason
Software issue identified in cases of initial patient setup with large (greater-than-or-equal-to 1.5 degrees) rotational deviations, between digitally reconstructed radiographs, and pre-delivery X-ray images. In these cases the transitional alignment algorithm may incorrectly calculate the new treatment table offset values, leading to an incorrect position for the subsequent treatment isocenter.
Code Info
UDI-DI: 00860183001504, Serial Numbers, ZUC17152, ZUC17153, ZUC18155, ZUC20157, ZUC20158, ZUC20156, ZUC21161, ZUC21162, using Treatment Delivery Software Version: 1.8.53
Distribution
Worldwide distribution - US Nationwide distribution in the states of AZ, MD, FL, CO and the countries of DEU, CHE, JPN.
Quantity
8