FDA Enforcement Class II Terminated

Eon - 3701 (Wall Charger) and 3711 (Portable Charger). Product Usage: Eon: Eon system is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back surgery syndrome, intractable low back and leg pain. Eon Mini: Eon Mini system is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back surgery syndrome, intractable low back and leg pain.

Recall: Z-1738-2013 · Reported July 24, 2013

Enforcement

Recall Number
Z-1738-2013
Event ID
63109
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
St. Jude Medical
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 24, 2013
Initiation Date
July 26, 2012
Classification Date
July 16, 2013
Termination Date
November 18, 2015
Address
6901 Preston Rd, N/A, Plano, TX, 75024-2508, United States

Description

Eon - 3701 (Wall Charger) and 3711 (Portable Charger). Product Usage: Eon: Eon system is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back surgery syndrome, intractable low back and leg pain. Eon Mini: Eon Mini system is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back surgery syndrome, intractable low back and leg pain.

Reason

St. Jude Medical has informed your doctor that a number of cases have been reported in which discomfort associated with heating occurred at the device site while patients were using the Charging System to charge their spinal cord stimulator. In three cases, patients received a burn to the skin (one second-degree and two first-degree burns) at the implant site. As of June 30, 2012, St. Jude Medica

Code Info

All serial numbers manufactured

Distribution

Worldwide Distribution - USA Nationwide Including Puerto Rico and countries of: Argentina, Australia, Belgium, Brazil, Canada, China, Chile, Colombia, Ireland, Israel, India, Malaysia, Mexico, Saudi Arabia, Singapore, South Africa, South Korea, Switzerland and United Kingdom.

Quantity

70,638 (combined total for Eon and Eon mini systems)