FDA Enforcement Class II Terminated

Leica Ceiling Mount MSF-1 for use with M841 and M501 operating microscopes. Article Number 10446781 utilizes a supply voltage of 230 V. Article Number 10446782 utilizes a supply voltage of 110 V. Intended to be used for improving the visibility of objects through magnification and illumination.

Recall: Z-1725-2014 · Reported June 18, 2014

Enforcement

Recall Number
Z-1725-2014
Event ID
66274
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Leica Microsystems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 18, 2014
Initiation Date
September 11, 2013
Classification Date
June 6, 2014
Termination Date
April 24, 2017
Address
1700 Leider Ln, N/A, Buffalo Grove, IL, 60089-6622, United States

Description

Leica Ceiling Mount MSF-1 for use with M841 and M501 operating microscopes. Article Number 10446781 utilizes a supply voltage of 230 V. Article Number 10446782 utilizes a supply voltage of 110 V. Intended to be used for improving the visibility of objects through magnification and illumination.

Reason

The ceiling mount supplier encountered an event in which the drive screw in the motor unit, which lifts and lowers the lower lift arm, broke. As a result, the lift arm fell down approx. 60 cm (23.6 inches).

Code Info

Leica Ceiling Mount MSF-1; Article Numbers: 10446781, 10446782

Distribution

Worldwide Distribution -- USA, including the states of AZ, IL, MA, MO, NC, OR, TX, VA, and WI; and, the countries of Australia, Austria, Chile, China, Colombia, Czech Republic, Greece, Israel, Italy, Japan, Malaysia, Netherlands, Norway, Oman, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, and United Kingdom.

Quantity

103 total