BioZorb Marker-BioZorb 3D Bioabsorbable Marker-Intended Use radiographic marking of sites in soft tissue. 1) F0405 BioZorb Marker 4cm x 5cm 2) F0404 BioZorb Marker 4cm x 4cm 3) F0331 BioZorb Marker 1cm x 3cm x 3cm 4) F0231 BioZorb Marker 1cm x 3cm x 2cm 5) F0221 BioZorb Marker 1cm x 3cm x 2cm 6) F0304 BioZorb Marker 3cm x 4cm 7) F0303 BioZorb Marker 3cm x 3cm 8) F0203 BioZorb Marker 2cm x 3cm 9) F0202 BioZorb Marker 2cm x 2cm
Enforcement
- Recall Number
- Z-1733-2024
- Event ID
- 94359
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Hologic, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 15, 2024
- Initiation Date
- March 13, 2024
- Classification Date
- May 3, 2024
- Address
- 250 Campus Dr, N/A, Marlborough, MA, 01752-3020, United States
Description
BioZorb Marker-BioZorb 3D Bioabsorbable Marker-Intended Use radiographic marking of sites in soft tissue. 1) F0405 BioZorb Marker 4cm x 5cm 2) F0404 BioZorb Marker 4cm x 4cm 3) F0331 BioZorb Marker 1cm x 3cm x 3cm 4) F0231 BioZorb Marker 1cm x 3cm x 2cm 5) F0221 BioZorb Marker 1cm x 3cm x 2cm 6) F0304 BioZorb Marker 3cm x 4cm 7) F0303 BioZorb Marker 3cm x 3cm 8) F0203 BioZorb Marker 2cm x 3cm 9) F0202 BioZorb Marker 2cm x 2cm
Safety Information: Complaints describe complications/adverse events that include pain, infection, rash, device migration, device erosion, seroma, discomfort, or other complications from feeling the device in the breast, and the need for additional medical treatment to remove the device
UDI-DI: (1) 15420045514065; (2) 15420045514058; (3) 15420045514041; (4) 15420045514034; (5) 15420045514027; (6) 15420045514010; (7) 15420045514003; (8) 15420045513990; (9) 15420045513983. All lots, including implanted devices and non-implanted devices within expiration date.
US Nationwide Distribution
53492