FDA Enforcement Class II Terminated

Nikon Metrology XTH cabinet x-ray Systems which includes variations STH225, XTH225ST, and MCT225.

Recall: Z-1727-2020 · Reported June 3, 2020

Enforcement

Recall Number
Z-1727-2020
Event ID
85498
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Nikon Metrology
Voluntary / Mandated
FDA Mandated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
June 3, 2020
Initiation Date
December 23, 2019
Classification Date
May 22, 2020
Termination Date
June 13, 2024
Address
12701 Grand River Rd, N/A, Brighton, MI, 48116, United States

Description

Nikon Metrology XTH cabinet x-ray Systems which includes variations STH225, XTH225ST, and MCT225.

Reason

The manufacturer identified a risk that their products might fail to comply with the cabinet x-ray system performance standard limit on x-ray emissions [21 CFR 1020.40(c)(1)] due to an insufficient safety margin for the means of attachment of a piece of interior x-ray shielding. Double sided industrial tape was used for attachment - there have been only 2 instances of detachment and although the piece of shielding detached it remained mostly in place supported by wires that pass through that piece of shielding. The repair is accomplished by adding a bracket to fasten the piece of shielding securely to the inside of the x-ray system. Mfr reports that no systems were found to have exceeded the emission limit when checked prior to repair.

Code Info

all STH225, XTH225ST, and MCT225 units)

Distribution

US Nationwide Distribution

Quantity

110 units