FDA Enforcement Class II Ongoing

MiniMed 670G Insulin Pump (MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782)

Recall: Z-1739-2026 · Reported April 8, 2026

Enforcement

Recall Number
Z-1739-2026
Event ID
98186
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Medtronic MiniMed, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 8, 2026
Initiation Date
February 13, 2026
Classification Date
April 2, 2026
Address
18000 Devonshire St, Northridge, CA, 91325-1219, United States

Description

MiniMed 670G Insulin Pump (MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782)

Reason

All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.

Code Info

Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 670G Insulin Pump: 643169946057, 763000096984, 763000090203, 763000165468, 763000190460, 763000190453, 763000216481, 763000187422, 763000187415, 763000165413, 763000165475, 763000367657, 763000289065, 763000264710, 763000283476, 763000283544, 763000283520, 763000283445, 763000393786, 763000289058, 763000374532, 763000283537, 763000365899, 763000371937, 763000283513, 763000165451, 763000165420, 763000165444, 763000165437, 763000165499, 763000165482, 763000165536, 763000165512, 763000165550, 763000165529, 763000235772, 763000165505, 763000235765, 763000165543, 763000257309, 763000283568, 763000283599, 763000289041, 763000289089, 763000289195, 763000289072, 763000289119, 763000289188, 763000289133, 763000289126, 763000289171, 763000289102, 763000404543, 763000289096, 763000289164, 763000404536, 763000324797, 763000264727, 763000324780, 763000283551, 763000283582, 763000374518, 763000374525, 763000374556, 763000374624, 763000374594, 763000374600, 763000399849, 763000399856, 763000408916, 763000408947, 763000371944, 763000365905, 763000374587, 763000408930, 763000283575, 763000157463, 763000203313, 763000252076, 763000283506, 763000411787, 643169946057, 763000096984, 763000190453, 763000283445, 763000395858, 763000278427, 763000090203, 763000283513, 763000365882, 763000283506, 76300028351303, 763000393786, 76300028351301, 763000166526, 643169939202, 763000190460, 763000283520, 763000072537, 763000256821, 643169939219, 763000187415, 763000283568: MiniMed670G System User Guide(M979513A011 1); MINIMED 630G and MINIMED 670G INSULIN PUMP SYSTEM(M042888C002 1); MINIMED 670G SYSTEM USER GUIDE(M988666A021 1, M979513A011 1, M985214A021 1, M985214A261 1, M974439A181 1, M974439A062 1, M974439A032 1, M974439A043 1, M974439A073 1, M974439A103 1, M974439A052 1, M020098C001 3, M020098C003 1, M974439A165 1, M974439A093 1, M974439A083 1); MINIMED 670G SYSTEM USER GUIDE mg/dL(M974449A111 1, M974449A073 1, M974449A103 1, M974449A023 1, M974449A093 1, M974449A083 1, M974449A113 1); MINIMED 670G SYSTEM USER GUIDE mmol/L(M974439A024 1).

Distribution

Worldwide - US Nationwide distribution including in the states of OR, CA, AZ, NY,TX,FL, OH, MO, TN,NJ, NC, SC, IL, VA, WA, DC, GA, PA, MS,IA, MI, NH, OK, SD, MT, WI, WV, MA, LA, MN, RI, AR, ME, AL, IN, UT, KY, CO, ND, MD, CT, NM, KS, ID, NV, DE, NE, VT, AK, HI, WY, PR, VI and the countries of Taiwan, New Zealand, Australia, Bangladesh, Brunei Darussalam, Cambodia, Indonesia, Malaysia, Maldives, Philippines, Thailand, Vietnam, Singapore, Pakistan, Hong Kong, China, Argentina, Aruba, Barbados, Bermuda, Bolivia, Plurinational State Of, Brazil, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Honduras, Mexico, Panama, Paraguay, Peru, Puerto Rico, Trinidad And Tobago, Uruguay, Venezuela, Canada, South Korea, India, Albania, Algeria, Andorra, Armenia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bosnia And Herzegovina, Botswana, Bulgaria, Canary Islands, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Faroe Islands, Finland, France, French Guyana, French Polynes., Georgia, Germany, Gibraltar, Greece, Guadeloupe, Hungary, Iceland, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Liechtenstein, Lithuania, Luxembourg, Macedonia, Malawi, Malta, Martinique, Mauritius, Mayotte, Moldavia, Montenegro, Morocco, Namibia, Netherlands, New Caledonia, Nigeria, Northern Ireland, Norway, Oman, Poland, Portugal, Qatar, Reunion, Romania, Russia, San Marino, Saudi Arabia, Serbia, Sierra Leone, Slovakia, Slovenia, South Africa, Spain, St.Pier,Miquel., Sudan, Sweden, Switzerland, Syrian Arab Republic, Tanzania, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Yemen, Zambia, Zimbabwe.

Quantity

84,248