FDA Enforcement Class I Ongoing

Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. ROBOTIC PACK, Medline SKU # DYNJ58475F; 2. URO/GEN ROBOTIC PACK, Medline SKU # DYNJ60803J.

Recall: Z-1728-2026 · Reported April 15, 2026

Enforcement

Recall Number
Z-1728-2026
Event ID
98598
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Medline Industries, LP
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
April 15, 2026
Initiation Date
February 27, 2026
Classification Date
April 7, 2026
Address
3 Lakes Dr, Northfield, IL, 60093-2753, United States

Description

Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. ROBOTIC PACK, Medline SKU # DYNJ58475F; 2. URO/GEN ROBOTIC PACK, Medline SKU # DYNJ60803J.

Reason

Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

Code Info

Medline SKU # DYNJ58475F, UDI/DI each 10198459572388, UDI/DI case 40198459572389, Lot Number: 25LMK173; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 26BMA268; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 26AMH928; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 25EMD876; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 25EMA519; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 25DMA846; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 25BMH004; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 25BMA218; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 24KMI208; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 24KMC007; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 24JMD850; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 24IMI939; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 24EDB694; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 24DDB793; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 24CDB207; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 24BDB439; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 23GDB756; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 23GDA532.

Distribution

Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).

Quantity

439 kits