B-CrossLaps/serum (B-CTx in serum) REF 11972308 160 100 tests Immunoassay for the in vitro quantitative determination of degradation products of type I collagen in human serum and plasma as an aid in assessing bone resorption. The test may be used as an aid in monitoring antiresorptive therapies (e.g. bisphosphonates, hormone replacement therapy - HRT) in postmenopausal women and individuals diagnosed with osteopenia. The electrochemiluminescence immunoassay ECLIA is intended for use on Elecsys and cobas e immunoassay analyzers.
Enforcement
- Recall Number
- Z-1727-2013
- Event ID
- 65163
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Roche Diagnostics Operations, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 24, 2013
- Initiation Date
- April 25, 2013
- Classification Date
- July 12, 2013
- Termination Date
- September 27, 2013
- Address
- 9115 Hague Rd, N/A, Indianapolis, IN, 46256-1025, United States
Description
B-CrossLaps/serum (B-CTx in serum) REF 11972308 160 100 tests Immunoassay for the in vitro quantitative determination of degradation products of type I collagen in human serum and plasma as an aid in assessing bone resorption. The test may be used as an aid in monitoring antiresorptive therapies (e.g. bisphosphonates, hormone replacement therapy - HRT) in postmenopausal women and individuals diagnosed with osteopenia. The electrochemiluminescence immunoassay ECLIA is intended for use on Elecsys and cobas e immunoassay analyzers.
Roche Diagnostics internal investigations in R&D have shown that the claims for biotin interference and the sample stability given in the Elecsys ¿-CrossLaps (¿-CL) assay package insert version 5, could not be reproduced. The ¿-CrossLaps test results contribute to the monitoring of antiresorptive therapy. False low or false normal results may lead to delayed diagnosis and treatment of osteoporosi
All lot numbers
Nationwide distribution: USA including states of: AL, CA, GA, FL, IL, IN, KS, KY, MA, MD, MI, MN, MO, NC, NY, OH, PA, TX, VA, VT, and WA.
3056 units