Stryker TPS Universal Driver REF 5100- 99 Rx , The Stryker Total Performance System (TPS) is intended for use in the cutting, drilling, reaming, decorticating and smoothing of bone and other bone related tissues for a variety of applications such as ENT, dental orthopedic, maxillofacial, spinal, and plastic surgery. The instruments are also used in the placement of screws, wires, pins, and other fixation devices.
Enforcement
- Recall Number
- Z-1740-2013
- Event ID
- 65456
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Stryker Instruments Div. of Stryker Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 24, 2013
- Initiation Date
- July 1, 2013
- Classification Date
- July 16, 2013
- Termination Date
- September 16, 2013
- Address
- 4100 E Milham Ave, N/A, Portage, MI, 49002-9704, United States
Description
Stryker TPS Universal Driver REF 5100- 99 Rx , The Stryker Total Performance System (TPS) is intended for use in the cutting, drilling, reaming, decorticating and smoothing of bone and other bone related tissues for a variety of applications such as ENT, dental orthopedic, maxillofacial, spinal, and plastic surgery. The instruments are also used in the placement of screws, wires, pins, and other fixation devices.
The safety margin values detailed in the Engineering Design were entered incorrectly into the programming software. Handpieces that have been programmed with the incorrect parameters could result in unintended activation. There is a potential for injury to the patient or user as a result of unintended forward or reverse activation. There is also the potential that the handpiece may not reach 100
Serial Number 0214200753
Worldwide distribution: US (nationwide) including states of: AZ, FL, GA, NC, NY, SC, TX and VA; and countries of: India, Netherlands and Japan.