FDA Enforcement Class I Ongoing

Vyaire Medical GmbH, Vyaire Twin Tube sample line 240 cm, Model Number V-707327; The TwinTube sample line 240cm is used in the Vyntus CPX. The TwinTube sample line 240cm is connected to the adapter for DVT. Twin Tube Probenschlauch (707004), which is sold outside the US only. Used in breath analysis for metabolic response study.

Recall: Z-1730-2024 · Reported May 15, 2024

Enforcement

Recall Number
Z-1730-2024
Event ID
94256
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Vyaire Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
May 15, 2024
Initiation Date
April 8, 2024
Classification Date
May 8, 2024
Address
Leibnizstr. 7, N/A, Hochberg, N/A, N/A, Germany

Description

Vyaire Medical GmbH, Vyaire Twin Tube sample line 240 cm, Model Number V-707327; The TwinTube sample line 240cm is used in the Vyntus CPX. The TwinTube sample line 240cm is connected to the adapter for DVT. Twin Tube Probenschlauch (707004), which is sold outside the US only. Used in breath analysis for metabolic response study.

Reason

Potential of the nozzle separating during patient use.

Code Info

UDI/DI 14250892903603, All lots prior to June 2023 or with a serial number before 04I00122 are potentially affected.

Distribution

US nationwide distribution.

Quantity

649 devices