CDI Blood Parameter Monitoring System 500 with Arterial blood parameter module. Intended for use during cardiopulmonary bypass procedures when continuous monitoring of arterial and/or venous pH, PCO2, PO2, K+, oxygen saturation, hematocrit, hemoglobin, and temperature are desired.
Enforcement
- Recall Number
- Z-1733-2013
- Event ID
- 60480
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Terumo Cardiovascular Systems Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 24, 2013
- Initiation Date
- February 27, 2013
- Classification Date
- July 13, 2013
- Termination Date
- February 17, 2016
- Address
- 6200 Jackson Rd, N/A, Ann Arbor, MI, 48103-9586, United States
Description
CDI Blood Parameter Monitoring System 500 with Arterial blood parameter module. Intended for use during cardiopulmonary bypass procedures when continuous monitoring of arterial and/or venous pH, PCO2, PO2, K+, oxygen saturation, hematocrit, hemoglobin, and temperature are desired.
Terumo discovered through its own testing that CDI 500 Blood Parameter Monitoring System software contains inaccuracies associated with a software algorithm, and a software update has been developed to address this issue and a premarket notification (K130425) has been submitted to the agency. Once clearance is received for this premarket notification from the US FDA, Terumo CVS will implement the
Catalog number: 500A* and serial numbers: 1211,1559-1563,1663-1666,1722-1727,1754-1758, 1776, 1795, 1951, 1952, 2016, 2112, 2117, 2190, 2246, 2273, 2274, 2322, 2355, 2356,2389-2391, 2393, 2394, 2397, 2634, 2700, 2801, 2804, 2860, 2890, 2891, 3011-3013, 3125, 3146, 3263, 3289, 3410, 3417, 3480, 3629-3631, 3646-3648, 3661, and 5349.
Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV and the countries of Jordan, CANADA, AUSTRALIA, Japan ,Hong Kong ,UNITED ARAB EMIRATES (UAE), BELGIUM, South Korea, Malaysia, Mexico, Philippines, Singapore, Taiwan, and Thailand.
3972 units since 4/26/12