FDA Enforcement
Class II
Terminated
KRD1 PedFuse Rods, Model: IM80100-13 Rev B Product Usage: Pedicle rods are cylindrical rods which act as a linking mechanism between pedicle screws. The rod sits with the pedicle screw tulip and they are locked into place with a set screw or locking cap. The rod may be placed along several pedicle screws. Component rods are used during implantation of pedicle screw system devices intended to provide immobilization and stabilization of spinal segments of skeletally mature patients.
Recall: Z-1737-2013
·
Reported July 24, 2013
Enforcement
- Recall Number
- Z-1737-2013
- Event ID
- 64989
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- SpineFrontier, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Telephone
- Report Date
- July 24, 2013
- Initiation Date
- December 30, 2011
- Classification Date
- July 16, 2013
- Termination Date
- January 8, 2014
- Address
- 500 Cummings Ctr Ste 3500, N/A, Beverly, MA, 01915-6516, United States
Description
KRD1 PedFuse Rods, Model: IM80100-13 Rev B Product Usage: Pedicle rods are cylindrical rods which act as a linking mechanism between pedicle screws. The rod sits with the pedicle screw tulip and they are locked into place with a set screw or locking cap. The rod may be placed along several pedicle screws. Component rods are used during implantation of pedicle screw system devices intended to provide immobilization and stabilization of spinal segments of skeletally mature patients.
Reason
One of the rods in each of two test constructs did not meet fatigue requirement as outlined in ASTM F1717-04 dynamic compression testing 5 million cycles.
Code Info
Lot Number: 002876-007R
Distribution
USA Nationwide Distribution including the state of TX
Quantity
2 rods